Breathing Circuits Flashcards

1
Q

Time Constants Equation

A

vol of circuit (L)/FGF L/min = min, 3 time constants to get to 95% of vaporizer settings

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Four Time constants?

A

98.2%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

One Time Constant?

A

63.2%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Two Time Constants?

A

86%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

What is more important in turbulent flow - gas density or gas viscosity?

A

Density

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Generalized turbulent flow

A

results when flow of gas through tube exceeds certain value = critical flow rate

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Localized turbulent flow

A

results when gas flow is below critical flow rate but encounters constrictions, curves, valves or other irregularities

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Critical Velocity

A

greatest velocity with which a fluid can flow through a given conduit without becoming turbulent

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What constitutes the breathing system?

A

Essentially FG outflow to scavenging

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Function of BC

A

o Direct oxygen to patient
o Deliver ax gas to patient
o Remove CO2 from inhaled breaths
o Means for controlling ventilation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Insufflation

A

gases delivered directly to patient airway

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Open Breathing Systems

A

patient inhales only mixture delivered by ax machine, valves direct each exhaled breath into atmosphere
 +/- RBB
 Rebreathing minimal, no CO2 absorbent

Higher FGF rates

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Semi Open Systems

A

exceeds minute ventilation, generally rebreathing does not occur (depends on FGF), 150mL/kg/min

No chemical absorption of CO2
RBB, unidirectional valves optional
Exhaled gases flow out or mix with FGF

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Semi Closed (most machines)

A

exceeds MVO2, allows partial rebreathing
 Part of exhaled gases passes into atmosphere, part mixes with FGF
 Chemical absorption of CO2, directional valves, bag
 22-44mL/kg/min

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Low flow?

A

10-15mL/kg/min

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Closed Systems

A

Complete rebreathing: 3-10mL/kg/min
Directly matches MVO2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

What is rebreathing?

A

composition of inspired gas mix of fresh gas, rebreathed gas

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

What is non-rebreathing?

A

composition of inspired gas same as fresh gas from machine

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

Rebreathing

A

inhale previously expired gases from which CO2 may or may not be removed

Possible to have complete rebreathing without increasing CO2

FGF: no RB if vol of fresh gas per minute > P’s minute volume

Mechanical dead space: vol in breathing circuit occupied by gases that are rebreathed without changing in composition

Breathing system components: arrangement can increase or decrease

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

What happens with excessively high FGFs in RB or NRB systems?

A

Minimal rebreathing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

What happens with low FGF in NRB?

A

No complete preventing of rebreathing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

Causes of increased FiCO2

A

-Decreased FGF
-Increased dead space
-expired absorbent
-stuck expiratory valve

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

Resistance

A

tracheal tube, important factor when determining WOB

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

Compliance

A

change in vol over change in pressure, measure of distensibility (mL/cm H2O)

most distensible components = reservoir bag, breathing hoses

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
Q

Effects of Rebreathing

A

o Heat, moisture retention
o Altered inspired gas concentrations

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
26
Q

Causes of increased inspired vol vs delivered vol

A

–FGF > than rate at which absorbed by patient or loss through leaks in BS when ventilator in use

–FGF delivered during inspiration added to VT delivered by ventilator

–Increased FGF, IE ratios; decreased RR

Eliminated by modern ventilators

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
27
Q

Decreased inspired vol vs delivered vol

A

Gas compression, distension of components during inspiration = wasted ventilation
–wasted vent increased with increased airway pressure, VT, BS vol, component distensibility

Smaller patients > larger patients

VT decreases DT leaks in BS
–Measure via comparing inspired, expired VT
–Measuring tidal volume at end of expiratory limb will reflect increase caused by FGF, decrease from leaks; will miss decrease from wasted ventilation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
28
Q

Features that Cause Discrepancy btw Inspired, Delivered Oxygen, Ax Gas Concentrations

A
  1. Rebreathing
  2. Air Dilution
  3. Leaks
  4. Ax agent uptake by BS components
  5. Ax agents released from system
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
29
Q

Discrepancy btw Inspired, Delivered Oxygen, Ax Gas Concentrations: rebreathing

A

 Depends on volume of rebreather gas, composition

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
30
Q

Discrepancy btw Inspired, Delivered Oxygen, Ax Gas Concentrations: air dilution

A

 Negative pressure in breathing system may cause air dilution if leak
* When fresh gas supplied per respiration <VT
 Concentration of anesthetic in inspired mixture to decrease, difficult to maintain stable anesthetic state

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
30
Q

Discrepancy btw Inspired, Delivered Oxygen, Ax Gas Concentrations: Leaks

A

Positive pressure in system will force gas out of system

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
31
Q

Discrepancy btw Inspired, Delivered Oxygen, Ax Gas Concentrations: Ax agent taken up by breathing system

A

 taken up or adhere to rubber plastics metal and CO2 absorbent
 Directly proportional to concentration gradient btw gas/components, partition coefficient, surface area, diffusion coefficient, square root of time

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
32
Q

Discrepancy btw Inspired, Delivered Oxygen, Ax Gas Concentrations: Ax agent released by breathing system

A

 Elimination of anesthetic agent from breathing system depends on same factor as uptake functions
 Essentially low output vaporizers after vaporizer turned off

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
33
Q

Connectors

A

join two pieces together
o Uses: extend distance btw patient, breathing system; change angle of connection btw patient and breathing system,
o Resistance increases with sharp curves, rough sidewalls
o Add dead space btw patient and breathing system
o increased number of locations disconnection can occur

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
34
Q

What are the fittings on the ETT, FG outlet?

A

15mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
35
Q

What are the fittings on everything but the ETT, FG outlet?

A

22mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
36
Q

Definition of a rebreathing system

A

Unidirectional flow of gas, means of absorbing CO2 from expired gases

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
37
Q

Full (complete) rebreathing

A

o Flow rates at or near patient MVO2
o 3-14mL/kg/min

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
38
Q

Partial rebreathing

A

o FGF > MVO2, less than that required to prevent rebreathing
o Low flow = 20-50mL/kg/min
o So-so (medium) flow = 50-100mL/kg/min
o High flow 100-200mL/kg/min

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
39
Q

Non (minimal) RB

A

o >150-200mL/kg/min, may result when circle systems used for small patients (<5kg) with FGF >1L/min
o Most modern vaporizers continue to function optimally down to 200-500mL/min

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
40
Q

Pros of Lower FGF rates

A

More economical in terms of oxygen, inhalant
Less environmental contamination by inhalants (halogenated hydrocarbons)
Improved maintenance of body temp

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
41
Q

Cons of Lower FGF

A

Decreased ax gas delivery
Time required to change ax concentration within circuit significantly increasess

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
42
Q

What are the components of a circle system?

A

fresh gas inlet
inspiratory one way valve, inspiratory/expiratory breathing tubes, expiratory valve
APL valve, reservoir bag
CO2 cannister

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
43
Q

What is true about low flows and anesthetic equilibrium?

A

Takes longer to achieve than high flows

Inspired gas concentrations will not reflect vaporizer settings until nearing equilibrium

Low FGF more commonly used in LA

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
44
Q

Fresh Gas Inlet

A
  • Site of gas delivery from common gas outlet of ax machine
  • After CO2 absorber, before inspiratory valve
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
45
Q

Inspiratory Valve (aka flutter valves)

A

opens via neg pressure from breath or pos pressure from vent -> gas moves from FGI, reservoir bag to valve in inspiratory limb

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
46
Q

Structure of OVW

A

o Clear dome – direct visualization of valve function
o Cage or guide mechanism (prongs) – prevent disc dislodgement
o Light weight valve with hydrophobic disc – prevent condensation from sticking, which increases resistance to opening
o Valve housing – seat, guides
o Removable cover – access for cleaning, repair, drying

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
47
Q

MOA OWV

A

Gas enters at bottom, raises disc from seat

Gas passes under dome, on through breathing system

Reversing gas flow causes disc to contact seat, preventing retrograde flow

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
48
Q

OWV expiration

A

closed, prevents exhaled gas from entering inspiratory limb -> forces entry into expiratory limb

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
49
Q

Positioning of OWV

A

Veritical or Horizontal

o Vertical valves decrease resistance to gas flow

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
50
Q

Matrx circle system

A

negative pressure relief valve, alternative path of gas flow (room air) to patient should inspiratory valve stuck closed

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
51
Q

Movement of valve

A

does not guarantee valve competence

Incompetent valve = less resistance to gas flow, flow of gas will go more easily through incompetent valve -> rebreathing

Expiratory > inspiratory, exposed to more moisture

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
52
Q

Inspiratory Port

A

Downstream of inspiratory unidirectional valve, 22mm OD male connector

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
53
Q

Breathing Hoses

A

Corrugated plastic or rubber inspiratory, expiratory limbs

Prevents kinking, allows expansion if BC subjected to traction, compression
o 2 x 22mm ID male connector ports for connection to breathing hoses

Connected to ETT via wye piece – 15mm ID female connectors

Dead space extends from wye piece to patient

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
54
Q

Does length of tubes affect dead space?

A

NO!

Length of tubes does not alter amt of dead space or rebreathing DT unidirectional gas flow

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
55
Q

Coaxial Systems

A

Decrease bulk assoc with breathing system, facilitate warming of inspired gases by expiratory gases
o Concentric, side by side
o Universal F circuit, can also be used as NRB

Inner (inspiratory) tube ends just before connection to patient
o Inner tube to patient, exhaled gases flow to absorber assembly via outer corrugated tubing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
56
Q

Disadvantages of coaxial systems

A

Increased resistance, increased dead space if leak in inner tube/difficult to catch, if gas flow reversed - increased resistance to exhalation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
57
Q

Y Piece

A

Three-way tubular connector with 2x22mm male ports for connection to breathing tubes, 15mm female patient connector for trach tube or supraglottic airway device

Most often permanently attached to breathing hoses

DEAD SPACE

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
58
Q

Expiratory Port

A

Upstream of expiratory valve, 22mm OD male connector port

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
59
Q

What is the most compliant part of a circle system?

A

Rebreathing Bag (application of LaPlace’s Law)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
60
Q

Expiratory Valve

A

Closes on inspiration, opens on expiration

Direct gas into expiratory limb of breathing system

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
61
Q

Rebreathing Bag

A

o 22mm female connector at neck
o Tail, +/- loop to hang for drying

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
62
Q

RBB Purpose

A

Compliant reservoir of gas that changes vol with patient’s expiration, inspiration

Allows gas to accumulate during exhalation, provides reservoir of gas for next inspiration
–Rebreathing, economical gas use, prevents air dilution

Assist, control of ventilation

Visual/tactile observation of patient’s spontaneous respirations

Protects from excessive pressures by being source of compliance in the system (LaPlace’s law)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
63
Q

MOA RBB

A

Addition of vol, pressure increases rapidly to peak then reaches plateau

As bag distends further, pressure falls slightly = max pressure that can develop in breathing system

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
64
Q

RBB Size?

A

Vol 5-10x VT (10-20mL/kg), ~ patient’s minute volume (VT x RR) -> no consensus

Usually attached via 22mm male bag port

ASTM standards:
o <1.5L bags: pressure shall not be <30, >50cm H2O when bag expanded 4x size
o >1.5L bags: <35, >60

*Latex free bags: allow higher pressures to develop than latex bags
*New bags: greater pressures when overinflated than old bags or if pre stretched

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
65
Q

What is another name for APL valve?

A

Heidbrink Valve

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
66
Q

Function of APL

A

Allows excess gas to escape from patient circuit during expiration
o If functioning properly, should escape of pressure >1-3cm H2O

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
67
Q

Anatomy of APL

A

Control part: controls pressure at which valve opens
–Spring-loaded disc
–Stem, Seat
–Control Knob
–Collection Device, exhaust port

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
68
Q

Spring Loaded Disc on APL

A

disc held onto seat via spring, threaded screw cap over spring allows variation of pressure exerted by spring on disc

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
69
Q

How spring loaded disc on APL opens

A
  • increased pressure in BS –> upward force on spring, upward force > downward force of spring –> disc rises, gas flows through valve
  • Cap maximally open: little resistance of spring
  • Weight, pressure of disc ensures reservoir bag fills before seat rises
  • increases pressure downstream –> increases pressure needed to open valve, PEEP
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
70
Q

Stem, Seat of APL

A

Threaded stem, variable contact with seat

Valve opens –> opening of seat comes larger, more gas escapes

Ensures unidirectional gas flow (no backflow of gas from scavenge flowing back into breathing system), supplies slight pressure to keep breathing bag inflated

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
71
Q

Control Knob of APL

A

Rotatory control knob: ASTM -> always closed clockwise, arrow indicator

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
72
Q

Collection device, exhaust port

A
  • Excess gases collected, directed to scavenge system via transfer tubing
  • Exhaust port: site of discharge, 19 or 30mm male connector
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
73
Q

APL Use

A

Fully open at all times unless PPV (can be isolated from system via bag switch), CPAP

Can be partially closed to prevent collapse of reservoir bag DT negative pressure/vacuum from scavenge system
o Risk of complete closure –> excessive pressure built up in system
o Ensure adequate inspiratory pressure

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
74
Q

Bag/Vent Selector Switch

A

Shift rapidly between manual respiration, automatic ventilation without removing bag or ventilator hose from mount
o Essentially 3 way stopcock

When selector switch in ventilator position, APL valve isolated from circuit –> does not need to be closed

Newer machines: turning ventilator on causes electronically controlled valves to direct gases into proper channels

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
75
Q

CO2 Canister Structure

A

Transparent to monitor color change of absorbent

Screens: hold absorbents in place

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
76
Q

Two canisters in series vs one

A

Smaller canisters allow more frequent changes in [FGF] to be reflected more quickly, improve ventilator performance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
77
Q

Large Canisters

A

Longer intervals btw absorbent changes
Risk desiccation if in absorbent for long time

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
78
Q

Small Canisters

A

Decreased likelihood of CO, compound A production (fresh absorbent with proper water content)
Decreased interval vol of BS
Must change more frequently

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
79
Q

Canister Housing

A

Space at top, bottom: promotes even distribution of flow through absorber

Space at bottom: accumulation of dust, water

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
80
Q

Canister Distribution Pattern

A

No difference whether gases enter at top or bottom

Start at inlet, progress down sides -> migrates throughout rest of canister

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
81
Q

Canister Baffles

A

annular rings, direct gas flow toward central part of cannister
o Compensate for reduced flow resistance along walls

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
82
Q

Side, center tube of canister

A

conduct gases to/from bottom of canister, return to patient

83
Q

Bypass of Canister

A

In older absorbers - allow exhaled gases to completely or partially bypass absorber
o Allow CO2 accumulation
o Canister removed, not replaced - creates bypass

84
Q

CO2 Absorbent MOA

A
  • Base neutralizing acid –> carbonic acid + CO2 forming water, carbonate; releases heat
    o Can predict dryness by measuring % of water in outflow gas
    o Very widely inability to absorb nitric oxide, nitrogen dioxide - Monitoring downstream from absorber
85
Q

High Alkali Absorbents

A

When desiccated, react with VAs –> CO, CA (sevo)

Do not change color when dry, capacity to absorb CO2 decreased by moisture

86
Q

Low Alkali Absorbents

A

reduced NaOH, KOH; smaller amts CO, Cmp A?

87
Q

Alkali Free Absorbents

A

–CaOH + small amts other agents to accelerate CO2 absorption, bind water
–No evidence of CO with any VA, little or no CA formation even if desiccated
–Indicator: changes color when dry
Once exhausted, do not revert to original color
o CO2 capacity < high alkali
–Do not deteriorate when moisture lost

88
Q

Lithium Hydroxide

A

o Reacts with carbon dioxide to form carbonate
o Does not react with anesthetic agents, even if desiccated
o Expensive, requires special handling - burns to eyes, skin, respiratory tract

89
Q

Indicators

A

Acid, base: color = pH-dependent, added to absorbent to signify when ability to absorb CO exhausted

Does not affect absorption

Ethyl violet most common, pH 10.3 -> bright, vivid color change

90
Q

Phenolphthalein

A

White –> Pink

91
Q

Ethyl Violet

A

White to Purple

92
Q

Clayton Yellow

A

Red to yellow
(Clay turns yellow)

93
Q

Ethyl Orange

A

Orange to yellow
(sunsets fade)

94
Q

Mimosa Z

A

Red to white
Empty the mimosa leaves a clear glass

95
Q

How granule size measured

A

Mesh Number
**Higher the mesh number, smaller the particles – most absorbents 4 to 8 mesh
**4 mesh strainer: 4 openings per square inch, 4 mesh will pass through but not through strainer with smaller holes (eg 8 mesh strainer)
**8 mesh strainer: 8 openings per square inch

95
Q

Pellets, small granules

A

greater surface area, decreases channeling along low-resistance pathways
 Potential for more resistance, caking

96
Q

How absorbent changes consistency

A

Fresh granules crumble more easily
Used absorbents become hard - formation of CaCO3

97
Q

Dust Production of Absorbents

A

Some granules fragment easily, producing dust

Excessive powder: channeling, resistance to flow, caking, may be blown through system to patient or cause system components to malfunction

Small amounts of hardening agent added to prevent

98
Q

Sodalime Composition

A

80% CaOH, 15% H2O, 4% NaOH

99
Q

How much CO2 can sodalime absorb?

A

25L CO2/100g sodalime

100
Q

Baralyme composition

A

20% BaOH, 80% CaOH +/- KOH

101
Q

How much CO2 can baralyme absorb?

A

27L CO2/100g baralyme
FIRE DANGER

102
Q

Amsorb composition

A

80% CaOH, 1% Ca sulfate hemihydrate, 3% CaCl, 15% water

103
Q

How much CO2 can amsorb absorb?

A

~12.5L CO2/100g

Minimal to no CO formation, minimal CA production, least amt of sevoflurane degradation

104
Q

What are the human recommendations for absorbents?

A

only non-desiccated absorbents with no KOH, little to no NaOH

105
Q

Haloalkene Formation

A

o Halothane degradation during closed circuit anesthesia
o Produces haloalkene 2-bromo-2-chloro-1, 1-difluoroethene (BCDFE)
–Nephrotoxic in rats

106
Q

Compound A Formation: Factors that increase Production

A

Low FGF**
KOH, NaOH absorbents
Higher absorbent temperature** (more exhaled CO2)
Higher concentrations sevoflurane**
Longer duration of ax
Dehydrated absorbents**
Use of barium lime**

107
Q

Factors that decrease production of compound A

A

Low alkali, alkali free absorbents
Lower absorbent temperature
(smaller canisters = lower temps)

108
Q

CO formation

A

Desflurane, enflurane, isoflurane passed through dry absorbent containing strong alkali

Sevoflurane degraded by absorbent, CO formed if temperature exceeds 80*C

109
Q

Most to least potential for CO formation?

A

Des > Enflurane > Iso&raquo_space;> Halothane

110
Q

Most common situation of CO formation?

A

Most commonly during first anesthetic of day on Monday, fresh dry gas flowing into the circle system over the weekend causing absorbent to become dehydrated
 Or machine don’t use very often

Concentration in BS varies with time, peak in first 60 minutes

111
Q

How detect CO formation?

A

Capnograph will appear normal (will not detect)

Pulse oximeter: read carboxyhemoglobin as oxyhemoglobin, unless COHgb levels very high then see slight drop (will not detect)

Not detected by currently used respiratory gas monitors

112
Q

What might suggest have a problem with CO?

A

Multi wave infrared analyzers may provide warning of isoflurane or desflurane breakdown by displaying wrong agents or mixed agents
* Unusually delayed rise or unexpected decrease in inspired concentration of VA in BS vaporizer setting
* Failed inhalation induction
* Inadequate anesthesia
* Unexpected decrease in inspired concentration

113
Q

Factors That Increase Production of CO

A

KOH, NaOH absorbents*
Dry absorbents
* – extended period of FGF during non-use
Des > En > Iso&raquo_space;> halothane**
Sevo if cannister >80
C
High FGFs: produce more but decreased concentrations bc remove it (also promote decissation)

114
Q

Factors that decrease production of CO

A

None if normally hydrated
CO2 absorption

115
Q

How to Minimize Risk of CO Formation

A

o Turn off all gas flows after each case, disconnect medical gas pipeline system at the end of the day
o Turn off vaporizers when not in use
o Change absorbent at least once a week, label with filling date, check date as part of daily machine checkout
o Machines not in use should not be filled with absorbent, fill with fresh before each use
o Supplying oxygen to a patient not receiving GA via circle system strongly discouraged
(Ideally FM that connected directly to oxygen pipeline system or auxiliary O2 flow meter on machine)
o Avoid using FG to drive breathing system components
o Negative pressure relief valve on closed scavenging systems should be checked regularly
o Failure of valve to pull room air may result in FGF from machine being drawn through absorbent if APL valve open
o Monitor temperature in the canister change absorbent if excessive heat detected

116
Q

Barium Hydroxide (BaOH), Baralyme

A
  • interaction with sevoflurane = temperatures of several 100*C
    o Reports of fires, melted components of canisters - removed from market
    o Soda lime: less elevated temperatures, fires involving desiccated soda lime reported
117
Q

Storage and Handling of Absorbents

A

–Supplied in several types of containers
–Once opened resealed ASAP to prevent absorbent reaction with CO2 in air, indicator deactivation, moisture loss
–Temps <32*F: moisture expands granules, fragmentation
–Gentle handling to avoid fragmentation, dust formation -> irritating to eyes, respiratory tract, Caustic to skin particularly when damp
–Important not to overfill, small space at top to promote even gas flow through canister

118
Q

How, When to Change Absorbent

A

Most Reliable Method = appearance of CO2 inspired gas
Indicator color change
Heat in canister
Detection of desiccation

119
Q

Color Change Limitations

A

 Peaking, regeneration seen with absorbance that contains strong bases
* Appears to be reactivated with rest
* Absorbent that shows exhausted color if allowed to at rest often show color reversal – false Impression of usefulness
 Absorbents without strong base change color when dried
 Channeling: absorbent along channels will become exhausted, CO2 will pass through canister
* Color change may not be visible if on inside of cannister
 Absorbent may not contain an indicator

120
Q

Ethyl Violet

A

undergoes deactivation even if stored in dark
* Deactivation accelerated in presence of light, especially high intensity or ultraviolet light

121
Q

Heat in Canister

A

Changes in absorbent temperature occur earlier than changes in color of the indicator

Some heat production should be apparent unless high FGF being used

If temp of downstream canister exceeds that of upstream canister –> change absorbent of upstream canister

122
Q

To and Fro Circuits

A

*Horses
*Waters’ cannister: patient breaths in, out of closed bag connected to ETT/face mask via canister containing absorbent
*FG introduced at patient end of system, APL usually mounted close by

Similar to Mapleson B

123
Q

Disadvantages of To and Fro

A
  • Absorbent at patient end becomes exhausted first –> increases functional dead space of system
  • Cumbersome, risk of inhalation of absorbent dust (resolved with filter)
124
Q

Logistics of To and Fro

A
  • Part of system btw p, soda lime = dead space –> vol must be kept to minimum
    o Canister must be close to patient’s head, technical challenges
    o Horizontal position of canister recommended
    o Must be well packed to avoid channeling/incomplete absorption
125
Q

NRB

A

Absence of unidirectional valves to direct gases, no device for absorbing CO2, FGF must wash CO2 out of circuit
o High FGF rates to flush CO2 from circuit
o Not used for patients exceeding 10kg = less economical DT high FGF
o Recommended FGF: 150-300mL/kg/min

126
Q

Maintenance of patients on NRB

A

Traditionally recommended patients <5kg DT lower resistance during breathing, less equipment dead space, smaller total circuit volume

Possible to maintain patients <5kg safely using rebreathing systems provided patients VT adequate to accentuate unidirectional valves

127
Q

What is also important about NRB circuits?

A

No clear separation of inspired, expired gases
o Rebreathing when inspiratory flow > FGF
o Best way to determine optimal FGF = monitoring ETCO2

128
Q

General Principle of NRB

A

During expiratory pause, high FGF pushes exhaled gas from previous expiration down exhalation conducting tube away from patient toward reservoir bag

Inspiration: inspires gas coming from FGI, exhalation conducting tube
–Under normal circumstances majority of inspired breath comes from FG in exhalation conducting tube

129
Q

Causes of RB with seemingly sufficient FGF

A

patient with rapid breathing –> inadequate expiratory pause to wash CO2 distal from patient end especially if large breath taken

130
Q

Advantages of NRB

A

–Simple, inexpensive, lightweight, not bulky – no moving parts except APL
–Easy to disinfect/sterilize
–Resistance usually low at clinically used FGFs
–WOB during SpV < circle system
–Variations in minute volume affect ETCO2 less than circle system
–Compression and compliance losses have less effect than compared to circle system
–Changes in FG concentrations = changes in insp gas concentration (no RB)
–No absorbent, no production of toxic compounds

131
Q

Disadvantages of NRB

A

–High FGFs: higher costs, increased pollution, difficulty assessing SpV
–Inspired heat and humidity tend to be low
–Can be difficult to determine optimal FGF
–Anything that decreases FGF risks RB
–ABC systems: APL close to patient, inaccesible to provider, SS use awkward
–E, F: difficult to use with SS, air dilution can occur with E
–Not suitable for patients with MH –> not possible to increase FGF enough to remove increased CO2 load

132
Q

Resp Gas Monitoring with Mapleson Systems

A

All but A have fresh gas inlet near patient connection port –> difficult to get reliable sample of exhaled gases

133
Q

Mapleson A System

A

–FGF away from patient, minimum 50-80mL/min
–Cannot be used with mechanical ventilator that vents excesses gases - entire thing will become dead space
–Most efficient with SpV

134
Q

Lack Modification of Mapleson A

A

Added expiratory limb, runs from patient connection to APL valve at machine end of system

Easier to adjust valve, facilitates scavenging excess gases

Slight increased WOB
 Coaxial, parallel tube arrangements

135
Q

How can a Mapleson A system be as efficient as a D during CMV?

A

If APL doesn’t vent during inspiration

136
Q

Basic Flow of Gas through a Mapleson A during SpV

A

Exhalation: GDS+ Galveolar flow toward bag as FGF flows into bag

Bag becomes full, pressure in system increases until APL opens, continuing FGF reverses direction of exhaled flow –> Galv vented first, followed by Gds if FGF high enough

–Lower FGFs: GDS retained within system, rebreathing
–If much lower FGFs: alveolar gas also retained

137
Q

Mapleson B

A

FGF 1.5-2.0x minute vent
FGI, APL located near patient end (B for Buddies)
RB away patient end separated from FGI by corrugated tubing

138
Q

Mapleson B Gas Flow - SPV

A

APL valve opened completely, excess gas vented through valve during exhalation

139
Q

Mapleson B Gas Flow - AsV, CMV

A

Closing APL sufficiently to allow lungs to be inflated, excess gases vented during inspiration

Slightly more efficient than Mapleson A bc fresh gas accumulates at patient end of tubing during exploratory pause

Variable performance during CMV

140
Q

Mapleson C

A

B without corrugation

SpV: almost as efficient as A when expiratory pause minimal
 Less efficient as expiratory pause increases
 FGF 1.5-2x minute ventil
o CMV: FGF 2.5-3x minute ventil

141
Q

Mapleson D

A

FGF 1.5/2-3x min ventil

FGF at patient end, bag/APL opposite
Easy gas scavenging

Sensor, gas sampling – btw corrugated tubing/T piece, corrugated tubing/APL, bag/mount, T piece and patient (adults)

Bidirectional PEEP valve – btw APL, corrugated tubing
* Some PEEP valves close with negative pressure application, SpV impossible with that kind of valve in system

T piece

142
Q

T piece in Mapleson D, E, F

A

3-way tubular connector with patient connection port, FG port, port for connection to corrugated tubing

143
Q

Bain Modification

A

Mapleson D

FGF coaxially inside corrugated tubing, ends at point where FG would enter if classic D

Outer tube clear, allows visual inspection of inner tube

Metal head with holes drilled in, allows for fixed position of reservoir bag, APL
MRI compatible versions: static compliance increased with decreased PIP, VT at same ventilator settings, increased PEEP

144
Q

Mapleson D SpV

A

APL open, excess gases vented during expiration

145
Q

AsV Mapleson D

A

partial closure of APL, excess vented during inspiration

146
Q

CMV Mapleson D

A

connect hose from ventilator in place of reservoir bag, closing APL, excess gases vented through ventilator spill valve

147
Q

ETCO2 with Mapleson D system

A

ratio of minute vol to FGF, absolute values

148
Q

Hazards of a Bain System

A

entire circuit becomes dead space
 Inner tube detached from connections
 FG supply tube kinked, twisted
 Incorrect system assembly
 Mix of FG, exhaled gas if defect in metal head

149
Q

Leak Test of Mapleson A

A

occlude patient end, close APL valve, pressurize system

Coaxial lack system: must test integrity of inner tube
* Blow down tube with APL valve closed – leak btw two limbs produces movement of bag
* Occlude both limbs at patient connection with APL open, squeeze bag – leak in inner limb –> bag collapses (gas escapes via APL)

150
Q

Low FGF with Mapleson D

A

Upon inspiration, previously expired gas still in tubing – not flushed out by FGF –> rebreathing

Higher humidity, less heat loss, greater fresh gas economy
Elevated ETCO2 will require more work by patient to return to normal levels

151
Q

How to decrease ETCO2 with low FGF with Mapleson D

A

high IE time ratio (prolongation of expiratory time allows more time for gas clearance)
slow rise in inspiratory flow rate (pulls FGF)
low flow rate during last part of exhalation
long expiratory pause (allows FGF to push exhaled gases toward APL)

152
Q

High FGF and Mapleson D System

A

 Little rebreathing bc fresh gas flushes exhaled gas from tubing
 Assoc with increased heat, humidity loss
 Better in patients with stiff lungs, poor cardiac performance, hypovolemia

153
Q

Relationship btw ETCO2 and FGF in a Mapleson D System

A

Low FGF: ETCO2 depends on FGF
High FGF: ETCO2 depends on ventilation

154
Q

Preuse check: Mapleson D

A
  • Occlude patient and, close APL valve, pressurize system
  • APL valve opened dash bag should deflate easily if valve, scavenging system working properly
155
Q

Mapleson E

A

o Does not have a bag
o Expiratory port, closed in chamber from which excess gases evacuated
o Numerous modifications of original T piece
o Use: more commonly used to administer oxygen or humidified gases to spontaneously breathing patients

FGF 2-3x minute ventilation
Resp cycle sequence of events similar to Mapleson D

156
Q

Mapleson E SpV

A

expiratory limb open to atmosphere
 No rebreathing if no exhalation limb
 If expiratory limb, FGF needed to prevent rebreathing equal same as Mapleson D
 Air dilution if no expiratory limb or volume of limb less than patient’s VT  FGF > peak inspiratory flow rate

157
Q

CMV Mapleson E

A

intermittent occlusion of expiratory limb, allow FG to inflate lungs
 no rebreathing, only FG inflates lungs
 No air dilution

158
Q

Hazards with Mapleson E

A

 intermittent occlusion of the expiratory limb for controlled ventilation arrow overinflation, barotrauma
 No bag –> No tactile feel during inflation as with other systems, no pressure-buffering effect
 No APL valve to moderate pressure in lungs

159
Q

What is a unique use of a Mapleson D system?

A

Modified Mapleson D system for one lung ventilation to apply CPAP to non dependent lung

160
Q

Mapleson F

A

Bag with mechanism for venting excess gases
–Hole in tail, side of bag that occluded by using finger to provide pressure
–Ventilator may be used in place of bag

APL valve placed near patient connection to provide protection from high pressure

FGF 2-3x minute vent

SS: enclose bag in chamber from which waste gas suctioned or attaching devices to relief mechanism in bag

161
Q

Modifications of Mapleson F System

A

o Ayre’s T piece: no scavenging

o Jackson Rees modification of Ayre’s Y piece: attached RBB

162
Q

Mapleson F SpV

A

Relief mechanism left open

163
Q

Mapleson F AsV

A

relief mechanism occluded sufficiently to distend bag, then squeeze bag

164
Q

Mapleson F CMV

A

Bag replaced by hose from ventilator

165
Q

Hazards of Mapleson D

A

 Same as for E
 Bag in system - excessive pressure less likely to develop
 Use of ventilator with ram of oxygen to produce inspiration with T piece system: disconnection at common gas outlet may not be detected by airway pressure monitor DT high resistance of fresh gas tubing

166
Q

Humphrey’s

A

Convert Mapleson A system to D/E to utilize most efficient modes
o A: spontaneous, DEF: CMV

167
Q

Humphrey’s Level up in A mode

A

RBB on ADE block connected to inspiratory pathway (Map A)

Breathing hose connecting block to patient = insp limb

Expired gas carried back along other limb, vented at APL

Minimizes any potential mixing of inspired, expired gas

APL away from patient end of system = decrease Apparatus dead space

168
Q

Humphrey’s Lever Down

A

o RBB, APL isolated from BS
o Inspiratory limb in A now delievers gas to patient end as a T piece
o Breathing hose returning gas to ADE block now functions as reservoir limb a T piece
o No RBB = Map E

169
Q

PEEP Valves

A

Electronically controlled, disposable, manually controlled

Add peak end expiratory pressure to system

170
Q

Fixed pressure PEEP valves

A

indicate amount of peep provide, >1 = additive effect

171
Q

Variable PEEP valves

A

adjust amount of peep using scale or manometer

172
Q

Unidirectional vs bidirectional PEEP valves

A

Bidirectional: second flow channel with own one way valve, recommended
–Incorrect orientation = no application of PEEP

Unidirectional peep valve incorporated incorrectly against flow of gas in inspiratory, expiratory limb –> block gas flow

173
Q

When not see PEEP indicated on pressure gauge?

A
  1. Incorrect application of bidirectional valve
  2. Pressure gauge on absorber side of expiratory unidirectional valve
174
Q

Filters

A

Protect patient from microorganisms, airborne particulate matter

Protect anesthesia equipment, environment from exhaled contaminants

ASA, CDC recommendations: no recommendation for placing in BS unless patient has infectious pulmonary disease

175
Q

Consequences of filters in pediatric systems

A

Increase resistance, relative inefficiency

176
Q

Mechanical Filters

A

(pleated hydrophobic): physically prevent microorganisms, particles from passing

177
Q

Electrostatic Filters

A

rely on electrostatic forces to hold organisms within loosely woven charged filter element, less effective than mechanical filters

178
Q

HEPA Grade Filters

A

High Efficiency Particulate Aerosol

trap >99.7% particles with diameter 0.3micrometers

179
Q

Problems with Filters

A

o Should not be placed downstream of humidifier, nebulizer - less efficient when wet
o increased resistance, dead space esp if condensation
o Obstruction (increased PIP)
o Leak
o Erroneous entitle gas concentrations, poor CO2 waveforms

180
Q

Administration of Bronchodilators

A
  • Shake well prior to administration, discharge maximal when canister upright
  • Slow deep inspiration followed by pause of 2-3s before exhalation - enhanced amount of medication deposited into airway
    o 30 to 60 seconds between puffs
181
Q

Humidification and Bronchodilators

A

high humidification causes aerosol droplets to increase in size –> rain out

182
Q

Goals of a circle system?

A
  1. Minimize absorbent decussation
  2. Maximum inclusion of FG in inspired mixture, maximum venting of alveolar gas
  3. Minimal consumption of absorbent – use absorbent as little as possible
  4. Accurate readings from respirometer placed in system
  5. Maximal humidification of inspired gases
  6. Minimum dead space
  7. Low resistance
  8. Minimal pull on airway equipment
  9. Convenience
183
Q

What are the three requirements of any circle system configuration?

A
  1. UNIDIRECTIONAL VALVES MUST GO BTW RBB AND PATIENT
  2. FGF CANNOT ENTER BTW EXPIRATORY VALVE AND P - WOULD NEVER REACH PATIENT
  3. APL VALVE CANNOT BE BTW PATIENT AND INSP VALVE - WOULD GET VENTED OUT BEFORE REACHING PATIENT
184
Q

Anestar: differences btw classic circle system

A

RBB near FGI, ventilator upstream to inspiratory valve; FG decoupling valve downstream of RBB

CMV: fresh gas decoupling, no spill valve

185
Q

Ohmeda ADU differences with classic circle system

A

o FGF enters system btw inspiratory unidirectional valve, Y piece
o RBB, APL, ventilator (spill valve) = classic circle system position

with continuous FGF, DT location of inspiratory valve, no retrograde FGF through absorbent

186
Q

Drager 6400 vs classic circle system

A

RBB near FGI, ventilator (piston) = exhalation side of absorber downstream of expiratory unidirectional valve

APL valve, mechanical exhaust valve = exhalation side of absorbent

FG control valve

Mechanical exhaust valve

187
Q

Fabius GS, Apollo BS

A

FG decoupling valve
Ventilator btw FG decoupling valve, inspiratory unidirectional valve
RBB btw expiratory unidirectional valve, absorber

188
Q

Dead Space of the Circle System

A

Patient port of Y piece to partition between inspiratory, expiratory tubing

When exhalation or inhalation starts, gases and breathing tubes move in opposite direction from usual flow until stopped by closure of one of unidirectional valves = backlash
o clinically insignificant if unidirectional valves competent
o Causes slight increase in dead space

189
Q

Heat, Humidity

A

Moisture: from exhaled gases, absorbent, water liberated from neutralization of carbon dioxide

Gases and inspiratory limb of circle system near room temperature
o Even at low FGFs, only 1-3*C above room temp

190
Q

How to Increase Humidity in a circle system

A

lower FGF
increasing ventilation
fresh gas upstream of absorber
wetting inspiratory tubing
humidifier smaller
canisters
coaxial breathing tubes

191
Q

What is the most important determinant of the internal vol in the circle system?

A

Canister size

192
Q

Relationship btw Inspired, Delivered Concentrations

A

No rebreathing: concentration of gases, vapors and inspired mixture will be close to those in fresh gas –> everything being delivered to patient will be newly added to system

Larger BS internal volume, greater the difference btw inspired, delivered concentrations

193
Q

Nitrogen

A

Before any FG delivered, concentration of nitrogen in BS ~80%
 Enters via exhaled gases, leaves via APL valve, ventilator spill valve, leaks
 Hinders establishing high [N2O] may cause low inspired oxygen concentrations

194
Q

Denitrogenation

A

using FGF for few minutes to eliminate most of nitrogen in system
 After denitrogenation, elimination by patient will proceed at slower rate

195
Q

Side Stream Analyzers and Nitrogen

A

If side stream gas monitor directs gases back to anesthesia circuit, nitrogen concentration may increase –> many analyzers entrain air as reference gas
 leak in sampling line can also entrain air

196
Q

carbon dioxide: with absorbent

A

should be near 0
 Unless failure of one or more unidirectional valves
 High FGF: limits increase in FiCO2

197
Q

CO2 without absorbent

A

depends on FGF, arrangement of components in circle system, ventilation

198
Q

Oxygen

A

affected by:
– rate of oxygen uptake by patient
–uptake/elimination of other gases by patient
–arrangement of components
–ventilation
–FGF, volume of BS
–concentration of oxygen in fresh gas

199
Q

Anesthetic Agents

A

Removed via absorption, degradation –> slower inductions, exposure of subsequent patient to VAs in system

Dry absorbent removes more than wet

200
Q

Things that affect relationship btw inspired, delivered concentrations of inhalants

A

 Uptake by patient
 Uptake by components of the system
 Arrangement of system components
 Uptake, elimination of other gases by patient
 Volume of system
 Concentration in FGF
 Degradation by absorbent
 FGF

201
Q

Low flow with circle system

A

technique in which a circle system with absorbent used at FG inflow of less than patients alveolar minute volume

Closed system anesthesia: form of low flow anesthesia in which FGF = uptake of ax gases, oxygen by patient, system and gas sampling – no venting by APL

202
Q

Requirements of Low Flow Ax with Circle System

A

Requires flow meter that provide low flows, vaporizers calibrated to be accurate at low FGF or VIC

Must continuously measure oxygen concentration

Must increase FGF to compensate for gases removed by monitor with side stream gas analyzer

203
Q

If closed circuit used, constant circuit volume achieved by:

A

RBB: increased size, increased FGF; increased size, decreased FGF

Ventilator with ascending bellows: adjust FGF so bellows below top of housing at end of exhalation

Ventilator with descending bellows: adjust FHF so bellows just reaches bottom of housing at end of exhalation

Closed system ax: high flows be used for 1-2’ at least 1x/hr to eliminate gases that have accumulated in system (Nitrogen, CO2)
o must increase FGF if rapid change in any component of inspired mixture desired

204
Q

Advantages of Low Flow with Circle System

A

o Significant savings, esp with VAs
o Less personnel exposure: decreased VA in OR, vaporizers filled less (less exposed during filling)
o Better environmental impact
o Lower FGF, longer takes for change in FGF to cause comparable change in inspired concentrations
o Inspired humidity increases, rate of fall of body temp decreases
o Dangerously high pressures in BS take longer to develop