131-humidity final Flashcards
(50 cards)
What is the point in the respiratory tract where inspired gas reaches body temperature, ambient pressure, saturated BTPS conditions
Isothermic Saturation Boundary
Which of the following factors causes the isothermic saturation boundary to shift farther down into the airways? I. Decreased ambient temperature II. Increased tidal volume III. endotracheal intubation a. 1 and 2 b. 2 and 3 c. 1 and 3 d. 1, 2, and 3
D all of the above
What is the primary goal of humidity therapy?
a. decrease airway reactivity to cold
b. maintain normal physiological conditions
c. deliver drugs to the airway
d. reduce upper airway inflammation
b. maintain normal physiological conditions
Indications for the warming of inspired gas include all of the following except:
a. treating a patient whose airways are reactive to cold
b. providing humidification when the upper airway is bypassed
c. treating a patient with a low body temperature (hypothermia)
d. reducing upper airway inflammation or swelling
ANS: D
Inhalation of dry gases can do which of the following?
- increase viscosity of secretions
- impair mucociliary motility
- increase airway irritability
a. 1 and 2
b. 3
c. 1 and 3
d. 1, 2, and 3
ANS: D
As the airway is exposed to relatively cold, dry air, ciliary motility is reduced, airways become more irritable, mucous production increases, and pulmonary secretions become inspissated (thickened due to dehydration).
Which of the following inspired conditions should be maintained when delivering medical gases to the nose or mouth?
a. 50% relative humidity (RH) at 20° to 22° C
b. 100% RH at 20° to 22° C
c. 80% RH at 28° to 32° C
d. 100% RH at 32° to 35° C
ANS: A
Which of the following inspired conditions should be maintained when delivering medical gases to the hypopharynx, as when administering oxygen by nasal catheter?
a. 50% relative humidity (RH) at 20° to 22° C
b. 100% RH at 20° to 22° C
c. 95% RH at 29° to 32° C
d. 100% RH at 32° to 35° C
ANS: C
See Table 35-1
What device adds molecular water to gas?
a. agitator
b. humidifier
c. nebulizer
d. atomizer
ANS: B
A humidifier is a device that adds molecular water to gas.
Which of the following inspired conditions should be maintained when delivering medical gases directly into the trachea through an endotracheal tube or a tracheotomy tube?
a. 50% relative humidity (RH) at 20° to 22° C
b. 100% RH at 37° to 42° C
c. 95% RH at 29° to 32° C
d. 100% RH at 32° to 35° C
ANS: D
See Table 35-1
Clinical indications for delivering cool humidified gas include which of the following?
- post-extubation edema
- upper airway inflammation
- croup (laryngotracheal bronchitis)
- epiglottitis
a. 2, 3, and 4
b. 1, 2, and 3
c. 1, 2, 3, and 4
d. 2 and 3
ANS: C
The delivery of cool humidified gas is used to treat upper airway inflammation resulting from croup, epiglottitis, and post-extubation edema.
. What is the most important factor determining a humidifier’s performance?
a. surface area
b. temperature
c. time of contact
d. gas flow
ANS: B
Temperature is an important factor affecting humidifier performance
Gas leaving an unheated humidifier at 10° C and 100% relative humidity (absolute humidity = 9.4 mg/L) would provide what relative humidity at body temperature?
a. about 100%
b. about 60%
c. about 40%
d. about 20%
ANS: D
Although the humidifier fully saturates the gas, the low operating temperature limits total water vapor capacity to approximately 9.4 mg/L water vapor, equivalent to approximately 21% of body humidity
What is the simplest way to increase the humidity output of a humidifier?
a. Increase the time that the gas and the water are in contact.
b. Increase the surface area between the water and the gas.
c. Decrease the water vapor pressure of the gas.
d. Increase the temperature of either the water or the gas.
ANS: D
Simply heating the humidifier to 40° C (Figure 35-3, right) increases its output to 51 mg/L, which is more than adequate to meet BTPS conditions
Which of the following types of humidifiers are used in clinical practice?
- heat-moisture exchanger
- passover humidifier
- bubble humidifier
a. 1 and 2
b. 2 and 3
c. 1 and 3
d. 1, 2, and 3
ANS: D
Active humidifiers typically include: (1) bubble humidifiers, (2) passover humidifiers, (3) nebulizers of bland aerosols and (4) vaporizers. Passive humidifiers refer to typical heat and moisture exchangers (HMEs).
. What is the goal of using an unheated bubble humidifier with oronasal oxygen delivery systems?
a. Raise the humidity of the gas to ambient levels.
b. Fully saturate the inspired gas to body temperature, ambient pressure, saturated (BTPS) conditions.
c. Cool the gas down to below room temperature.
d. Fully saturate the inspired gas to ambient temperature, ambient pressure, saturated (ATPS) conditions
ANS: A
Unheated bubble humidifiers are commonly used with oronasal oxygen delivery systems (see Chapter 38) the goal is to raise the water vapor content of the gas to ambient levels.
Why should you NOT heat the reservoirs of a bubble humidifier used with an oronasal oxygen delivery system?
a. Condensate will obstruct the delivery tubing.
b. Heating will melt the reservoir or cause a fire.
c. Heating will absorb the extra water vapor.
d. Heating will causes too much aerosol impaction
ANS: A
Heating the reservoirs of these units can increase humidity content but is not recommended because the resulting condensate tends to obstruct the small-bore delivery tubing to which they connect
To protect against obstructed or kinked tubing, simple bubble humidifiers incorporate which of the following?
To protect against obstructed or kinked tubing, simple bubble humidifiers incorporate which of the following?
a. HEPA outlet filter
b. pressure relief valve
c. automatic hygrometer
d. electronic alarm system
ANS: B
When checking an oxygen delivery system that incorporates a bubble humidifier running at 6 L/min, you occlude the delivery tubing, and the humidifier pressure relief immediately pops off. What does this indicate?
a. malfunctioning humidifier
b. normal, leak-free system
c. malfunctioning flowmeter
d. leak in the delivery tubing
ANS: B
If the system is obstructed at or near the patient interface and the pop-off sounds, the system is leak free
A design that increases surface area and enhances evaporation by incorporating an absorbent material partially submerged in a water reservoir that is surrounded by a heating element best describes what type of humidifier?
a. bubble
b. wick
c. cascade
d. hygroscopic
ANS: B
Typically a wick is placed upright with the gravity-dependent end in a heated water reservoir.
Which of the following are types of heat-moisture exchangers (HMEs)? Which of the following are types of heat-moisture exchangers (HMEs)? 1. simple condenser 2. hygroscopic condenser 3. hydrophobic condenser a. 1 and 2 b. 2 and 3 c. 1, 2, and 3 d. 3
ANS: C
There are three basic types of HMEs: (1) simple condenser humidifiers, (2) hygroscopic condenser humidifiers, and (3) hydrophobic condenser humidifiers
A heat-moisture exchangers has an efficiency rating of 80%. What does this mean?
a. Of the exhaled water vapor, 80% returns to the patient on inspiration.
b. The inspired temperature is 80% of the expired temperature.
c. Of the exhaled water vapor, 20% returns to the patient on inspiration.
d. The device provides 80% relative humidity at body temperature.
ANS: A
According to the National Institute of Standards and Technology of the U.S. Department of Commerce, a gas cylinder that is color-coded brown and green should contain which of the following?
He Oxide
The gauge on an E cylinder of O2 reads 800 psig. About how long would the contents of this cylinder last, until completely empty, at a flow of 3 L/min
Formula: Pressure x cylinder factor / flow l/min
Answer: 74.666 or 1 hour 14 min
Note the cylinder factor is 0.28 for E cylinder
What do the components of a liquid O2 bulk storage tank include?
I. inner and outer steel shells, separated by a vacuum
II. safety system that vents O2 if warming occurs
III. vaporizer system that converts liquid O2 to gas
IV. pressure-reducing valves to lower pressure to 50 psig
A. I, II, and III
B. II and IV
C. I, II, III, and IV
D. III and IV
C-all of the above