SP-Fix Flashcards

0
Q

What issues might happen with the Lateral Barrel Holder?

A

They arms of the holder don’t match up. Plate can go beyond the barrel as it articulates down. Do not completely lower the barrel before the plate is lowered.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
1
Q

can a surgeon mix and match plate sizes?

A

yes

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

How could an implant come unlocked during surgery?

A

Pin isn’t in the locked position, or when going to cut the rod the surgeon rocks the pin cutter back and forth and the pin/rod becomes unlocked.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

What issues might the scrub have loading the Arch plate?

A

They might load them with the curve of the plate cephalad instead of caudal.

Not locking the rod to the barrel or plate properly.

Might try to load the locking plate first instead of the pivot plate. Always load the pivot plate first, then the locking plate.

The locking plate is always on the left side.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Does the scrub tech need to load the AP inserter in the caddie?

A

Yes, always.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

When reading the distractor spindel, do you down size the implant ?

A

yes

How well did you know this?
1
Not at all
2
3
4
5
Perfectly