Analysis and testing Flashcards

1
Q

What is analytical method transfer? How do you transfer assay of a tablet?

A
  1. Change Control:
    Initiate a change control process to evaluate and document the need for the transfer, assessing regulatory and quality impacts.
    1. Gap Analysis & Impact Assessment:
      Compare key aspects between the transferring and receiving laboratories, including equipment (e.g., HPLC models), columns, environment, and analyst competence. Any identified gaps are addressed before transfer.
    2. Transfer Protocol Development:
      A detailed method transfer protocol is drafted, including:
      • Scope and responsibilities of both labs.
      • Acceptance criteria, typically covering parameters such as accuracy, precision, linearity, and specificity, aligned with ICH Q2.
      • Sample transport conditions to maintain integrity (e.g., temperature, humidity).
      • Training requirements for analysts at the receiving lab.
    3. Method Transfer Execution (Assay Example):
      For an HPLC assay of a tablet:
      • Samples (e.g., reference standards and tablet batches) are analyzed at the receiving lab.
      • Method performance is assessed against ICH Q2 parameters, including:
      • Accuracy: Using spiked placebo samples at 80%, 100%, and 120% of the label claim with 3 replicates per level. Evaluate % recovery, aiming for 98-102%, with %RSD ≤ 2%.
      • Precision:
      • Repeatability: Same analyst, same day, %RSD ≤ 2%.
      • Intermediate precision: Different analysts and days within the same lab, %RSD ≤ 3%.
      • Linearity: At least 5 concentration levels across the working range, calculate coefficient of determination (r² ≥ 0.99).
      • Specificity: Confirm separation of the analyte from excipients, impurities, or degradation products, with resolution > 1.5 between peaks.
    4. Data Analysis and Comparison:
      • Use statistical tools, such as t-tests or ANOVA, to compare results between the transferring and receiving labs, ensuring no significant difference.
      • All data are reviewed against the predefined acceptance criteria in the protocol.
    5. Deviation and CAPA Management:
      Any deviations during the method transfer (e.g., out-of-specification results or method performance failures) are investigated, root causes identified, and CAPAs implemented. Depending on the findings, the method transfer may be repeated.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Analytical method transfer to India

A
  1. Change Control:
    Initiate a formal change control to document and manage the method transfer, evaluating the impact on regulatory licenses, GMP certificates, and product quality.
    1. Gap Analysis:
      A senior analyst from the transferring laboratory conducts a gap analysis, comparing critical factors such as equipment, facilities, analytical techniques, and staff competence between the two laboratories.
    2. Impact Assessment:
      Assess the potential impact of the transfer on:
      * Marketing Authorizations (MA) if applicable,
      * GMP compliance and certification,
      * GCP requirements (for clinical products),
      * Finished product quality.
    3. Transfer Protocol Development:
      Develop a detailed protocol covering:
      * The method to be transferred,
      * Acceptance criteria,
      * Staff training at the receiving lab,
      * Sample transportation conditions,
      * Handling of deviations and CAPA during the transfer process.
    4. Validation Activities:
      Validation of the method at the receiving lab should cover key ICH Q2 elements, including:
      * Accuracy:
      * Assess recovery at 3 concentrations (e.g., 80%, 100%, 120%) with 3 replicates each.
      * Acceptance: Mean recovery within predefined limits (e.g., 98–102%) and %RSD ≤ 2%.
      * Linearity:
      * Demonstrate a linear relationship between concentration and response across 5 levels.
      * Acceptance: Coefficient of determination (r²) ≥ 0.99.
      * Precision:
      * Repeatability: Same analyst, same day, multiple replicates;
      * Intermediate Precision: Different analysts and/or days;
      * Reproducibility: If applicable, comparison across different laboratories.
      * Acceptance: %RSD typically ≤ 2–3%.
      * Specificity:
      * Confirm the method can separate the analyte from excipients, impurities, or degradation products.
      * Acceptance: Resolution between critical peaks ≥ 1.5.
      * LOD and LOQ:
      * Determine based on signal-to-noise ratio:
      * LOD ~3:1,
      * LOQ ~10:1.
      * Robustness:
      * Evaluate the method’s reliability under small deliberate variations (e.g., minor flow rate, temperature changes).
    5. Sample Transport:
      Define transport and storage conditions for sample transfer from the manufacturing site to both transferring and receiving labs, ensuring sample integrity.
    6. Comparative Study:
      Analyze the same batch samples at both labs and perform a paired t-test to statistically confirm there are no significant differences between results.
    7. Prospective (Traditional) Validation:
      Conduct a prospective validation at the receiving laboratory using at least 3 production batches, to demonstrate consistent method performance.
    8. Deviation Handling:
      Any deviations during the transfer must be investigated promptly, root causes identified, CAPAs implemented, and documented in the final transfer report.

Summary:
Although regulatory guidance (e.g., MHRA blog, Orange Guide) mainly emphasizes transfer process controls, it is considered best practice to include relevant ICH Q2 validation parameters such as accuracy, precision, linearity, specificity, LOD/LOQ, and robustness. This ensures that the method is scientifically sound, fit for purpose, and compliant at the receiving laboratory.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Where would you find guidance on how to validate chromatographic testing methods (HPLC)

A

To validate chromatographic testing methods (such as HPLC), the key references include:
* ICH Q2(R2) – “Validation of Analytical Procedures” (updated version, 2023),
* ICH Q14 – “Analytical Procedure Development” (for novel substances or new products),
* EU GMP Chapter 6 – “Quality Control,”
* EU GMP Annex 15 – “Qualification and Validation.”

These documents provide the regulatory expectations and technical requirements for method validation.

Of course! You’ve covered the key ideas really well — you just need a more polished and professional structure, correct technical English, and slightly more clarity (especially for viva or audit situations).

Here’s your revised, clean model answer:

Model Answer:

Q: Where would you find guidance on how to validate chromatographic methods such as HPLC?

To validate chromatographic testing methods (such as HPLC), the key references include:
* ICH Q2(R2) – “Validation of Analytical Procedures” (updated version, 2023),
* ICH Q14 – “Analytical Procedure Development” (for novel substances or new products),
* EU GMP Chapter 6 – “Quality Control,”
* EU GMP Annex 15 – “Qualification and Validation.”

These documents provide the regulatory expectations and technical requirements for method validation.

Key Requirements for HPLC Method Validation:

You must develop a validation protocol that includes:
* Clear definition of the analytical method,
* Predefined acceptance criteria for each parameter,
* Staff training requirements,
* Deviation handling, investigation, and CAPA in case of failures during validation.

During HPLC Method Validation, confirm the following elements (ICH Q2 parameters):
1. Accuracy:
* Measures how close the test results are to the true value.
* Perform at 3 different concentrations (e.g., 80%, 100%, 120% of target concentration) with 3 replicates each.
* Calculate the mean % recovery for each level.
* Acceptance: Mean % recovery between 98% and 102%.
2. Linearity:
* Demonstrates a linear relationship between analyte concentration and response (e.g., peak area).
* Prepare at least 5 concentration levels.
* Calculate the coefficient of determination (r²).
* Acceptance: r² ≥ 0.99.
3. Precision:
* Measures how close individual results are to each other.
* Repeatability: Same analyst, same equipment, same day.
* Intermediate Precision: Different analysts, different days, possibly different equipment.
* Reproducibility: (only if across different laboratories; less common for single-site validation).
* Acceptance: %RSD ≤ 2% (may allow ≤3% for more complex methods).
4. Specificity:
* Ability to clearly identify and quantify the analyte without interference from impurities, degradants, or excipients.
* Acceptance: Resolution between critical peaks ≥ 1.5.
5. Limit of Detection (LOD) and Limit of Quantitation (LOQ):
* Determined based on signal-to-noise ratio:
* LOD: S/N ≈ 3:1,
* LOQ: S/N ≈ 10:1.

Summary Viva-Style Sentence:

“I would refer to ICH Q2(R2), ICH Q14, EU GMP Chapter 6, and Annex 15.
For HPLC validation, I would develop a protocol covering method definition, acceptance criteria, staff training, deviation handling, and CAPA.
The key validation elements I would confirm are accuracy, precision, linearity, specificity, LOD, and LOQ, following ICH Q2 principles.”

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Can you describe the parameters that would be validated for an HPLC method? and what stats tool you’d use?

A

According to ICH Q2(R2), the key validation parameters for an HPLC method are:
1. Accuracy:
* Measures how close the test results are to the true value.
* Acceptance criteria: Mean % recovery between 98–102%.
* Statistical tool: Calculate mean % recovery across replicates.
2. Linearity:
* Demonstrates that the method response is directly proportional to analyte concentration across a defined range.
* Acceptance criteria: Coefficient of determination (r²) > 0.99.
* Statistical tool: Perform linear regression analysis and calculate r².
3. Precision:
* Measures the closeness of individual results under the same conditions.
* Includes:
* Repeatability (same analyst, same equipment, short time),
* Intermediate precision (different analysts, different days).
* Acceptance criteria: %RSD ≤ 2% for repeatability; may allow slightly higher (≤3%) for intermediate precision.
* Statistical tool: Calculate %RSD (Relative Standard Deviation).
4. Specificity:
* Confirms that the method can distinguish the analyte from impurities, degradation products, or matrix components.
* Acceptance criteria: Resolution between critical peaks ≥ 1.5.
* Statistical tool: Use system suitability parameters (e.g., USP resolution formula).
5. Limit of Detection (LOD):
* The lowest amount of analyte detectable, not necessarily quantifiable.
* Acceptance criteria: Signal-to-noise (S/N) ratio approximately 3:1.
* Statistical tool: Measure S/N ratio.
6. Limit of Quantitation (LOQ):
* The lowest amount of analyte that can be quantitatively determined with acceptable precision and accuracy.
* Acceptance criteria: Signal-to-noise (S/N) ratio approximately 10:1.
* Statistical tool: Measure S/N ratio.
7. Robustness:
* Assesses the reliability of the method when small deliberate variations are introduced (e.g., slight changes in flow rate, column temperature, mobile phase composition).
* Often evaluated by checking if %RSD remains within acceptable limits (typically ≤2%) when conditions vary.
* Statistical tool: Calculate %RSD and/or compare means across small method variations.

Short viva-style Summary Sentence:

“For HPLC method validation under ICH Q2, I would confirm accuracy, linearity, precision, specificity, LOD, LOQ, and robustness.
I would use statistical tools such as mean % recovery, coefficient of determination (r²), %RSD, signal-to-noise ratio, and resolution calculations to assess method performance.”

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

You are working as a QP at a site and have been asked to support your QC team in carrying out method transfer of a HPLC method. How would you go about this?

A

To support the method transfer of an HPLC method, I would follow a structured, compliant, and risk-based approach, guided by:
* ICH Q2(R2) – Validation of Analytical Procedures
* ICH Q14 – Analytical Procedure Development (if applicable)
* EU GMP Annex 15 – Qualification and Validation
* EU GMP Chapter 6 – Quality Control
* MHRA Orange Guide (2022) and MHRA Inspectorate blog

Step-by-step approach:
1. Initiate Change Control:
Formally raise change control to manage and document the method transfer, including impact and risk assessment.
2. Conduct Gap Analysis:
Performed by a senior analyst from the transferring lab, assessing:
* Equipment comparability (e.g. HPLC model, detector),
* Analytical methods and materials,
* SOPs, staff competence, and lab environment.
3. Impact Assessment:
* Assess whether the receiving lab is named in the Marketing Authorisation (MA); if not, a variation may be required.
* Evaluate GMP/GCP compliance, transport and handling of samples, and any potential impact on the finished product.
4. Develop Transfer Protocol:
The protocol must define:
* The method and scope of the transfer,
* Predefined acceptance criteria,
* Training requirements for analysts,
* How deviations will be handled (investigation + CAPA),
* Sample transport conditions,
* The comparative study design.
5. Conduct Comparative Testing:
* Analyze the same sample batch at both transferring and receiving labs.
* Perform statistical comparison using a paired t-test (or ANOVA if appropriate),
to ensure there is no significant difference in results.
* This forms a key part of confirming method equivalence.
6. Perform Validation Activities (ICH Q2):
Validate the method at the receiving lab for the following parameters:
* Accuracy:
Mean % recovery between 98–102% across 3 concentrations (e.g. 80%, 100%, 120%) in triplicate.
* Linearity:
Regression analysis over 5 concentration levels; r² ≥ 0.99.
* Precision:
Repeatability and intermediate precision; %RSD ≤ 2%.
* Specificity:
Resolution between analyte and closest peak ≥ 1.5.
* LOD / LOQ:
Based on signal-to-noise ratio: LOD ≈ 3:1, LOQ ≈ 10:1.
* Robustness:
Assess impact of small, deliberate changes (e.g., flow rate, temperature); ensure results remain within %RSD ≤ 2%.

Final statement (for viva impact):

“As a QP, I would ensure the method transfer is executed under change control, using a risk-based approach in line with ICH Q2 and Annex 15.
The comparative study, supported by a t-test, is a key step to confirm equivalence between labs.
My role is to ensure the receiving lab can perform the method reliably, maintaining regulatory compliance and product quality.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Any other types of method transfer other than comparative testing?

A
  1. Comparative Testing
    • The most common: both sending and receiving labs test the same samples and results are compared.

Other types include:
2. Re-Validation (Full or Partial)
* If the method cannot be reliably transferred or if there are changes (e.g., equipment is very different), you may need full or partial re-validation at the receiving site.
3. Transfer Waiver
* If justified (e.g., the same company group, same equipment, trained personnel, method already validated and proven equivalent), a formal transfer may be waived with proper documented risk assessment.
4. Co-Validation (Concurrent Validation)
* Sometimes the method is transferred and validated during routine sample testing — used mainly when the method is very robust or urgency is high. It’s riskier and needs strong justification.

You also mentioned Process Verification —
* Process verification usually refers more to manufacturing process (not analytical method transfer). It’s about verifying that a process consistently produces a product meeting its specifications — not transferring analytical methods.
* So, process verification would not typically be classified as a method transfer type.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Stability assay is OOS (one in spec one slightly out of spec); batches on the market; what do you do? Oncology drug everything else is in spec.

A

Initial Information Gathering:
0. Gather key information:
* What is the product, indication, and patient group (oncology patients are highly vulnerable)?
* Are there alternative treatments available on the market?
* What test showed OOS?
(E.g., assay slightly outside spec — critical for efficacy.)
* Are other tests (e.g., impurities, dissolution) within specification?
* Confirm stability trends over time.

Immediate Actions:
1. Quarantine the affected batch on-site immediately as a precaution, while investigation proceeds.
2. Initiate OOS Investigation:
* Phase 1a Investigation (Analyst Level):
* Check for clear assignable causes:
* Calculation errors, wrong SOP version, expired reagents, equipment malfunction.
* Phase 1b Investigation (Supervisor/Senior Analyst Level):
* Detailed checklist: Recheck calculations, equipment calibration/maintenance records, reagents and solutions.
* Hypothesis testing: If needed, reanalysis with predefined and QA-approved protocols to rule out hypothetical causes.
* If no assignable cause found:
Proceed to Phase 2 — product quality investigation (Deviation raised).

Deviation Management:
* Open a deviation record:
* Describe “who, what, when, where, and how” the OOS occurred.
* Perform impact assessment:
* Marketing Authorisation (MA):
* Batch is OOS — impacts authorised specifications.
* GMP:
* Possible process or equipment failure; systematic issue risk.
* Patient safety:
* Consult medical team:
* Is the small % deviation in assay clinically significant?
* What is the risk to the patient (underdosing, treatment failure)?
* Market risk:
* Oncology drugs often have no alternatives — risk/benefit must be weighed very carefully.

Additional Immediate Actions:
* Review stability data of other batches — trend analysis.
* Test retention/reference samples from released batches.
* Increase testing points to strengthen statistical reliability.
* Consider additional testing (e.g., impurities) depending on risk.

Recall Decision:
* Organize a recall risk assessment meeting with a multidisciplinary team:
* QA, QC, QP, Regulatory, Medical, Manufacturing, and Supply Chain.
* Review findings and decide:
* Recall class (Class I, II, III based on patient risk),
* Scope (all batches affected?),
* Level (wholesale, pharmacy, patient).
* Contact DMRC (Defective Medicines Report Centre, MHRA) early for advice and notify if recall is decided.

Root Cause Analysis (RCA):

Conduct thorough RCA covering:
* QMS: Deviations, CAPA records, change controls, OOS history, PQR data.
* Personnel: Training, analyst qualification.
* Equipment/Facility: Equipment calibration, maintenance (PPM) logs, validation status.
* Process: Process validation, any interventions during manufacturing?
* Materials: Supplier quality — any change/issues with starting materials or packaging?
* Documentation: Batch Manufacturing Records (BMR) for anomalies.
* Self-inspection findings: Any relevant audit observations.

CAPA and Follow-up:
* Open CAPAs based on the identified root cause:
* Prevent recurrence,
* Improve control measures,
* Conduct effectiveness checks.
* Finalize the investigation report.
* Submit a final report to the DMRC (including root cause, impact, recall outcome, and CAPAs).
* Update regulatory authorities as required if patient risk remains acceptable.

Closing Statement (for viva impact):

“In this situation, as a QP, my first priority is patient safety, followed by regulatory compliance and product quality assurance.
A thorough investigation, multidisciplinary risk assessment, and regulatory communication must be conducted, leading to appropriate CAPA to prevent recurrence.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

MHRA OOS guidance - tell me the process

A

MHRA OOS Investigation Process

Phase 1a: Immediate Laboratory Investigation
• The original analyst and their supervisor investigate if there is an obvious laboratory error.
• Examples: wrong calculation, wrong dilution, equipment power failure, spillage, sample prep error.
• If an obvious assignable cause is found, invalidate the original result, document fully, and repeat the analysis following a justified protocol.
• If no assignable cause is found, proceed to Phase 1b.

Phase 1b: Extended Laboratory Investigation
• A more detailed, structured investigation is conducted — usually by a senior analyst/supervisor and QA oversight.
• Use a pre-approved checklist to guide the investigation.
• Hypothesis testing may be conducted — for example, checking equipment, reagents, sample prep to rule out possible causes.
• A protocol must be prepared and QA approved before any re-testing or further investigation is carried out.
• Hypothesis testing is to support the investigation, not to justify re-testing to replace the original result unless clear assignable cause is found.

If assignable cause is still not found after Phase 1b:
• The result must be considered a true result unless proven otherwise.
• Initiate a Product Quality Investigation (e.g., Deviation or Non-Conformance) under the site’s Quality System.
• Assess batch impact (potentially batch rejection, further stability testing, or recall risk if batch already distributed).

Notes:
• Phase 2 (if needed): Full manufacturing and system-wide investigation if it points towards manufacturing issues.
• MHRA expects real-time QA oversight during OOS investigation.
• Root Cause Analysis (RCA) and CAPA must be documented.
• Trend analysis should also be considered (e.g., is it a one-off or part of a wider trend?).

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

dissolution failure

A

Step 1: Initiate OOS Investigation
* Conduct an OOS investigation in accordance with MHRA expectations and site OOS SOP.
* Immediately quarantine all affected batches (both released and unreleased) to prevent further distribution.

Phase 1a — Initial Laboratory Investigation
* The original analyst and supervisor conduct an immediate investigation.
* Check for obvious errors such as calculation mistakes, incorrect sample preparation, instrument malfunctions (e.g., power outage).
* QA approval is not required for Phase 1a.
* If an assignable cause is identified (e.g., proven calculation error), invalidate the result following documented procedure.
* If no assignable cause is found, proceed to Phase 1b.

Phase 1b — Extended Laboratory Investigation
* Senior analyst/supervisor conducts a detailed structured investigation, using a pre-defined checklist (e.g., reagent quality, equipment calibration, environmental conditions).
* Hypothesis testing may be performed to rule out probable causes.
* A written protocol must be approved by QA prior to conducting any hypothesis testing.
* If no assignable cause is found after Phase 1b, treat the OOS result as a true result and escalate.

Step 2: Open Deviation
* Raise a formal deviation or non-conformance.
* Document the situation clearly:
* Who found it, When, Where, How the deviation was identified.
* Cross-reference the deviation report to the OOS investigation.
* Identify all batches affected.

Step 3: Conduct Impact Assessment
* Marketing Authorisation (MA):
* Failure to meet dissolution specifications may breach the MA — review SmPC/registered specifications.
* GMP:
* Consider whether there may be a systemic failure (process, equipment, environmental control).
* Patient Safety:
* A dissolution failure may impact the drug’s bioavailability, leading to under-dosing or lack of efficacy.
* Classify the deviation as Major (or Critical if there is direct patient impact).

Step 4: Organise Recall Decision Meeting
* If batches have already been released:
* Convene a recall committee including QA, Regulatory Affairs, Medical Team.
* Decide on recall class and level (likely Class 2 or 3; patient-level recall if needed).
* Confirm classification and recall strategy internally before external action.

Step 5: Notify Authorities
* Contact the Defective Medicines Report Centre (DMRC) at the MHRA before starting the recall.
* Early notification is essential even if investigation is ongoing.

Step 6: Conduct Root Cause Analysis (RCA)
* Use an Ishikawa (Fishbone) Diagram to systematically investigate potential causes.
* Cover all key elements:
* Quality Management System
* Facilities and Equipment
* Personnel
* Documentation
* Manufacturing Process
* Quality Control
* Suppliers/Outsourcing
* Self-Inspection/Internal Audit Findings

Step 7: Implement Corrective and Preventive Actions (CAPA)
* Define clear CAPAs based on the root cause.
* Include effectiveness checks (e.g., trend monitoring, re-training, system upgrades).

Step 8: Complete Reporting
* Prepare a closing investigation report.
* Submit final reports and updates to the DMRC as required, including recall effectiveness checks.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

You are moving testing from a site in India to UK – how would you do this?

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

After completion of method transfer there is an OOS on the first batch tested in UK lab but Indian lab is showing passing result – what do you do?

A
  1. Initiate OOS Investigation (UK Receiving Lab)
    • Conduct a full OOS investigation following MHRA guidance and site SOP.
    • Determine whether the OOS result is due to an identifiable laboratory error (e.g., analyst error, equipment failure) or represents a true batch failure.
    • If no assignable cause is found, treat the OOS result as valid.
  2. Quarantine the Batch
    • Immediately place the affected batch under quarantine.
    • If any product has been released, perform a risk assessment and consider product recall if necessary.
  3. Investigate Method Transfer Execution
    • Review the method transfer protocol and acceptance criteria.
    • Compare critical method parameters and equipment between the Indian and UK labs.
    • Assess whether the method is truly robust and reproducible across sites — differences in instrumentation, analyst technique, or sample preparation may reveal the issue.
    • Confirm whether the sending lab results (India) were accurately and compliantly reported.
  4. If the UK OOS is Confirmed as a True Result:
    • Reject and dispose of the batch under QA oversight.
    • Document the rejection decision with cross-reference to the OOS investigation and batch records.
  5. Root Cause Analysis (RCA)
    • Conduct RCA to understand the discrepancy between sending and receiving lab results.
    • Use tools like the Ishikawa diagram to explore:
    • Method variability,
    • Analyst competency,
    • Equipment calibration and sensitivity,
    • Environmental factors,
    • Sample handling and preparation.
  6. Corrective and Preventive Actions (CAPA)
    • Raise a deviation in the QMS.
    • Implement appropriate CAPAs, such as:
    • Re-assessment or re-validation of the analytical method,
    • Analyst re-training,
    • Instrument qualification review,
    • Enhanced oversight during early-phase testing after transfer.
  7. CAPA Effectiveness Check
    • Define measurable success criteria (e.g., consistent passing results on future batches, stability of system suitability parameters).
    • Monitor trends in analytical data from subsequent batches.
  8. Regulatory and Quality Oversight
    • If product was released or distributed, assess need to notify MHRA via DMRC.
    • Document investigation, RCA, CAPA, and final decision in a comprehensive investigation report.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Give examples of what you would look at in Phase 1a/b?

A

Phase 1a – Immediate Laboratory Checks (No QA Approval Needed)

Conducted by the original analyst and supervisor to identify any obvious errors:
* Calculation errors (e.g., incorrect formula or unit conversion)
* Sample mix-up or mislabelling
* Equipment failure (e.g., power outage, instrument crash during analysis)
* Incorrect reagent or diluent used
* Incorrect sample preparation technique (e.g., pipetting error, dilution mistake)
* Missed steps in the test method or SOP deviation
* Obvious analyst error (e.g., misreading a balance or spectrophotometer)

Phase 1b – Extended Laboratory Investigation (QA Approval Required Before Hypothesis Testing)

Conducted by senior analyst/supervisor with QA oversight, using a structured checklist:
* Out-of-date reagents or reference standards
* Equipment calibration out of date or instrument drift
* Preventive maintenance (PM) overdue (e.g., HPLC pump not serviced)
* Method ruggedness issues (e.g., sensitivity to small changes in temperature or pH)
* Analyst variability — review of training and performance records
* Review of chromatography system suitability results or baseline noise
* Environmental conditions during analysis (e.g., humidity, temperature excursions)
* Software/system access issues or audit trail anomalies
* Hypothesis testing (e.g., repeat test with fresh reagents, different column or analyst — if pre-approved by QA)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Turns out it’s a true OOS – can you release the batch?

A

No — if the OOS result from the transferring lab is confirmed as a true result (i.e., not due to analytical error), the batch must not be released.

Why:
* Under EU GMP Annex 16, and MHRA expectations, a Qualified Person must ensure that each batch complies with the specifications described in the Marketing Authorisation or clinical trial documentation before certification.
* If a confirmed OOS result exists (either during in-process or finished product testing), it means the batch does not comply with its specification.
* This constitutes a quality defect, and the batch must be rejected unless a scientifically justified, regulatory-approved action is taken (e.g., reprocessing with requalification — not retesting to invalidate a true result).

Even if another lab (e.g., UK receiving lab) obtains a passing result, the confirmed true OOS result from the transferring lab cannot be disregarded.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

That batch was scrapped and we released another batch, 3 months later there is an OOT for an impurity – what are your thoughts?

A

Immediate Actions:
1. Quarantine any affected batches:
* Immediately quarantine any batch currently awaiting release.
* Identify all batches already released to market that may be impacted by this impurity OOT.
2. Review Other Analytical Results:
* Check the full analytical profile (e.g., assay, dissolution, other impurities) to ensure no other abnormalities.
3. Initiate OOT Investigation:
* Perform a structured OOT investigation according to site SOP.
* Determine if the OOT is:
* A genuine trend (e.g., slow increase in impurity)
* Or due to analytical/laboratory variation (e.g., instrument drift, analyst error).
4. If OOT is Confirmed as True:
* Use statistical analysis:
* Extrapolate impurity levels over time.
* Apply linear regression or 95% confidence interval to predict when the batch would breach specification (OOS).
* Assess remaining shelf-life viability.

Deviation Management and Impact Assessment:
* Deviation:
* Open a formal deviation in the QMS system linked to the OOT finding.
* Impact Assessment:
* MA Compliance:
* Although currently within specification, projected future OOS is a potential MA breach.
* GMP Compliance:
* Potential systemic issues to be investigated — manufacturing process, contamination risks, API quality, environmental controls.
* Patient Safety:
* Evaluate whether the impurity is toxic or otherwise harmful. Check toxicology data (e.g., ICH M7 if genotoxicity).
* Market Impact:
* Assess whether alternative products are available.
* Patient Impact:
* Identify patient groups affected — vulnerable groups (e.g., paediatrics, oncology) need greater caution.

Recall Decision:
* If extrapolation shows the batch will soon exceed specification, even if currently compliant:
* Organise a recall meeting with QA, Regulatory Affairs, Medical, and Management teams.
* Likely Class 3 recall (precautionary)
* Level: pharmacy/hospital recall (as product is still within specification at the time of action).
* Notify the Defective Medicines Report Centre (DMRC) before any public action if needed.

Root Cause Analysis (RCA):

Use an Ishikawa Diagram to systematically investigate:
Category
Areas to Investigate
QMS
Deviation logs, PQR, Change Control history
Facilities/Equipment
Temperature excursions, environmental trends, maintenance logs, cleaning records
Process
Cleaning validation, line clearance failures, manufacturing records (BMRs)
Quality Control
Reference standard stability, lab equipment performance, audit trails
Personnel
Analyst and operator training records
Supplier Management
API quality (CoA, retesting, reprocessing, audits)
Self-Inspection
Any relevant audit findings linked to manufacturing, storage, or testing
Corrective and Preventive Actions (CAPA):
* Implement CAPAs based on root cause findings:
* Strengthen supplier qualification.
* Improve cleaning processes.
* Revise method validation if necessary (e.g., impurity stability).
* Additional stability testing or reduced shelf-life if appropriate.
* Conduct effectiveness checks:
* Monitor future batches for similar trends.
* Additional sampling/testing if needed.

Regulatory Communication:
* Submit a closing investigation report to MHRA DMRC, especially if recall or public health risk is involved.
* Update stability protocols if necessary.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Describe the recall procedure .

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

You have a stability batch and there was an oos at 18 months for Tablet, shelf life was 24 months. Cancer drug.

A

Why will you not reject this batch when I said I will not, then justify – cancer drug

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

A routine stability batch OOT for assay at 18 months for another product, which drilled more detailed about the concern with MDI.

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

What confidence interval, its calculation and why do you keep 95% confidence interval?

A

A confidence interval is a statistical range that is likely to contain the true population parameter (e.g. mean, slope, intercept) with a specified level of confidence — most commonly 95%.

In simple terms:

A 95% confidence interval means: “We are 95% confident that the true value lies within this range.”

It’s not that 95% of the data lies in this range — it’s that if the experiment were repeated many times, 95% of those intervals would include the true value.

How is it calculated?

For a mean, the confidence interval is calculated using:

CI = x̄ ± (t × s/√n)

Where:
• x̄ = sample mean
• t = t-statistic (from Student’s t-distribution for the chosen confidence level and degrees of freedom)
• s = sample standard deviation
• n = number of observations

So, the CI widens when:
• Your data is more variable (larger standard deviation),
• You have fewer samples (smaller n),
• You choose a higher confidence level (e.g., 99%).

Why use 95% confidence interval?
• 95% is a regulatory standard accepted by authorities like ICH, EMA, MHRA, FDA, etc.
• It provides a good balance between certainty and practicality — high enough confidence without requiring an unmanageably large sample size.
• Specifically, in ICH Q1E, confidence intervals are used when performing regression analysis for stability data, especially to justify shelf life assignments.

Example: A 95% one-sided confidence limit on the regression line is used to ensure that the assay remains above the lower specification limit throughout the proposed shelf life.

Viva-ready sentence:

“A confidence interval defines the range within which the true population parameter is likely to fall.
I use the 95% confidence interval because it is the regulatory standard per ICH Q1E for stability data and provides sufficient assurance that the drug product will remain within specification throughout shelf life.”

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

How would you set up a stability programme for a tablet product?

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

Where would you find guidance bracketing and matrixing ? Can you please explain what do they mean?

A

Where to find guidance?

You can find formal guidance in ICH Q1D – Bracketing and Matrixing Designs for Stability Testing of New Drug Substances and Products.

What do bracketing and matrixing mean?

Both bracketing and matrixing are statistical design approaches used in stability studies to reduce the total number of samples tested, while maintaining confidence in the data and regulatory compliance. They are especially useful when multiple strengths, pack sizes, or container types are involved.

Bracketing:
* Definition: A design in which only the extremes (i.e., highest and lowest strengths, or largest and smallest pack sizes) are tested, assuming the intermediate strengths or configurations will behave similarly.
* When used: When all other variables (e.g., formulation, manufacturing process, container closure system) are the same across the range.
* Example: For a product with 50 mg, 100 mg, and 200 mg strengths — only 50 mg and 200 mg are placed on stability, skipping 100 mg.

Matrixing:
* Definition: A design where a subset of the total number of stability samples is tested at each time point, with each batch/strength/container tested at different intervals.
* Purpose: Reduces the number of tests over time without compromising the ability to detect degradation trends.
* Example: For 3 batches × 3 strengths × 3 pack types = 27 combinations, matrixing allows you to test only a portion of these at each time point (e.g., 12 out of 27), rotating across time points.

Viva-Style Summary Sentence:

“Bracketing and matrixing are statistical approaches described in ICH Q1D, used to reduce the number of stability tests while maintaining confidence in shelf-life estimation. Bracketing involves testing only the extremes of strength or pack size, and matrixing involves testing only a subset of samples at each time point, based on a defined design.”

21
Q

You receive a call on Friday evening from QC informing about OOT for tablet assay testing at 9 months. Shelf life is 36months. What are your concerns?

22
Q

You have now been told that batch will fail at 12 months. What do you do?

23
Q

Which type of stats would you use to estimate shelf life?

A

I drew a graph to explain this and the assessor was happy.

24
Q

Are you allowed to perform extrapolation?

A

Shelf-life extrapolation is only allowed when stability trends are clear and statistically supported. ICH Q1E requires regression analysis (95% CI), a minimum of 3 batches, and testing of batch poolability  . The extension beyond the last data point is limited (typically ≤ +6 m with stats, or +3 m without)  . Accelerated/intermediate results must show stability to permit extrapolation  . EMA/MHRA guidance (including Q1A, Q1E and clinical trial guidelines) echoes these rules   . All extrapolations must include a commitment to obtain longer-term data to verify the shelf life.

25
How would you perform a method transfer of Assay method to another lab? What stats would you use?
26
What are the difference between biological and chemical tests methods?
27
Please explain what ELISA is?
ELISA stands for Enzyme-Linked Immunosorbent Assay. It is a quantitative or semi-quantitative analytical method used to detect and measure biological substances such as hormones, proteins, peptides, and antibodies, based on antigen–antibody interactions. ⸻ How it works: 1. A microplate (usually 96-well) is coated with a capture antibody or antigen specific to the target analyte. 2. The sample is added — if the target analyte (e.g., hormone, protein) is present, it binds to the coated antibody/antigen. 3. A secondary antibody, linked to an enzyme, is added and binds to the analyte. 4. A chromogenic substrate is introduced — the enzyme catalyses a color change. 5. The intensity of the signal (measured as absorbance) correlates with the amount of analyte present in the sample. ⸻ Types of ELISA: • Direct ELISA • Indirect ELISA • Sandwich ELISA (commonly used for large molecules like proteins or cytokines) • Competitive ELISA ⸻ Viva-style summary sentence: “ELISA is an immunoassay technique used to quantitatively or semi-quantitatively measure biological molecules like hormones or proteins. It uses antigen-antibody binding followed by an enzyme reaction to produce a measurable signal.”
28
What statistical methods would you use for chemical and biological analysis with examples?
29
HPLC for unknown impurity at 0.1%, OOS at 18 months stability, what do you do? Shelf life is 24 months
Step 0: Gather Context * What is the product, its indication, and patient group (e.g., oncology, paediatrics)? * Are there market alternatives? * What other stability parameters (e.g., assay, known impurities, dissolution) look like? * Review prior stability trend data for the impurity (e.g., was it increasing gradually?). * Confirm whether the impurity was present at time zero (manufacturing issue) or only appears during storage. ⸻ Step 1: OOS Investigation (per MHRA OOS Guidance) * Phase 1a: Analyst-level checks — method, system suitability, sample prep, calculations, equipment calibration, reagent integrity. * Phase 1b: Supervisor-led review using an approved checklist — reprocessing using the same sample (not re-sampling), cross-checking method parameters, etc. * If no assignable cause is found, proceed to: * Hypothesis testing (e.g., using a second column or mobile phase) — protocol must be QA-approved in advance. * If still unresolved: raise a formal deviation — possible product quality issue. ⸻ Step 2: Immediate Containment and Risk Review * Quarantine affected batch(es) in stability chambers and market stock (if still in distribution). * Investigate the identity and structure of the unknown impurity: * Use MS/NMR techniques to identify. * Check if a Permitted Daily Exposure (PDE) is available or derivable. * Review toxicological data (consult QPPV and medical team). * Check: * Reference/retention samples for comparison. * Previous stability batches for similar impurity trends. * Customer complaints / signal detection systems for any related findings. * Environmental monitoring trends (if contamination suspected). ⸻ Step 3: Impact Assessment * Marketing Authorisation (MA): * This is an OOS for an unknown impurity, which may breach ICH Q3A(R2) limits — reportable to the MHRA and possibly to other global regulators. * GMP Impact: * Could indicate process control failure, cleaning validation failure, or contamination. * Patient Risk: * Unknown impurity = unknown toxicity. * Precautionary principle applies — involve QPPV and medical to assess benefit-risk. * Finished Product Risk: * Identify potentially affected batches (e.g., same lot, line, equipment, raw material). ⸻ Step 4: Recall Decision * Convene a recall risk assessment meeting with multidisciplinary team: QA, QP, QC, Medical, Regulatory, Supply Chain, and QPPV. * Based on outcome: * Recall class: likely Class III (no immediate safety risk, but quality defect). * Recall level: depends on distribution (pharmacy, hospital, wholesaler). * Notify MHRA DMRC before executing recall. * Consider DASH portal notification to DHSC if shortage expected. ⸻ Step 5: Root Cause Analysis (RCA) Use a structured method such as an Ishikawa diagram to evaluate: * QMS: prior deviations, change controls, CAPAs, PQR trends * Facility/Equipment: cleaning validation, qualification, maintenance logs, line clearance, temperature excursions * Documentation: batch records, cleaning records, analytical raw data * Personnel: training and supervision * Process: contamination control, hold times, process validation * Suppliers: COA review, supplier change, transportation deviations * Storage/Distribution: temperature/microbiological excursions * Audit Findings: any previous GMP findings relevant to the impurity? ⸻ Step 6: Corrective and Preventive Actions (CAPA) * Based on root cause, implement: * Supplier audits * Strengthened incoming goods testing * Enhanced cleaning validation or EM monitoring * Method revalidation (if analytical issue) * CAPAs must be tracked, effectiveness checked, and documented. ⸻ Step 7: Regulatory Reporting * Submit a closing investigation report to the MHRA DMRC, including: * OOS details * Root cause * Patient risk assessment * Recall outcome * CAPAs taken ⸻ Final Viva-Style Summary: “For an unknown impurity OOS at 18-month stability, I would investigate per MHRA OOS guidance, involve the QPPV and medical team, assess patient and MA impact, and convene a recall risk meeting. I would identify the impurity, evaluate risk, and report to MHRA. Root cause analysis would drive CAPAs, with effectiveness checks and regulatory follow-up.”
30
How will you calculate potency in a cell based product, how will you go about to test robustness in ELSIA method and what stats will you use in the case you want to test the following: *Different temperatures *Different incubation times *Different dyes
Potency in a Cell-Based Product: Potency is defined as the biological activity of the product and is a critical quality attribute (CQA), especially for cell-based therapies such as CAR-T products. It should reflect the intended mechanism of action (MoA). In the case of a CAR-T product, a common potency assay would be a cytotoxicity assay, where the ability of CAR-T cells to lyse target cancer cells (e.g., CD19+ lymphoma cells) is measured. One approach is to calculate potency by comparing the activity of the test sample against a qualified reference standard, expressed as a relative potency ratio. For example, you may assess the percentage of cancer cell lysis (using a dye release or flow cytometry-based assay) and report the ratio of the test product’s activity relative to that of a reference CAR-T batch. This can be further expressed using dose-response curves and EC50 values. ⸻ Robustness Testing in ELISA Method: Robustness is part of method validation under ICH Q2 and reflects the method’s reliability under a variety of deliberate variations. For ELISA: • Identify critical method parameters (e.g., incubation temperature, time, reagent concentration, washing steps). • Deliberately vary them within a realistic range (e.g., ±5°C, ±10% incubation time, alternate dyes or detection systems). • Assess the impact on the assay result (e.g., optical density or calculated concentration). • A common acceptance criterion is that the %RSD of replicate measurements should be ≤2% (or a defined acceptance range based on method variability and clinical relevance). ⸻ Statistical Tools for Robustness Evaluation: To evaluate the effect of each variable, the following statistical tools can be used: • Different Temperatures: • Use %RSD to assess repeatability within each temperature group. • Apply t-tests (if comparing two temperatures) or ANOVA (if more than two) to determine if there is a statistically significant difference in mean results. • Different Incubation Times: • Use %RSD for internal precision. • Apply ANOVA to assess whether varying incubation time has a statistically significant impact on assay performance. • Different Dyes (e.g., detection labels in ELISA): • Use t-tests to compare performance metrics (e.g., OD values or calculated concentrations) between dye conditions, assuming data is normally distributed. • Alternatively, use non-parametric tests (e.g., Mann-Whitney U) if normality cannot be assumed.
31
Difference between biological and chemical methods
Differences Between Chemical and Biological Methods: 1. Molecular Size and Structure: * Chemical products are typically small molecules with well-defined, simple structures. * Biological products (biologics) are large, complex molecules such as proteins, enzymes, monoclonal antibodies, or hormones. 2. Method of Production: * Chemical products are manufactured via controlled chemical synthesis. * Biologics are produced using living cells or organisms through biological processes, e.g., antigen–antibody interactions, cell cultures, or recombinant DNA technology. 3. Stability and Variability: * Chemical molecules tend to be relatively stable and consistent between batches. * Biologics are inherently more variable and sensitive to changes in manufacturing conditions. Therefore, tighter in-process controls are essential (e.g., precise control of bioreactor parameters such as temperature, pH, oxygen, and CO₂ levels). 4. Sterilisation and Process Design: * Chemical products can often undergo terminal sterilisation (e.g., heat or radiation). * Biologics are usually produced in closed systems to prevent contamination, as many cannot tolerate terminal sterilisation—aseptic processing is typically required.
32
a beta blocker product stability batch fail at 18M: *What is the OOS/OOT procedure?
Immediate Actions and Data Review First, recognise that an 18‑month stability OOS (out‑of‑specification) result on a marketed batch immediately triggers an investigation per MHRA policy  . Immediately ask: What was the actual assay value versus spec? Review previous stability results (e.g. at 6, 12 months) for trends or Out‑of‑Trend (OOT) signals. Examine the analytical data and method: confirm the correct, latest method version was used; check chromatograms, standard/reference identities, calculations and equipment logbooks. Verify sample chain of custody and storage conditions. Evaluate possible patient impact: e.g. if Atenolol content is low, patients may be undertreated. In summary, collect all relevant data quickly (prior time‑points, method validation, recent equipment checks, distribution history) to scope the problem  . Quarantine and Containment All remaining stock of the affected batch – both stability samples and manufactured product – must be held under quarantine immediately to prevent use. Label the batch as “under investigation/OOS” and stop any further distribution or dispensing. Retain all unused stability samples (and any reference standards or reagents used) for re-testing. If any product has been released or shipped, inventory it and notify distributors or warehouse to hold supply pending investigation. Essentially, treat the batch as nonconforming until the investigation is complete. This aligns with GMP requirements to “not delay” handling of a defective product . (No references needed for standard holding procedure.) OOS/OOT Investigation (MHRA Phases I–III) MHRA guidance prescribes a stepwise investigation of OOS/OOT results . The investigation is conducted in phases: * Phase 1A – Analyst check (Immediate): The analyst who ran the assay first reviews raw data for obvious errors. Check calculations, weighing and dilution logs, instrument settings, power failures, column issues, leaks, sample mix‑ups or spills . Verify the correct sample was tested and recorded. If a simple lab error is found (e.g. calculation mistake, instrument fault, expired standard), correct it and re-calculate; if this resolves the OOS, document and close Phase I (no formal report needed beyond lab records). The MHRA notes that clear causes like technician error or equipment failures can invalidate a result . * Phase 1B – Supervisor check (Checklist review): If no obvious error was found, the QA/manager or supervisor conducts a joint checklist review with the analyst . Using the MHRA’s investigational checklist, verify key points: correct method (version, solvent), instrument calibration and system suitability, sample identity and integrity (chain of custody, storage), reagent/standard quality, analyst training and any deviations. Check that the retention/sample storage conditions (stability chamber temperature) were maintained. Restrict this initial review to data/equipment records only. Once the checklist is complete and documented, proceed to re‑measurement: perform a repeat assay on the original sample solution or additional aliquot as per SOP . Any retest should follow the documented hypothesis plan. If the repeat is in spec, evaluate whether the original was an outlier or whether to invalidate it based on evidence. If re-test confirms OOS, or if doubts remain, move to Phase II. * Phase 2 – Full Investigation: If Phase I finds no clear cause, launch a formal, hypothesis‑driven investigation under QA oversight . First, execute a manufacturing investigation: review batch records for process deviations, raw material certificates (API assay, impurity limits), equipment logs (granulator, tablet press), environmental controls, etc. Simultaneously, plan a laboratory investigation: write and QA‑approve a protocol detailing the hypothesis (e.g. degradation, sampling error, method issue) and testing plan . This plan must document exactly which samples/standards to retest, how (e.g. new extractions, alternate method), and data evaluation criteria . Perform additional analyses to confirm or eliminate each hypothesis: for example, retest another portion of the retained stability sample, prepare fresh solution from the sample, check alternate assay methods or stability‑indicating procedures, and test working standards. Also consider testing retention samples from other time points or other batches. MHRA emphasizes multiple hypotheses (e.g. equipment, method, sample prep) must be tested systematically . Any corrective retesting must not compromise the integrity of the remaining reference samples. * Phase 3 – Conclusion and CAPA: If, after investigation, the result remains outside spec with no assignable lab error, then the batch is formally rejected (batch disposition = “fail”), and broader actions follow . The Phase III investigation reviews all findings: it evaluates whether other batches or ongoing stability studies might be affected and determines product disposition  . MHRA guidance states “once a batch has been rejected there is no limit to further testing… and the decision to reject cannot be reversed” . QA/Quality Unit should document all testing done and evaluate process implications. The Qualified Person (QP) must conclude whether the batch can ever be released (usually not, if one assay failed) . Crucially, CAPAs (Corrective/Preventive Actions) are identified and implemented: for example, process corrections, analytical method fixes, equipment maintenance or training. All conclusions, including impact on other products and CAPAs, are documented in the final OOS report . Regulatory Reporting and Recall Because this is a stability fail on a marketed product, it is a “suspected defect” that likely requires immediate notification to the MHRA (Defective Medicines Report Centre) and possibly a recall. Under GMP Chapter 8.15 and MHRA DMRC guidance, any defect that could lead to patient harm must be reported without delay . A stability OOS should not be held until the next timepoint – report it promptly . The licence holder should submit a defect report via the MHRA’s Yellow Card system or direct DMRC contact, including a description of the defect, batch/product details (PL number, batch number, expiry, distribution dates, quantities on hand), and an initial risk assessment  . The report should state whether a recall is recommended. If product has been distributed, initiate a recall plan in consultation with QA and DMRC. Classify the recall level per MHRA guidelines: here, an Atenolol assay failure (likely lower potency) could reduce efficacy (risk of mistreatment but not immediate life‑threat). This scenario typically falls under a Class 2 Medicines Recall  (“defect may cause harm but is not life-threatening”). MHRA’s table clarifies Class 2 is for “defect may cause harm; not life-threatening” . A Class 1 (NatPSA) patient‑alert would be for severe risk. The licence holder decides recall scope (patient, pharmacy, wholesale) based on risk and stock distribution. Notify the DMRC promptly with details of the recall actions and affected stock. If a MHRA-wide Medicines Recall Notification is warranted (widespread supply or serious risk), the agency may issue an alert . Throughout, coordinate with pharmacovigilance (for any reported lack of efficacy) and follow MHRA’s recall guidance. Root Cause Analysis (Fishbone Diagram) Use a systematic root-cause approach (e.g. Ishikawa “fishbone” diagram) covering all potential factors. On the diagram, consider categories such as Equipment (HPLC calibration, stability chamber sensors, blending/compression machines), Materials (API potency, impurity content, excipient quality, packaging), Methods (analytical method validation, stability‑indicating power, sample preparation, filter use), Manpower (analyst training, oversight, deviations in procedure), Environment/Facilities (laboratory conditions, stability chamber temperature/humidity control, transport/storage conditions), Process Controls (manufacturing process parameters, in-process sampling, batch uniformity), and Documentation (errors in recording, SOPs, label mix-ups). Brainstorm all plausible causes: for example, a decline in Atenolol content could be due to degradation (e.g. moisture or light effects), a formulation inconsistency (blending or coating issue), analytical underestimation (expired reference standard or incomplete extraction), or simple operator error . For each hypothesis, gather evidence: review equipment maintenance logs, re-test calibration standards, analyze retained API and excipients, and inspect stability chamber logs. MHRA guidance encourages testing hypotheses one by one (e.g. retesting with fresh standards, verifying sample prep, re-analyzing in a different lab) to confirm or eliminate each potential cause . Common root‑causes to explicitly consider (and have been noted by MHRA) include technician/calculation errors, sample or standard prep mistakes, analytical method issues, equipment failures, or procedural deviations . CAPA Development and Effectiveness Checks Based on the root‐cause analysis, implement targeted CAPAs. For example, if an unstable assay procedure or calibration issue was found, revise and revalidate the method or replace faulty equipment. If manufacturing issues (e.g. tablet homogeneity or moisture) are implicated, adjust process parameters or supplier specs. If human error (training or oversight) was a factor, retrain staff or strengthen SOPs. Document each CAPA with a clear action plan, responsible person, and timeline. Include preventive measures (e.g. more frequent stability chambers calibration, updated sampling SOPs) to avoid recurrence. Plan effectiveness checks: for instance, run extra stability points, audit the process, or perform trending on subsequent batches to verify that the issue is resolved. Update the Quality Management System records with the CAPAs and monitor their implementation. MHRA guidance stresses that corrective and preventive actions be recorded alongside the investigation conclusion . Investigation Closure and Reporting to DMRC Once all testing, root-cause analysis and CAPAs are complete, finalize the investigation. Prepare a formal investigation report summarizing findings, justification for the OOS, root cause(s) identified, actions taken, and the final disposition of the batch (usually rejection and destruction if the OOS holds). The QP (or QA head) reviews and signs off. If a recall was executed, include the recall/reconciliation report. Submit the final report to the DMRC, as MHRA requires the full conclusion of investigations into defects . Maintain all records in the batch file and GMP quality systems. Finally, confirm closure by verifying that CAPAs are completed and effective (e.g. no recurrence of similar OOS/OOT on future batches). References: MHRA’s current OOS/OOT guidance provides a detailed phased flowchart  ; MHRA inspectorate and DMRC communications emphasize timely reporting and root‑cause rigor   . Adhering to these steps (including full documentation and notifying DMRC) ensures compliance and patient safety.
33
Explain HPLC analysis and instrumentation?
34
What is regression analysis
Regression analysis is used to determine whether a dataset shows a linear relationship between variables. This is often assessed by calculating the coefficient of determination (R² value). An R² value greater than 0.99 indicates a very strong linear correlation. This method is commonly used in pharmaceutical stability studies to demonstrate that assay results, degradation products, or potency trends remain consistent over time.
35
Tell us Recall procedure and Types of recall?
1. Initiate Recall and Convene Recall Meeting A recall is initiated upon detection of a serious product defect. A multidisciplinary recall meeting is promptly organised, involving representatives from Pharmacovigilance (PV), Medical, Quality Assurance (QA), Manufacturing, and Regulatory Affairs. 2. Conduct Risk and Impact Assessment The team reviews the nature of the defect, assesses patient safety risks, supply chain impact, and potential regulatory obligations. Based on this, the recall class (I, II, or III) and level (wholesale, retail, patient) are determined in line with MHRA and the Guidelines for the Notification and Recall of Defective Medicines. 3. Notify the MHRA/DMRC The Defective Medicines Report Centre (DMRC) at the MHRA is contacted promptly. Formal notification is submitted using the Yellow Card system and the Defective Product Reporting Form. The recall must not proceed until the DMRC agrees on the recall level and communication content. 4. Execute the Recall Once approved, the recall is executed. Communications are sent to affected stakeholders with clear instructions for return, segregation, and quarantine of defective stock. Distribution records are reviewed to ensure traceability and completeness. 5. Reconciliation and Disposal All returned products are reconciled against distribution data. Any affected stock held on-site is segregated and clearly labelled. Defective stock is securely quarantined and disposed of in accordance with SOPs and waste regulations. 6. Root Cause, CAPA and Effectiveness Check A formal root cause investigation is conducted. Corrective and Preventive Actions (CAPAs) are raised for any process gaps identified. Effectiveness checks are performed to ensure CAPAs prevent recurrence. 7. Closure and Final Reporting A final recall report is prepared and submitted to the DMRC, including reconciliation data, investigation outcomes, CAPAs, and effectiveness verification. The recall is only closed when DMRC formally accepts the outcome.
36
Your lab informs you of a related substance peak at 12M on a stability trial that is unexpected. What are you going to do?
1. Gather Initial Information: * Ask: What is the related substance? Confirm if it’s a known degradant, process impurity, or new/unknown peak. * Determine: Is the result OOS (Out of Specification) or OOT (Out of Trend)? * Confirm: What is the product, indication, dosage form, and patient population? This informs the risk level (e.g., oncology, paediatric). * Check if the value is still within the ICH specification or justified limits. 2. If OOS or OOT, Initiate Formal Investigation: * Follow MHRA Guidance for OOS Investigations (aligned with the FDA 2006 OOS Guide). * Conduct Phase 1a/1b review: analytical error, instrument calibration, sample prep, standard, etc. * If no assignable cause is found, escalate to Phase 2: cross-functional team investigation (QA, QC, production).
37
Where is the guidance for OOS investigations? How would you conduct one?
GOV.UK - OOS guidance
38
OOT at 6 months for impurity for Solid Oral Dose
1. Initiate OOT Investigation: * Open a formal OOT investigation. * Conduct Phase 1a (by analyst): Check for analytical errors (e.g., sample prep, instrument calibration, calculation). * If inconclusive, escalate to Phase 1b (with supervisor/QA): Review method validity, equipment status, reagent quality, analyst training, etc. * If found to be a true OOT, proceed to Phase 2 – full investigation. 2. Evaluate Trend and Predict Shelf-Life Impact: * Review historical stability data. * Apply statistical trend analysis (e.g., regression) and calculate 95% confidence intervals to estimate when the impurity will breach specification. * Consider increasing stability test frequency (e.g., monthly) to monitor further progression. 3. Product Risk Assessment and Regulatory Impact: * Assess clinical risk based on impurity identity, ICH Q3B thresholds, and indication (e.g., chronic use, paediatrics). * If the product may breach its specification before expiry, initiate recall discussions (Class and Level) and contact MHRA/DMRC as needed. * Ensure QA oversight and consult with regulatory team on required notifications. 4. Deviation and Cross-Batch Impact Assessment: * Open a deviation and perform impact assessment: * Review all batches using the same API batch, packaging component, or manufacturing conditions. * Check reference samples and other stability batches for similar trends. * Review recent manufacturing or storage changes. 5. Root Cause Analysis and CAPA: * Conduct RCA: Focus on API impurity profile, vendor change, environmental storage, or formulation robustness. * Perform supplier audit or data review if linked to raw materials. * Implement CAPA, including long-term preventive measures. * Document and perform effectiveness check (e.g. future stability batches, audit closure).
39
Phase 1a/1b investigation – asked for specifics of what would be required
1. Phase 1a – Performed by Analyst (Within 1 Working Day): The aim is to identify any obvious, assignable errors. Actions typically include: * Re-checking calculations and data transcription. * Confirming correct sample, method, and dilution were used. * Verifying equipment was calibrated, powered, and functioning. * Checking for sample preparation errors or analyst deviations from SOP. * Ensuring correct reference standards and reagents were used and in-date. This phase should be documented contemporaneously. If an assignable cause is found, and it is scientifically justified, the test may be repeated with QA approval. 2. Phase 1b – Performed with Supervisor/QA and Checklist: If Phase 1a is inconclusive, a supervised review using a structured checklist is carried out. This includes: * Reviewing equipment qualification and maintenance/calibration status. * Checking method validation/transfer suitability for the matrix and product. * Reviewing training and competency of the analyst. * Confirming consumables’ quality and expiry date. * Reviewing environmental conditions and logbooks. * Ensuring no mix-up of samples, standards, or reagents. If no assignable cause is found after 1b, the issue may be escalated to Phase 2 (full investigation), which involves QA, production, and possibly a cross-functional team.
40
Phase 2 – no issues at DP manufacturing site API site had reworked the batch What documentation would you expect for rework? What is the difference between reprocessing and rework? Considerations for market action
1. Documentation Expected for Reworked API Batch: The QP must ensure the following documentation is available and reviewed prior to certification of the finished product: * Rework justification (with scientific rationale and risk assessment). * Deviation record (approved and QA-reviewed). * Approved rework protocol (including batch-specific procedures and additional testing). * Analytical data showing that the reworked API meets all registered specifications. * Impurity profile comparison (pre-/post-rework) to assess against ICH Q3A thresholds. * Stability data or bracketing/justification if stability is affected. * QP confirmation that the rework is in line with the Marketing Authorisation (MA) or supported by a post-approval change. * Reference to the QTA if a third-party API manufacturer is involved, clarifying who holds responsibility for approving and overseeing rework.Reprocessing Repeating an established process step to bring material back into specification. Recrystallising, drying again. Acceptable if within validated process. Must be documented. Rework Introducing a process not part of the approved manufacturing SOP to bring material into compliance. Using a different solvent or pH adjustment not in MA. Higher risk. Must be justified, documented, and may need regulatory notification or variation. 3. Considerations for Market Action: If there is concern over the reworked API and product quality cannot be assured: * Contact DMRC (MHRA) to agree recall class and level. * Recall should not be executed without MHRA/DMRC agreement. * Patient safety risk assessment must consider: * New impurity profile. * Indication (e.g., critical medicines, paediatrics). * Batch size and distribution status. * If alternative product not available: * Coordinate with the Marketing Authorisation Holder (MAH) and regulatory team. * Notify the Department of Health and Social Care (DHSC) through the DASH Portal to support shortage management.
41
You receive a call from your QC lab you have an OOS and an OOT for a low assay result on a dry powder inhaler product it's for a 12 month stability time point - what do you do?
a. Further information on questioning – 2 sets of assay results one OOT and one OOS, but mean within specification and the mean is what has been registered b. Discussed other time points, all ok c. Discussed other tests – expelled dose was also low 73% - medical have advised no lack of efficacy would be noted d. No increase in complaints from PV e. Asked about market – informed that all batch is used within 3-4 months of being certified and placed on market f. Went through oos/t investigation
42
g. Phase 2 found out that maintenance on the sealing jaws of the dpi tablet sealing during batch - wasn't sealing correctly h. Temperature was high i. Some unsealed / some too high temp exposure
j. Went into recall for any remaining stock on the market
43
You receive a call that a tablet on stability has failed dissolution stage 1
a. Went through OOS/T – were interested in phase 2 investigation b. Went through investigation and potential probable root causes c. Stopped me at API as had another scenario later on this .
44
What is analytical method transfer?
Analytical Method Transfer is the documented process of transferring a validated analytical procedure from a transferring (sending) laboratory to a receiving (testing) laboratory, to ensure the method can be performed reliably at the new site without the need for full revalidation. ⸻ Purpose: * Ensure consistent, accurate, and reproducible results at the receiving lab. * Required when: * Adding a new QC site. * Outsourcing to a contract lab. * Transferring testing within a company network. ⸻ Governing Guidance: * ICH Q2 (R2) – Validation of analytical procedures (ensure method is suitable and robust). * EU GMP Chapter 6 – Quality control. * EU GMP Annex 15 – Qualification and Validation (section on method transfer). * MHRA Inspectorate Blog – Practical expectations and common deficiencies. ⸻ Controlled by Change Control (CC): * A formal change control should be raised and approved. * The change control should include: * Gap analysis: Between transferring and receiving labs (equipment, reagents, training). * Impact assessment: On licence compliance, GMP status, and patient risk. * Method transfer protocol: Outlines method, number of replicates, sample types, acceptance criteria. ⸻ Typical Method Transfer Protocol Includes: * Description of test(s) (e.g., assay, related substances, dissolution). * Reference standards, samples to be tested, and testing schedule. * Acceptance criteria (e.g. %RSD, mean recovery, system suitability). * Transport and storage conditions of test items and standards. * Staff training and qualification at receiving site. * Handling of deviations or OOS results. * CAPA in case of failures or non-conformance.
45
How do you transfer assay of a tablet?
1. Initiate Change Control (CC): * Open a formal change control to manage the transfer. * Justify the transfer (e.g., new testing site, outsourcing, business continuity). * Appoint responsible parties from both the transferring lab and receiving lab. 2. Conduct Gap Analysis and Impact Assessment: * Senior analytical staff perform a gap analysis covering: * Equipment differences * Reagents and standards * Environmental conditions * Analytical experience * Conduct an impact assessment on: * Product licence (MA/variation requirements) * GMP compliance * Finished product release timelines and patient impact 3. Prepare Analytical Method Transfer Protocol: Include: * Defined test: e.g., Assay of API by HPLC * Samples and reference standards to be tested * Test method (as per validated procedure) * Acceptance criteria (e.g., % recovery, %RSD, system suitability) * Number of replicates or runs to be performed * Transport and storage conditions for samples and standards * Any special handling instructions * Deviation management process * Requirement to raise CAPA if the transfer is unsuccessful * Training requirements for receiving lab analysts 4. Ensure Prerequisites Are Met: * The receiving lab must complete equipment qualification (IQ/OQ/PQ). * Analysts must be trained and qualified. * Perform partial revalidation as needed to demonstrate suitability (aligned with ICH Q2(R2) and Annex 15), typically covering: * Accuracy * Precision (repeatability, intermediate) * Linearity * Specificity * Robustness (if environmental or procedural conditions differ) 5. Execute the Comparative Study: * Test a common sample batch in both labs under identical conditions. * Compare results using statistical methods (e.g., t-test) to confirm equivalence. 6. Review and Conclude Transfer: * QA reviews all results and confirms success. * Document in a transfer report. * Update SOPs, method files, and quality agreements as needed.
46
Explain T test in detail
A t-test is a statistical method used to compare the means of two sets of data to determine if they are significantly different from each other. It is commonly used in analytical method transfer to assess if the receiving lab generates results equivalent to the transferring lab. ⸻ Step-by-Step Explanation of a t-test: 1. Establish Hypotheses: * Null Hypothesis (H₀): There is no significant difference between the two sets of results. * Alternative Hypothesis (H₁): There is a significant difference. 2. Collect Data: * Use the same sample measured in both labs. * For each lab, calculate the mean (x̄) and standard deviation (SD). 3. Calculate the t-value: * For unpaired data (e.g., different analysts or conditions), use: t = \frac{|\bar{x}_1 - \bar{x}_2|}{\sqrt{\frac{s_1^2}{n_1} + \frac{s_2^2}{n_2}}} * Where: * \bar{x}_1, \bar{x}_2 = means of lab 1 and 2 * s_1, s_2 = standard deviations * n_1, n_2 = number of replicates 4. Determine Degrees of Freedom (df): * For unpaired tests: df \approx \frac{\left( \frac{s_1^2}{n_1} + \frac{s_2^2}{n_2} \right)^2}{\frac{(s_1^2/n_1)^2}{n_1-1} + \frac{(s_2^2/n_2)^2}{n_2-1}} 5. Find the Critical t-value from the t-distribution table: * Based on degrees of freedom and desired confidence level (e.g., 95% = p < 0.05). 6. Interpret the Result: * If the calculated t-value < critical t-value, fail to reject H₀ → no significant difference. * If calculated t-value ≥ critical t-value, reject H₀ → there is a significant difference. Alternatively, compare the p-value to your threshold (usually 0.05). If p < 0.05, it’s statistically significant. ⸻ Application in Method Transfer: * If the t-test shows no significant difference between labs, the method transfer can be considered successful. * This adds confidence in inter-lab equivalency, especially for critical tests like assay or content uniformity.
47
What parameters are validated for an HPLC assay method?
1. Specificity • Ability to measure the analyte response in the presence of potential impurities, excipients, or degradation products. • Demonstrated using blank, placebo, and forced degradation samples. 2. Linearity • Assesses the correlation between analyte concentration and detector response. • Typically performed at 5 concentration levels (e.g. 80%–120%), with r² ≥ 0.999. 3. Accuracy (Recovery) • Measures closeness of test results to the true value. • Performed by spiking known quantities of API into placebo at 3 concentration levels (e.g. 80%, 100%, 120% of target concentration), with 3 replicates each. • Total: 9 independent sample preparations, evaluated for % recovery (typically acceptable range: 98–102%). 4. Precision • Includes: • Repeatability: Same analyst, instrument, short time frame (usually 6 injections). • Intermediate precision: Different analysts, instruments, days within the same lab. • Reproducibility: Optional, between different labs. • Evaluated using %RSD. 5. Range • Interval where the method demonstrates suitable accuracy, precision, and linearity. • For assay methods: typically 80% to 120% of label claim. 6. Robustness • Assesses the method’s resilience to small, deliberate variations (e.g. ±10% flow rate, pH variation, mobile phase composition). • Identifies critical method parameters. 7. Solution Stability • Confirms that standard and sample solutions remain stable throughout analysis. • Evaluated at different time points (e.g. 0, 6, 24 hours).
48
When are LOD (Limit of Detection) and LOQ (Limit of Quantitation) required, and are they needed for assay methods?
• LOD and LOQ are not required for assay methods. Assay methods quantify the API at high concentrations (e.g. ~100% of label claim), where sensitivity is not the focus. • LOD and LOQ are required for methods that detect or quantify low levels of impurities or degradation products, especially when the impurity may have safety implications. Use cases: • Required for: • Impurity profiling (e.g. genotoxic impurities, degradation products) • Residual solvent analysis • Limit tests (e.g. elemental impurities) • Not required for: • API or drug product assay • Content uniformity (typically) LOD determines the lowest amount of analyte that can be detected (not necessarily quantified). LOQ is the lowest amount that can be quantified with suitable accuracy and precision.