What sizes and bends are there
F18 85 deg, F22_20, F22_180
Lengths of pe catheter and right heart catheter
F18 is 105 cm and the F 22 is 77 cm
sheath size for both
F18 uses at 22 fr and the F22 uses a 26 fr.
what is behind your office
intercostal waterway 3000 mile system of connected bays rivers inlets and canals that run along the atlantic and gulf coasts. there are two 1. atlantic Runs for norfolk va to key west.
2. gulf - brownsville texas to apalachicola
How do you do the vacuum lock (vacuum lock rocker)
for engaging large clot you and lock the back switch.
two setting of the volume switch limited
10 cc /30 cc
what is the tube on the top of the device
9f accessory port used for ancillary devices or over the wire exchanges as well as picutres thought the device
How big is the funnel once exposed
11 mm fr is mmx 3 so 33 fr
What did Moriarty call it
“a melding of devices” based on angiovac.
setup steps
- back of the handle there is a lock mechanism unlock it and insert the obturator
- lock then flush the obturator and the side port.
- place over a wire via na approaprate sheath and advacne to pa. unlock the back handel and remove the obturator.
- need to bleed back on the system prior to use via the accesory port on teh handle to deaid
preferred side to start
right, frequently you can do left to right
emerging the inner funnel cannual
unsheath it
pulls
mainly doing 10 cc, 30 save for main pa to clain main pa
- multipurpose 6 gr ( 9fr catheter 7 fr sheath)
- oversheath back to the pulmonic valve and then flip
- picutres on the left. torque and advance with tech doing 10 cc then back to a 30 cc pulls
what do you do for lollipopping on the funnel
- bring funnel cannula down. Squeez cannular an lock. hold funnel still and advance oversheath pull funnel out bid advance
tell me about the new hemostasis valve on the back
3 stops - wide open, wire access, and fully closed (rarely used)