Hardware control Flashcards

1
Q

High field scanners typically have field strengths greater than or equal to:

A

1.5 Tesla

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Consistent room temperature of 65-75°F, humidity of 50-70%, and sufficient air flow in the MR scan room is recommended to

A

Make patients less vulnerable to the RF heating effects

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

In a cross sectional view of the MR magnet design, which coils are closest to the patient?

A

Radiofrequency coils

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

The liquid cryogens are cooled in a superconducting magnet at a temperature of

A

4° Kelvin

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Which of the following is NOT a magnetic field associated with MR imaging?

A

Gravitational earth magnetic fields

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

The purpose of shimming the magnet is to

A

make the B0 field as homogenous as possible

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

The array processor is responsible for:

A

Reconstructing the images using the Fourier transform

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

: In order for energy to be efficiently transferred between a transmitter and receiver, the two must be at the same

A

Frequency

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

The _____ is located within an equipment room.

A

RF power amplifier

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

During a quench of the MR system, patients and employees should be evacuated from the scan room to avoid

A. Damage to tympanic membranes

B. Frostbite

C. Asphyxiation

D. All of the above

A

all of the above

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Field strength at magnet isocenter is measured in units of:

A

Tesla

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

The unit of measurement of the magnetic field surrounding the periphery of the MR scanner is expressed as:

A

Gauss

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

The intensity limit for a static magnetic field in clinical use is limited by the Food and Drug Administration (FDA) to _____________?

A

4 Tesla

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

The patient image display with sequence parameter visualization is located in the

A

Scan control room

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

The local RF coil used to image a human brain is typically located/kept in the __________.

A

Magnet room

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

The CPU used to plot and adjust slices in an MR sequence is located:

A

Scan control room

17
Q

Radiofrequency (RF) shielding can be achieved by lining the scanner room walls with

A

Copper

18
Q

The location of the coils installed in the MRI scanner, from the inner point closest to the patient to the outer most edge, are

A

Radiofrequency coils, gradient coils, shim coils, main magnet

19
Q

The liquid cryogens most often used to maintain field strength in a super conducting MR system are:

A

Helium

20
Q

The coil construction in an MR system, from the innermost aspect to the outer:

A

Radiofrequency coils, gradient coils, shim coils, main magnet

21
Q

Doubling the magnetic field strength will:

A

Yield an increase of double the SNR

22
Q

: What type of magnet is used with most systems currently manufactured

A

Superconducting

23
Q

: Image 78 is an example of what type of MR scanner?

A

Open magnet

24
Q

Image 79 is an example of which MR system?

A

Superconducting magnet

25
Q

The direction of the magnetic field in Image 79 is:

A

Horizontal

26
Q

: Image 82 was acquired at what field strength?

A

1.5 Tesla

27
Q

Passive shielding can be accomplished by lining the MR room with ___________.

A

Steel

28
Q

Active shielding can be defined as:

A

Superconducting windings within the MR scanner, reducing the fringe magnetic field

29
Q

_________ would be used to contain the RF in an MR magnet room.

A

Copper

30
Q

___________ could be used to line the walls and/or floors of the MR magnet room to reduce the fringe field, and is defined as __________ shielding

A

Steel; passive