Chapter 2 x-ray imaging Flashcards

1
Q

dominant source of scatter in x-ray imaging

A

compton
coherent is < 5 %

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

ratio of scatter x-rays to primary x-rays exiting a patient

A

5:1 in abdominal radiography

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

what does scatter do?

A

reduce contrast
-does not affect spatial resolution or mottle

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

say a lesion transmits half the x-rays of adjacent tissues. What is the contrast for a scatter to primary ratio of 0:1 and 1:1?

A

0:1 - contrast is 50%. 100 primary photons and 50 for lesion
1:1 contrast is 25 %. scatter does nothing. Get 200 photons for adjacent tissue and 150 for lesion

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

how is scatter radiation minimized?

A

scatter removal grids

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

what reduces/increases scatter

A

-scatter increases with increasing field size
-collimation reduces scatter. Also benefits patients as reduces dose to patient by restricting field size
-increases with increased patient thickness
-max patient thickness without a grid is 12 cm
-less scatter with high Z materials like bone (PE effect dominates)
-scater is less with lower tube voltage because PE will dominate over Compton

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

is scatter as concerning with small fields?

A

No, because most of the scattered photons will miss the image receptor

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

tube voltage, scatter, and grid required for extremity x-ray

A

55 kV, very low scatter, no grid

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

tube voltage, scatter, and grid required for skull x-ray

A

80 kV, moderate, grid

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

tube voltage, scatter, and grid required for chest x-ray

A

120 kV, high, grid

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

tube voltage, scatter, and grid required for abdomenx-ray

A

80 kV, very high, grid

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

what are anti-scatter grids made of?

A

parrallel bars of highly attenuating material like lead. strips are filled with low attenuating material like graphite or Al to keep strips aligned.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

where are grids placed

A

between patient and image receptor

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

what is grid ratio

A

ratio of strip height H over gap between the strips D (i.e HD)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

common grid ratios in radiography and fluoroscopy

A

~10

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

common line density in radiography and fluoroscopy

A

~ 45 lines/cm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

are grids focused?

A

yes, and have diverging strips that must be used at a specified distance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

what is a reciprocating grid?

A

grid moves during the treatment, rendering them invisible

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

what is the device that moves the grid?

A

Bucky

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

define primary transmission

A

% of primary (useful) radiation that passes though grid
-grids transmist about 70% of primary photons and absorb the rest

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

how much of the scattered photons do grids absorb?

A

absorb 90 %, transmit 10%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

what is Bucky factor

A

-ratio of radiation incident on the grid to transmitted radiation
-increase in patient dose due to use of a grid

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

busky factor for abdominal radiography

A

5

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

define contrast improvement factor

A

ratio of constrast with a grid to contrast without a grid, usually >/= 2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
artifacts with grid
grid cut-off. -caused by improper alignment, wrong focal spot to film distance for the focused grid, inverted grids
26
depedence of primary transmission, scatter transmission, image contrast, and patient dose with grid ratio
primary transmission increases for smaller grid ratio scatter transmission increases for smaller grid ratio dose decreases with smaller grid ratio contrast decreases with smaller grid ratio
27
when does geometric magnification occur?
when there is air gap between patient and image receptor
28
what happens to primary vs scattered photons when using an air gap?
primary photons will reach the image receptor and be geometrically magnified scattered photons are less likely to reach the image receptor usually with air gaps, do not require additional scattere removal
29
disadvantage of air gaps
-causes magnification which increases focal spot blurring- can require a smaller focal spot
30
where is magnification imaging typically used?
mammography and neuroradiology -can improve lesion visibility of a microcalcification cluster or a small blood vessel
31
what does analog film consist of?
-thin emulsion supported by a Mylar base -emulsion contains many small silver halide grains
32
briefly describe how film works
-x-rays sensitive the silver halide grains -exposed films containing latent image are processed through developer solution then a fixer solution -in the developer solution, sensitivzed silver halide grains are reduced chemically and form a speck of silver -unexposed grains are removed in the fixer -fixer also makes image permanent
33
what corresponds to light/dark regions when viewing a film?
-silver grains (where dose was delivered) appear dark because loight cannot prenetrate
34
what is film blackening
pattern represents the distribution of x-rays reaching the image receptor
35
describe film density
film density of 1 transmist 10% of visible light; density of 2 transmits 1%, density of 3 transmits 0.1%
36
optical density
-log(Itransmit/Iincident) or log (Iincident/Itransmit)
37
what is the target average film density in film radiography?
1.5
38
what is fog and base
level of blackening in absense of any radiation exposure base is density of film base alone -unexposed film has a base + fog level of 0.2 OD units
39
what is max film OD
3 -occurs when all grains in the film have been sensitized and reduced to silver grains during development
40
what is characteristic curve? screeen film
represents relation betwen radiation intensity incident on film (Kair) and resultant film OD
41
regions of screen film characteristic curve?
toe= low exposure region shoulder = high exposure region linear region in between toe and shoulder -only in linear region do you get good contrast
42
what is the amount of film contrast proportional to?
gradient (steepness) of linear portion of characteritic curve
43
what is latitude?
range of radiation intensities that result in satisfactory contrast
44
what is contrast film and where is it used
has steep cruve, narrow latitude used in mammo
45
where are wide latitude films used?
-shallow curves -used in chest radiography to capture wide range of exposures
46
how much of the incident x-rays does a film along system absorb?
1 %
47
what are intensifying screeens for?
contain phosphor crystals to absorb 50 times more of the incident x-rays than radiographic film
48
how do intensifying screens work?
-each x-ray absorbed in screen produces hundreds of visible light photons -screens convert x-ray pattern to light pattern
49
what do intensifying screens do for exposure times and patient doses?
-decrease exposure time and patient dose 50 X compared to film alone
50
what is spectral matching
some screens emit mainly blue light, others green light -light output from the screen and the light sensitivty of the film must be matched
51
pros and cons of thick vs thin screens
-thick screens absorv more x-rays but increase image blur because the light diffuses more before striking the film -thin screes absorb less x-rays but yield sharper images
52
where do white artifacts come from?
stuff stuck between the screen and film- light from the screen cannot reach the film
53
what happens if Kair is too low or too high?
too low = underexposed film too high = overexposed want Kair of about 3 uGy
54
explain automatic exposure control and how it affects patient dose
-AEC keeps Kair fixed -higher kV = less skin dose because photons are more penetrating and therefore less photons are required to get appropriate Kair (therefore mAs is reduced) -vice versa also true
55
how does filtration affect patient dose?
more filtration= higher energy photons = reduced mAs and dose when Kair is fixed
56
what is a scintllator?
emits light when exposed to radiation
57
scintillator absorbs a 35 kEv photon- how many light photons are produced?
1000
58
what are radiographis screens in mammo made from?
Gd2O2S (gadolinium oxysulfide)
59
2 compoments of flat panel detectors?
scintillator (CsI) coupled to array of light detectors (2000x2500)
60
why is CsI a commonly used scintillator?
k-edges are 36 kEv (Cs) and 33 keV (I)- good match for average x-ray energies of 40 keV (for 80 kVp) in radiography
61
are scintillators direct detectors?
no, indirect because x-rays are absorbed to produce light and this light is converted to charge
62
what are image intensifier input phosphors?
scintillators, usually made of CsI
63
do scintillators produce blur?
yes, because of spreading of light from x-rays -can be manufactured into columns which restrict the amount of light diffusion and reduce detector blurring
64
how do photoconductors works?
-absorb x-rays which are converted into charge. A voltage collects the charge -electronic signal is proportional to x-ray energy deposited
65
are flat panel detectors that use photoconductors direct detectors?
yes, because liberated charge is directly collected
66
what is most common photoconductor?
selenium k-edge is 13 keV so poor x-ray absorber at higher energies (> 80 kV). Alternatives are PbI and HgI -Se is good in mammo (average energy 20 keV)
67
pros and cons of thick vs thin scintillators
thick- high x-ray absorption, low patient dose, lots of blur thin- low x-ray absorption, high patient dose, less blur
68
characteristic curve for a digital detector?
linear over wide range of intensities (no toe or shoulder) -image brightness and contrast is independent of x-ray intensity, unlike screen-film -doubling x-ray intensity doubles the signal detected
69
do photoconductors or scintillators have sharper images?
photoconductors- charge doesn't spread out, thus less blur
70
explain photostimulable phosphors (computer radiography)
-made of barium fluorohalides (BaFBr) -x-ray photons are absorbed in the phosphor- absorbed energy is stored and creates latent image -plates are read out with red lasers to release the stored energy, that is emitted as blue-green light -amount of blue-green light from each pixel represents number of x-rays reaching this pixel -detected light pattern (5 million pixels, 2000x2500) is digitized and stored -white light erases the photostimulable phosphor plates so they can be reused
71
pros and cons of thick vs thin photostimulable phosphors
thick is impractical because of unavoidable light scatter. The read out (red) light scatters into adjacent pixels and releases light, increasing blur (i.e. light from different pixels scatters into "wrong" pixels) -thin absorbs fewer photons than scintillators- increases patient dose
72
is photostimulable phosphor as sharp as scintillator or photonconductor image?
No
73
briefly explain how gas detectors work (this is clearly for the non RO folks)
-ionization chamber -incident x-rays liberate electrons in a gas (Compton, PE- then those electrons lose their energy with collisons with atoms, producing more electrons) -total electron charge liberated is the signal- electrons are collected by positive anode
74
what would imaging gas detectors use for gas?
-not air -high Z gases -xenon (k-edge 35 keV)
75
why are gas detectors operated at high pressures?
-improve ability to absorb x-rays
76
are gas detectors used for clinical x-ray imaging systems?
rarely
77
what x-ray intensities can digital detectors handle? screen film?
-digital detectors handle x-ray intensities 100 X lower and 100 x higher than screen - film -screen-film requires Kair of 3 uGy (or else underexposed or overexposed_
78
dynamic range of digital radiographic systems vs screen-film
10,000:1 vs 40:1 for screen film
79
what happens if a digital detector is exposed to too little radiation?
-noisy image -won't be "light" -when digital detector signal is low, electronic noise becomes important and may degrade image quality
80
what happens to digital detector is exposed to too much radiation?
-image quality rarely affected but patient gets more dose
81
what does ADC do?
-analog to digital converter can also have digital to analog converters
82
what is a bit and a byte
computers store data in bits (can be 0 or 1) byte is 8 bits
83
what is required to display digital data?
digital to analog converter -computer graphics board is example of DAC
84
what is kilobyte
1024 bytes (2^10 bytes)
85
how much data do medical images contain?
megabytes (1024 kilobytes)
86
storage requirement for radiology department is on order of..
terrabyte (1024 GB)
87
what is baud rate?
rate of information transfer in bits/second
88
how long do computers take to transmit a chest x-ray (10 MB)?
ethernet- 10 Mbps - 1 minute fast ethernet - 100 MBps - < 10 s Gigabit ethernet- 1,000MBps - < 1 s
89
what is matrix size
# of pixels in a given dimension
90
what are pixels?
individual picture elements in a 2D image
91
what is information content of an image?
number of pixels * number of bytes/pixel
92
explain how pixels are coded using 1 or 2 byte
1 byte = 8 bits, 2 bytes = 16 bits -256 (2^6) shades of grey can be coded for 8 bits -adding a bit doubles number of shades of grey; removing a bit halves number of shades of grey
93
how much memory would a 1024x1024 matrix use?
1 MB for 1 byte coding, 2 MB for 2 byte coding
94
typical matrix size, byte per pixel, and nominal image size (MB) for nuc med
128x128, 1, 1/64
95
typical matrix size, byte per pixel, and nominal image size (MB) for MRI
256x256,2, 1/8
96
typical matrix size, byte per pixel, and nominal image size (MB) for digital ultrasound
512x512,1,1/4
97
typical matrix size, byte per pixel, and nominal image size (MB) for ncomputed tomography
512x512,2,1/2
98
typical matrix size, byte per pixel, and nominal image size (MB) for TV digital photospot/DSA
1024x1024,2,2
99
typical matrix size, byte per pixel, and nominal image size (MB) for digital readiography
2560x2048,2,10
100
typical matrix size, byte per pixel, and nominal image size (MB) for digital mammo
4096x6144,2,50
101
what is soft copy display?
present images on flat panel monitors
102
typical luminence of monitors
300 cd/m^2
103
where are 2 MP vs 3 MP pixel monitors used?
3 MP- digital display radiographs 2MP- MRI, CT 2MP color- NM, ultraosund, PET/CT
104
who developed test patterns to evaluate monitor performance? what is the test?
society of motion picture and television engineers -intensities range from 0% to 100% in 10% increments -display is good when 5% square in 0% background (dark) and 95% square in 100% background (white) can both be visualized
105
what are individual pixels assigned in digital imaging?
-location -gray scale value by group of bits
106
what offers identical grayscale image display on different monitors?
DICOM grayscale standard display function
107
what does lookup table do? ie LUT
-straight line that maps pixel values to image intensities
108
what does window level define?
center (midpoint) of window width and therefore overall brightness
109
what does window width display?
range of gray scale values displayed, resultant image contrast -all pixels with values below this range register as black and all those above as white
110
what happens to image contrast within the window range as the window narrows?
-increases
111
what does image inversion do?
make dark regions appear white and vice versa
112
what is unsharp masking?
-subtract a smoothed version from original and add it to replicate original -improves visibility of tubes, lines cathethers -increases noise and can introduce artifacts
113
what does histogram equalization do?
eliminates while and black pixels that contribute little diagnostic info and expands remaining data to full display range
114
what is energy subtraction?
-sutract projection radiographs obtained at 2 energies -ex: 2 radiographs obtained at high and low kV can be subtracted to eliminate bone-like structures and improve depiction of lung and soft-tissue -or produce bone-only image, to help see calcified and non-calcified lung nodules
115
are grids tyically used for bedside chest radiographs?
no, difficult to align
116
explain grid cutoff
when grids are not aligned corrently, the resultant images exhibit grid cutoff artifacts showing light regions where no x-rays reached the image receptor
117
what is gray scale display function?
relates pixel value to monitor brightness -provides identical grayscale image display on different monitors
118
typical bucky factor in adult abdo radiography
5
119
films with steep gradient result in images with improved...?
contrast
120
screen films absorb how many more xrays than film alone?
50X more film will absorb 1 %. Film-screen will absorb 50%
121
increasing film quality does what to patient entrance Kair?
reduces it since fluence is more penetrating, less is required at entrance to obtain what is required at the film
122
which has best resolution? photoconductor scintillator photostimulable phosphor
photoconductor because charge produced is not dispersed during the detection process -charge from scintillators is dispersed
123
what types of images require the least storage?
MRI