Anesthesia workstations and components are covered by _____
ASTM (American Society for Testing and Materials)
Computer driven machines should be turned off and restarted how often?
Every 24 hours
Only what should be plugged into machines electrical outlets?
Anesthesia monitors only
Backup battery power is _____
At least 30 minutes
When gas travels from high pressure cylinder to the machine a _____________ decreases pressure to _______
Pressure regulator decreases pressure to 50 psi
Hanger yoke 3 functions
- Orients and supports the cylinder
- Provides a gas-tight seal
- Ensures unidirectional flow
Components exposed to high pressure
- hanger yoke
- cylinder pressure gauge
- cylinder pressure regulators
___________ allow gas to enter machine but prevent gas from exiting the machine when no cylinder in yoke
Check valves
Cylinder and pipeline pressure gauges are __________
Bourdon gauges- metal needles which stretch with pressure and point to correct pressure
Pneumatic components of intermediate pressure system
- master switch
- oxygen failure devices
- flow control devices
Accessories supplied by oxygen (4)
- Ventilators
- Accessory flowmeters
- Jet ventilation
- Oxygen flush
Flush buttons direct flow to ___________ at ______lpm.
______ require machine to be on
Common gas outlet at 35-75 lpm
Do not require master switch to be on
4 flush hazards
- Leaks
- Sticking and obstructing flowmeter output
- Barotrauma
- Awareness
Machines possess a pneumatic or electronic alarm device that sounds a warning when oxygen supply pressure decreases below _________
30 psi
The ________ will decrease supply pressure of other gases as the oxygen supply pressure decreases below _________
Fail-safe valve (oxygen pressure sensor shut-off valve)
Below 20-25 psi
Reducing decides just upstream from flow valves decrease pressure a second time to _______
15-19 psi
Machine components exposed to low pressure system
Flowmeter tubes
Vaporizers
Check valves
Common gas outlet