Anesthesia Machine Checkout Flashcards

(33 cards)

1
Q

Step 1

A

Verify backup ventilation equipment is funtioning

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Purpose of verifying backup equipment

A

If the machine fails, the AMBU bag is the only way to provide positive pressure ventilation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Step 2

A

Check Oxygen cylinder supply (at least 1000 psi)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Purpose of checking Oxygen cylinder supply

A

It is your backup 02 supply if the pipeline fails

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Step 3

A

Check central pipeline supplies (About 50 psi)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Purpose of checking central pipeline supplies

A

To ensure adequate gas supply

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Step 4

A

Check initial status of low pressure system (everything off)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Purpose of checking low pressure system

A

To get the machine ready for the next steps

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Step 5

A

Perform a leak check of low pressure system

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Purpose of performing a leak check on the low pressure system

A

It checks for leaks in the flow meter tubes and vaporizers

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Step 6

A

Turn the machine ON

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Purpose of turning the machine on

A

The flow meters don’t work when the machine is off

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Step 7

A

Test the flow meters

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Purpose of testing the flow meters

A

To make sure the floats don’t get stuck and give false readings and to make sure that the hypoxic guard mechanism prevents FiO2 setting less than 25%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Step 8

A

Adjust and check the scavenging system

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Purpose of checking the scavenging system

A

A loose of improperly adjusted vacuum will release waste gases into the OR
A broken negative pressure valve will allow excessive vacuum and create sub-atmospheric pressure
A broken positive pressure valve will cause buildup of waste gases

17
Q

Step 9

A

Calibrate the O2 monitor (21% - >90%)

18
Q

Purpose of calibrating the O2 monitor

A

To prevent delivering a hypoxic gas mixture to the patient and to know when a hypoxic gas mixture is being delivered (via alarm)

19
Q

Step 10

A

Check the initial status of the breathing system

20
Q

Purpose of checking the initial status of the breathing system

A

An improperly attached circuit can cause ventilation problems
Exhausted CO2 absorbent causes rebreathing
Changing the absorbent can cause a circuit leak

21
Q

Step 11

A

Perform a breathing system leak check

22
Q

Purpose of checking the breathing system for a leak

A

To see if closing the APL valve or using the O2 flush creates a leak (high pressure situations)
Use the fresh gas flows to measure the amount of the leak

23
Q

Step 12

A

Test ventilation systems and unidirectional valves

24
Q

Purpose of testing the ventilation systems and unidirectional valves

A

Ensure the ventilator works properly
High gas flows will hide a leak
Stuck inspiratory valve will increase pressure in the machine
Stuck expiratory valve will increase the pressure inside the patient’s lungs

25
Step 13
Check capnometer
26
Purpose of checking the capnometer
to ensure that it works properly
27
Step 14
Check the final status of the machine
28
Purpose of checking the final status of the machine
to ensure that you are ready to start the case once the patient is in the room
29
Three tips for troubleshooting leaks
- Start from the patient end and move backward - Check connections - Adjust CO2 absorber
30
What to do when the CO2 waveform is distorted or nonexistent
- Check to make sure the sample line is tight and functioning by breathing into it - Make sure line is not occluded or bent
31
What to do if the suction is not working/weak
- Check the wall suction - Make sure the canister top is not loose - Make sure all ports are closed - Tighten suction tubing
32
Which steps need not be repeated after initial checkout?
Steps 1-9
33
Which four steps are often skipped?
Steps 5 (low pressure leak check), 8 (checking the scavenging system), 9 (calibrating the O2 monitor), and 13 (checking and calibrating alarms)