chpt 13,14,15 Flashcards

(54 cards)

1
Q

sterilization documents is not required for items that undergo immediate use steam sterilization

A

false

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

instruments must be cleaned before being sterilized for immediate use

A

true

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

PPE is not required when using IUSS to clean items

A

false

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

IUSS is the sterilization method of choice for single use items

A

false

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

following a procedure on a patient with confirmed or suspected creutzfeldt Jakob disease,IUSS is recommended to sterilize the instruments

A

false

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

IUSS was devoloped to process items when a facility does not have enough instrumentation to perform surgery

A

false

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

to sterilize an item using IUSS the manufacturers IFU must state that the device can be sterilized using an IUSS cycle

A

true

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

when processing items for immediate use items do not need to be disassembled prior to cleaning

A

false

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

when using a sterilization or disinfection process at the point of use, item preparation is not critical as when preparing items for terminal sterilization

A

false

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

all items sterilized or HLD at the point of use mut be carefully monitored and documented

A

true

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

the process by which instruments are steam sterilized for immediate use is called

A

immediate use steam sterilization

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

items to be disinfected or sterilized at the point of use must be

A

properly cleaned per the manufacturers instructions for use

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

the best way to transport items that have been processed using immediate use steam sterilization IUSS at the POU is

A

in a rigid container designed for IUSS

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

what is the length of time after sterilization that an IUSS item can be used

A

IUSS items must be used immediately

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

implantable devices

A

should not be sterilized using IUSS unless there is a tracking system in place

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

what is terminal sterilization

A

the process by which surgical instruments and medical devices are sterilized in their final containers allowing them to be stored until needed

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

IUSS documentation should include the

A

patient ID, sterilizer number and cycle , name of instrument

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

both AORN and the joint commission recommend that

A

the use of IUSS be minimized or decreased

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

when loading a steam sterilizer basins should be

A

placed on edge

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

the higher level of bioburden on the object the

A

more difficult it will become to sterilize

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

when combining loads,the proper way to load a sterilizer is to place

A

rigid containers on the bottom shelf

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

the three main phases of a terminal sterilizer cycle are

A

conditioning, exposure and exhaust

23
Q

steam sterilization is the most commonly used method for sterilization in healthcare because

A

cycles are fast and expensive

24
Q

what is the weakest part of the steam sterilizer

25
what is the coolest part of the steam sterilizer
drain
26
how frequently should a sterilizer chamber drain strainer be cleaned
daily
27
the most common reason for steam failure is
lack of steam contact with the contents of the load
28
for sterilization peel pouches should be placed
on edge paper to plastic
29
how long can items that have undergone immediate steam use steam sterilization IUSS be stored
they cannot be stored
30
to ensure successful sterilization the sterilization chamber should not be
overloaded
31
items to be disinfected or sterilized at the point of use must be
properly cleaned per the manufacturers instructions for use
32
the best way to transport items that have been processed using immediate use steam sterilization at the point of use is
in a rigid container designed for iuss
33
what is the length of time after sterilization that an iuss item can be used
IUSS items must be used immediately
34
which of the following statements about IUSS is true
it should only be used when there is not time to process items using the wrapped method
35
Both AORN and TJC recommend that
the use of IUSS be minimized and decreased
36
implantable devices
should not be sterilized using IUSS unless there is a tracking system in place
37
when transporting items that have undergone IUSS it is required that they be
transported in a way that reduces the potential contamination
38
what is terminal sterilization
the process by which surgical instruments and medical devices are sterilized in their final containers allowing them to be stored until needed
39
when loading a steam sterilizer , basins should be
placed on edge
39
IUSS documents should include the
patient id sterilizer number and cycle name of the instrument sterilized
39
the higher the level of bioburden on an object, the
more difficult it will it be to sterilize
40
when combining loads , the proper way to load a sterilizer is to place
rigid containers on the bottom shelf
41
which of the following does not affect sterilization
how the user department tells the sterile processing SP staff to sterilize it
42
sterile processing technicians need to understand the anatomy of a steam sterilizer to
understand how the sterilizer operates
43
the weakest part of a steam sterilizer is the
door
44
the three main phases of a terminal steam sterilizer cycle are
conditioning, exposure and exhaust
45
the most common reason for stream sterilization failure
lack of steam contact with contents of the load
46
the coolest area in steam sterilizer is
drain
47
which of the following is not part of the sterilizer physical monitor printout
statement of guarantee of sterility
48
factors that can cause sterilant contact failure are
devices with soil remaining
49
one common cause of a clogged drain screen is
tape
50
sterilizercart can be unloaded
when a infrared device shows that the items have cooled to room temperature
51
what can create a pathway for microorganisms to travel from the outside of a package of to the inside of a package
moisture
52
how frequently should a sterilizers chamber drain strainer be removed and cleaned
daily