Knife Flashcards
1
Q
EK-417D
Working length: O cutter length: I cutter length: Max diameter: Min passable channel:
A
Working length: 2300 +/- 100mm O cutter length: 4 +/- .6mm I cutter length: 2 +/- .3mm Max diameter: >= 2.4mm Min passable channel: 2.8mm
2
Q
EK-425D
Working length: O cutter length: I cutter length: Max diameter: Min passable channel:
A
Working length: 2300 +/- 100mm O cutter length: 4 +/- .6mm I cutter length: 1.5 +/- .2mm Max diameter: >= 2.4mm Min passable channel: 2.8mm
3
Q
I Knife Recommended Output Power Upper GI Tract AutoCut mode Forced coag mode
A
AutoCut mode: 20-30W
Forced coag mode: 30-50W
4
Q
I Knife Recommended Output Power Lower GI Tract AutoCut mode Forced coag mode
A
AutoCut: 20W
Forced coag mode: 20-30W
5
Q
EK-417D
I knife length
A
2 +/- .3mm
6
Q
EK-425D
I knife length
A
1.5 +/- .2mm
7
Q
O Knife Recommended output power Upper GI tract AutoCut Forced Coag mode
A
AutoCut: 30-40W
Forced Coag mode: 40-60W
8
Q
O Knife Recommended output power Lower GI tract AutoCut Forced Coag mode
A
AutoCut: 20W
Forced Coag mode: 20-30W
9
Q
EK-417D
O Knife Length
A
4 +/- .6mm
10
Q
EK-425D
O Knife length
A
4 +/- .6mm
11
Q
Rated cutting voltage
A
1600Vp (3200Vp-p)
12
Q
Rated coag voltage
A
2900Vp (5800Vp-p)
Caution: higher voltage is not allowed
13
Q
Contraindications
A
- patients with high anesthetic risk
- patients who are unable to tolerate the endoscopic examination
- patients with cardiac pacemaker
- patients with severe comprise of the vital organs including heart, lungs, and kidneys
- patients with severe anemia or infection or patients with other high risks
- patients who are taking anti-coagulant IR anti platelet agents or patients with coagulation disorders
- patients with non-lifting signs, namely, injection of normal saline into the sub mucosa of the lesion base cannot result in local uplift