Cardiology measurement Flashcards
(218 cards)
Describe the oscillometric measurement of blood pressure
The cuff is coupled to an oscillometer which measures the pulse of the brachial artery transmitted through the air-filled tubing
As it inflates, the oscillometer stores the maximal amplitude of oscillations.
The cuff inflates above the systolic (i.e. when the oscillometer no longer sees any oscillations)
The cuff is gradually deflated until oscillations are 25-50% of their maximal amplitude: this is the systolic.
The cuff is gradually deflated until maximal amplitude is reached: this is the MAP
The cuff is deflated until the amplitude decreases again by 80% or more; this is the diastolic.
What sources of innacuracy are there in NIBP measurements
Wildly inaccurate: 95% CI for NIBP within the normal range is 15mmHg
Even more inaccurate in the extremes of blood pressure; over-estimates low blood pressure and under-estimates high blood pressure
Impossible to calibrate
Cuff size is a major influence of measurement
Oscillometer is confused by arrhythmia, shivering, or tremor.
What 2 methods of calibration are involved in using arterial lines
Static calibration
Dynamic calibration
What are the essential steps to static calibration of an arterial line
Zeroing the transducer
Checking and adjusting the gain
Checking for time stability
How do you zero an arterial line transducer
Zeroing - sets the zero reference point for pressure measurement
Technique
- Line off to patient - 3 way tap
- Position the tap at the level which is to be used as the zero reference point
- Open the tap in the transducer dome to air so the transducer is exposed to atmospheric pressure
- press the zero button the monitor
- Close the tap on the dome
- Open the 3 way tap so there is a continuous column of fluid between the tranducer and arterial lumen
What is involved in dynamic calibration of arterial lines
Resonant frequency
Damping coefficient
What are the advantages of optimal damping?
What is the otimal damping coefficent
0.64
If the pulse rate was 60bpm, what is the fundamental frequency? What therefore must the resonant frequency of the system be above in order to avoid resonance?
1 Hz
10Hz is the upper limit of the frequency of the 10th harmonic; as the resonant frequency must be >0.64 of this the answer is 15Hz
brandis 264
What is the SI unit of pressure
pascal
What is the typical pressur ein a hospital gas pipeline
400kPa
4 atmospheres
60psi
What is the difference between gauge pressure and absolute pressure
What is an ECG
graphic representation of myocardial electric potential against time used for monitoring and diagnosis of heart disorders
Components to producing an ECG
ECG electrodies
Cables
Amplifier
Processor
Monitor and recording device
How many ECG electrodes are used
3, 5, 10
What makes up an ECG electrode - why is this important
◦ Electrodes are disposable, thin layers of silver electrode and silver chloride on its surface covered in a gel that is rich in chloride ions - this combination results in a stable electrode potential that odes not interfere with recording
◦ 10mm diameter thin and broad, conducting gel to improve skin contact, high sampling rate 10000 - 15000 Hz to detect pacing spikes
What happens at the level of an ECG electrode to cause detection?
◦ Small changes in potential difference at the skin surface cause polarisation fo the silver/silver chloride electrode
What properties of ECG cables are important
each electrode will have a cable that returns the signal to the monitor
◦ Insulated to avoid eddy currents from surrounding electrical/magnetic sources
Why si amplification necessary for ECGs?
- Amplification of electrode signals is required as although a change in myocardial potential may be 120mV (from -90mV to +30 mV during an action potential intracellularly, and extracellular charge the inverse), the amplitude of a QRS at the skin surface is 1-2mV
What is an ECG lead
◦ Lead is a measure of the potential difference between two electrodes examining the heart’s potential difference changes at different angles
‣ The electrode at one end of the lead acts as the positive terminal while the other a negative terminal e.g Lead 1 LA is positive and RA negative
‣ Depolarisation towards the + terminal or repolarisation away results in positive deflection in the ECG
‣ E.g. Lead 1 measuring potential difference between LA and RA
What direction does depolarisation and repolarisation run in leads? Which is postivie and which is negative?
◦ Lead is a measure of the potential difference between two electrodes examining the heart’s potential difference changes at different angles
‣ The electrode at one end of the lead acts as the positive terminal while the other a negative terminal e.g Lead 1 LA is positive and RA negative
‣ Depolarisation towards the + terminal or repolarisation away results in positive deflection in the ECG
‣ E.g. Lead 1 measuring potential difference between LA and RA
What is the indifferent electrode? What charge does it have>
◦ Use of the indifferent electrode allows for unipolar analysis against a single lead and this is utilised for the chest leads but also for the augmented leads; it is always the negative electrode
Describe some of the processing elements to an ECG that aims to reduce interference
‣ This is also the site of common mode rejection of interference - where sources of electrical noise which affect each electrode equally are eliminated
‣ Differential amplicfication amplifies the difference between electrode elads rather than absolute voltages
‣ High frequency filters reduce - muscle and mains current interference
‣ Low frequency filters - screen respiratory movements
‣ Wider range frequency used for diagnostic 0.05 - 100Hz
‣ Reduced frequency used for monitoring 0.5 - 40Hz
What does 1cm represent vertically on an ECG?
1mV