Acronyms Flashcards
(104 cards)
COC stands for?
certificate of conformance
RWP stands for?
rated working pressure
WPS stands for?
welding procedure specification
What standard governs welding and brazing procedures, activities, and personnel qualifications?
ASME Boiler and Pressure Vessel Code (BPVC) Section IX
AWS stands for?
American Welding Society
The distance between the outlets on a typical lubricator design ranges between?
14 to 17 inches.
The most common RWP for lubricators are:
2,000 psi, 3,000 psi, 5,000 psi, and 10,000 psi
Chapy impact testing is required for the design validation of materials and welds intended for service at or below __°F?
0°F
What is the minimum KE rating allowed on a lubricator?
1,150 in-lb (130 J)
Do design changes require validation and verification?
A design that undergoes a substantive design change becomes a new design requiring design verification as specified in 5.11 and design validation as specified in 5.12.
Who determines whether a design change is substantive or non-substantive for validation and verification purposes?
A qualified person after considering (as a minimum):
a) stress levels (static and dynamic) of the modified or changed components;
b) material changes;
c) functional changes.
When introducing substantive changes to the design of a subcomponent of an assembly, shall re-validation and re-verification be performed on the top level assembly?
No. This may be done by testing only the changed component or series of components, rather than the entire assembly. The test shall simulate the loading conditions that would be present if the entire assembly were tested. The supplier/manufacturer shall document the detailed test results and analysis that demonstrate that the component test simulates the required loading conditions. Evaluation results shall be approved by a qualified person other than the person performing them, and records of the results shall become a portion of the design documentation.
What is the prescribed time frame within which a user or purchaser should anticipate the receipt of design validation and verification documents from the supplier or manufacturer?
Documents and data may be in any type of media, e.g. hard copy or electronic media. All documents and data shall be available for review and/or audit by the user/purchaser within 30 days of a request.
What is the prescribed timeframe for data retention?
Design documentation shall be retained for a minimum of 5 years after a lubricator design has been obsoleted.
Who is responsible for providing the Functional Specification?
The user/purchaser shall prepare a functional specification to order products that conform with this specification and specify the following requirements and operating conditions, as applicable, and/or identify the supplier/manufacturer’s specific product. These requirements and operating conditions may be conveyed by means of a dimensional drawing, datasheet, or other suitable documentation.
Who is responsible for providing the Technical Specification?
The supplier/manufacturer shall prepare a technical specification that conforms to the requirements defined in the functional specification. If the technical specification does not fully meet the functional requirements, the supplier/manufacturer shall identify the differences to the user/purchaser. The supplier/manufacturer shall also provide to the user/purchaser the product datasheet defined in 6.2.5 and have available an operator’s manual and design documentation as defined in 5.1.2 and 6.2.
Measuring and monitoring equipment used during the testing shall be calibrated as defined in 6.8. All measurements for acceptance shall be within the calibrated range(s) of the testing equipment. All pressures are defined as gauge unless otherwise specified and shall be recorded on time based equipment.
For inspection, measuring, and testing equipment without a calibration history, what is the maximum allowed calibration interval?
Calibration intervals shall be a maximum of 3 months until a recorded calibration history can be established. Intervals may then be lengthened or shortened based on documented repeatability, amount of usage, and calibration history.
What is the maximum verification interval for standards used in the calibration process of inspection, measuring, and testing equipment?
Calibration standards used to calibrate measuring equipment shall be checked and approved at least once every 3 years by an independent outside agency with traceability to the applicable recognized national or international standards agency.
What is the maximum allowed calibration interval for inspection, measuring, and testing equipment with a satisfactory calibration history?
Intervals may then be lengthened or shortened based on documented repeatability, amount of usage, and calibration history. The calibration interval shall not be increased by more than twice the previous interval and shall not exceed 1 year.
Lubricators furnished to 11PL specifications may be
For product marking and identification purposes, is sharp V-stamping an acceptable process for any part or location within a lubricator?
Marking shall use low-stress (dot, vibration, or rounded V) methods. Conventional sharp V-stamping is acceptable in low-stress areas, which may include the OD of flanges. Sharp V-stamping is not permitted in a high-stress area unless subsequently stress relieved per supplier/manufacturer specifications.
Body impact rating is defined as _________
A rating based on the kinetic energy (KE) rating of the body of the lubricator design in the event that a KE absorption device is not functioning.
When required by the supplier/manufacturer for the material or welds, impact tests shall meet the requirements of _______?
ASTM A370 using the Charpy V-notch technique