Ch 15-18 Flashcards
(131 cards)
2 factors contributing to Compton scatter
kVp (affects beam penetrability)
Volume of irradiated material (FS and patient thickness)
3 affects on interactions increasing kVp has
Increased transmission
Decreased photoelectric absorption
Increased Compton scatter
Interactions that happen in the body when x-rays get absorbed
An interaction between x-rays and matter characterized by an incident electron with slightly greater energy than the binding energy of the electrons in the inner shells, ejecting an electron from the inner shell while being absorbed in the reaction, resulting in an ionized atom
Photoelectric absorption
3 affects on patient dose increasing kVp has
Decreased dose
Decreased photoelectric absorptions because x-rays have more energy to pass through the body
Increase in kVp typically accompanied by reduction in mAs
Affects on image quality increasing kVp has
Low/longer scale of contrast (many shades of gray)
2 affects on interactions decreasing kVp has
Decreased transmission and scatter
Increased photoelectric absorption
Affect on interactions decreasing kVp has
Shorter scale/increase in contrast (black and white)
Don’t have as much penetrating ability so x-rays absorbed
What does increasing field size affect?
Increases volume tissue irradiated and results in increased scatter and patient dose
What does decreasing field size affect?
Decreases beam quantity, scatter, and shortens scale of contrast image
4 beam restrictors
Aperture diaphragm
Cones
Cylinders
Collimators
Flat sheet of metal, usually lead, with an opening cut in the center and attached to the x-ray tube port
Simplest of all beam-restricting devices and different diaphragms are needed to accommodate different receptor sizes and distances
Fixed field size
Aperture diaphragm
Circular aperture diaphragms with metal extensions and has an extension that flares or diverges, with the upper diameter smaller than the bottom flared end
Most effective means of scatter control
Fixed field size
Cones
Circular aperture diaphragms with metal extensions and has an extension that flares or diverges, with the upper diameter smaller than the bottom that doesn’t flare
Fixed field size
Cylinders
A set of lead shutters at right angles to one another that move in opposing pairs
Collimator
Devices tailored for a specific use during a given procedure, designed to restrict the beam to a specific shape for a particular examination
Used to absorb scatter produced by patient
Must check vendor for information for digital systems before using lead blockers or masks
Ancillary devices
Get rid of off-focus radiation
Upper collimators
Radiation produced in tube someplace other than the anode
Photons that were not produced at the focal spot; extrafocal radiation
Off-focus radiation
Uses light reflected off mirror to project coverage of x-ray beam
Proper adjustment of mirror necessary to accurately display location of exposure field
X-ray beam coincide testing should be part of quality control (QC) program
Needs to be accurate within 1/2 inch (2% of SID)
Light field
Reduce penumbra
Bottom shutters
Unsharp shadow around the sharp shadow
A geometric unsharpness around the periphery of the image
Penumbra
An automatic collimator that adjusts to the size and placement of the cassette
Possible to override (can reduce beam to smaller field size than cassette size)
Positive Beam Limitation (PBL) devices
Adjusts to the size and placement of the cassette
Automatic collimator
2 ancillary devices
Lead blocker
Lead mask
A sheet of impregnated rubber that can be cut to any size or shape
Shields
Lead blocker