ABG, O2 analyzers, Capnography Flashcards
(33 cards)
-ABG Kit
-Most ABG kits come prepackaged and contains all necessary equipment
-The things you will find in kit -Gauze -Band-aid -Provided Label (if you dont have anything from the chart then write PT info on label -Alcohol Prep and lodine -Needle cap and bung -Syringe with heparin and vented so blood feels the syringe -1.23 gauge needle and 1.18 gauge needle
Possible hazards when drawing an ABG
• Hematoma-Not holding correct amount of pressure.
• Cutting of the artery/Arterial laceration
• Hemorrhage-Clot breaking lose or not allowed to clot then therapist leaves the room.
• Vasovagal reaction/nerve response.
• The artery has a cluster of nerves that wrap around the artery, thus when drawing from the radial artery— those nerves causes tremendous amounts of pain.
Causing the sympathetic nervous system response to pain
• Losing a limb is the biggest complication when drawing an ABC. The patient can lose the limb is the artery is sliced.
Best Places to stick
-1st choice Radial
-2nd Choice Brachial -3rd Femoral
Site Selection
-Prior to sticking with an ABG all therapists must obtain a modified Allen’s test which is used to ensure the PTs have collateral circulation/ blood flow at the wrist. Can also use ( anesthesia medication) if PT or doctor request it.
The Angles of ABG
-Radial Artery 45 degrees
-Brachial Artery 60 to 90 degrees -Femoral Artery 90 degrees
How long to hold pressure
-Site must have continuous pressure for at least 5 minutes
-Up to 30 minutes on PT on blood thinners
Erroneous Values from Air Bubble
Handling
• Blood gas specimen should collected anaerobically
• Expel air bubbles immediately
In Vivo Values Air Contamination
pH. 7.40 To 7.45
PCO2. 40. To. 30
PO2. 95. To. 110
• The amount of blood should be at least
1 - 2cc within the syringe.
• Therapist should run sample ASAP.
• If sample is placed in ice it will keep longer.
• Glass syringe-1 hour
• Plastic syringe-15 minutes
When moving the ABG
• Transport ABG to the run location in a biohazard bag.
Internal parts of ABG machine.
Electrodes
• Utilized to Measure values of ABC
pH, PC02, РОг
All other blood gas values are Calculated
The pH Electrode
• Called Sanz Electrode
PCO2 Electrode
• Called Severinghaus Electrode
PO2 Electrode
Called Clark Electrode
• Labeled as a polarographic electrode
• Sometimes cleaning of the tip with pumice may be required, because the polypropylene attracts protein thus causing buildup.
Calibration Process
• 2-Point Calibration
• Uses low concentration and a high concentration range is used at both ends of the physiological range to be measured (every 30 minutes)
• Multiple-Point Calibration (3 or more points)
• Used to Verify whether the gas analyzer is linear/within range or not. (Every shift)
• Has two separate buffers that provide approximate values of:
• 6.840 buffer
• referred to as the zero point or low point buffer
• 7.384 buffer
• high point or slope point buffer
• Standard deviation for pH is
+ .005 or 99.5
PO2 & PCO2 Electrode
• Tank One
• Low CO2 (5%) - balance
• High 02 (12% or 20%)
• Balance Nitrogen
• Tank Two
• High CO2 (10%) - slope
• 02 (0%)
• Balance Nitrogen
• Must convert tank concentration from % to mmHg
(Pe - PH2O) x tank concentration = mm Hg
(760 - 47) x 0.12 = 85.65 mm Hg
Quality Control Plotting
• In control
• Trend
• Random Error
• Out of Control
Quality control l • In control
• All QC runs are within the acceptable SD
Quality Control• Trend
• All QC runs within the acceptable +SD but trending towards one side of the SD
Quality Control • Random Error
• All QC runs within the acceptable +SD except for one run
Quality Control • Out of control
• Two or more QC runs are out of the acceptable +SD
Oxygen Analyzers
Oxygen analyzers are a huge part of developing and maintaining appropriate care. They measure the concentration of oxygen before or while patients are using oxygen. If the oxygen isn’t correctly delivered, then the patient is receiving the correct amount.. Polarographic analyzers use a Clark electrode to measure oxygen. Although it actually measures PO2, the polarographic analyzer usually displays the percent oxygen.
These devices are accurate at any altitude, provided the instrument is calibrated at that altitude.