Electron dosimetry Flashcards

(51 cards)

1
Q

What is the relevant documentation for small field electron dosimetry?

A

TG 70

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

For a planar electron beam incident on a skin lesion that is above the flat skin surface, where would a dose hotspot be created?

A

lateral to the lesion (see Fig 32 of TG 25), beneath the nominal skin surface. The reason is due to the scatter generated in the above-skin lesion, scattering laterally into the air and towards the nominal skin surface.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

For oblique/curved surfaces irradiated by an electron beam, what affect does this have on the isodose curves?

A

The isodose curves roughly follow the curvature of the external contour. As the angle of incidence increases, dmax moves more towards the surface. This means the R90 coverage is going to be challenging to achieve in these oblique regions.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

How is electron beam quality specified?

A

R50

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

For absolute dosimetry in photon beams, the chamber is positioned at a depth of 10cm. For electrons, how is the reference depth specified? Provide the equation.

A

dref (cm)=0.6R50-0.1

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

What are the two protocols used for absolute absorbed dose to water for electrons and photons? What do we use in Australia?

A

AAPM TG 51, and IAEA TRS 398 rev1. In Australia we use TRS 398

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

According to TG 70, the reference field size for electron absolute dosimetry is 10x10cm2. As the beam energy increases, R_50 increases. Above what beam energy/R50 value should the field size for absolute dosimetry be increased to 20x20cm2 or greater?

A

Beam energies above 20 MeV, corresponding to a R50 >8.5 cm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

When a air-filled ionisation chamber is used to measure a depth dose profile of an electron beam, what is this curve called? What correction needs to be applied to get PDD?

A

PDI, can be converted to PDD using stopping power ratio tables.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

If an electron depth dose curve is measured with a solid state device, is the measured data a PDI or PDD? Why?

A

PDD (diode represents the dose directly), as the relative stopping power ratio between water and silicon is essentially independent of electron energy, and hence depth.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

For a broad beam electron field PDD, what sort of chamber is used?

A

PPC

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

For absolute dosimetry of a broad electron beam, what sort of chamber is used for absolute dosimetry?

A

Cylindrical ionisation chamber

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Are shielded or unshielded diodes recommended for electron beam dosimetry? Why?

A

Unshielded diodes are recommended for electron beam dosimetry. Shielded diodes have a high Z material around the sensitive region to attenuate low energy photons in a photon beam which contributes to an overresponse of the detector (remembering the mass energy absorption curve with energy of water: silicon)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Does mass attenuation coefficient (also known as mass absorption coefficient) pertain to photons or electrons?

A

photons, is the rate at which photons are attenuated across a cross-section/density of the attenuating material. Quantity is used for shielding calculations

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

what dosimetric quantity does mu_tr/p multiplied by photon energy fluence give?

A

Kerma= sum of all the kinetic energies of all primary charged particles released by photons, per unit mass.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

For depth dose beam scanning in water, should the scan direction be towards or away from the surface?

A

Scan direction should be towards the surface to reduce the effect of meniscus formation.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

For an electron field size, how is the output factor defined?

A

ratio of the dose per monitor unit on the CAX at dmax for a measurement field (M_field) to the dose per monitor unit for the reference field size taken at dmax.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

Before performing an output factor measurement, what information is needed?

A

PDD data

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

What happens to dmax when the field size becomes too small to produce lateral electronic equilibrium at the central axis?

A

dmax shifts closer to the surface, hence dmax for the same beam energy at different field sizes is likely different. Hence if measuring OPFs, PDDs should be performed to determine dmax for each field size.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

What effects can using an extended treatment distance (SSD) for electron beams?

A

(1) decrease in output factor (inverse square)
(2) reduced dose uniformity within the field
(3) widening of the penumbra (relative to measurements at the calibrated SSD of 100 cm)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

Dosimetrically, what is observed for small electron fields (3 points)?

A

When field size is <= radius required for lateral scattering equilibrium:
- dmax shifts towards surface
-output decrease
- CAX PDD decreases

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

As a rule of thumb, what is the equation given in TG 70 used to determine if an electron field can be considered small?

A

field radius, r,=0.88 x sqrt(E’_0)

Where average energy (E_0) can be determined using the PDD.

22
Q

If you were creating a library of small cut out information for electron fields, what information is relevant to include?

A

depth of dmax, depth dose data, output factor, and penumbral and isodose information.

23
Q

Tissue heterogeneities penetration of the beam and hence affect the dose distribution because of differences in what?

A

Stopping power, which are a result of different densities and tissue compositions

24
Q

what are the two types of electron beam dose calculation algorithms currently available in commercial TPS?

A

(1) pencil beam, (2) Monte carlo

25
What condition is considered when selecting an appropriate electron energy for a treatment?
Electron energy is selected so it covers the deepest portion of the target, to be treated to at least 90% of the given dose.
26
When using a cylindrical chamber for electron beam scanning, what is the generally accepted effective point of measurement equation used to place the chamber at the EPOM?
d=d_cent-0.5r where d=depth
27
For electron beams, as the field size increases what happens to the PDD? (see Fig 14, TG 25)
dmax shifts closer to the surface, and drop off after dmax occurs quicker with depth. As field size decreases, the PDD drop off tangent moves to the LHS of the graph. All graphs meet at the practical range point. Reason for field dependence on depth dose: lateral scattering of electrons
28
As the electron beam energy increases, what happens to the surface dose and why?
Surface dose increases
29
For internal shielding, a lead shield can be placed beyond the target to protect the underlying normal structures. e.g. treating lip or eyelid. Electron backscatter produces significant dose enhancement near the tissue-lead interface. How can this dose enhancement be minimised?
Using low atomic number material between lead and overlying tissue to absorb the backscattered electrons. Layer order: tissue, wax, lead, tissue
30
Photon beam question: What is the Mayneord F factor equation used for?
Inverse square law correction of PDDs from one SSD to another. For example, you have PDD data obtained at 100cm SSD but wish to treat the patient at 115cm SSD. Given the Mayneord F factor equation, the relative increase or decrease in % will be calculated for a given depth, e.g. 10cm. Mayneord’s F Factor= [(SSD1+dmax)^2* (SSD2+d)^2]/[(SSD1+dmax)^2* (SSD2+d)^2)] Quantitative example: for 6MVbeam (dmax=1.5cm) changing SSD from 100 to 150cm will result in PDD(10cm) increase by 5%
31
What is the 2,3,4 rule of thumb when it comes to electron beam range concepts?
x2: dmax/R100 x3: R90 x4: R50 x5: R10
32
When lateral scatter equilibrium exists (broad beam), how does the depth dose curve change for a fixed electron beam energy?
PDD is essentially independent of field size
33
A electron field can be considered 'small' when the field dimensions are smaller than the what?
Practical range of electrons (Rp)
34
What is the approximate energy loss per cm in water for electron beams?
2 MeV/cm
35
What is the equation 'rule of thumb' for estimating the practical range (Rp) depth for an electron beam of energy E?
Rp (cm)= E (Mev)/2 E.g. Rp for 10 MeV beam = 10/2=5 cm
36
For low-energy beams all the isodose curves buldge. For higher energy beams what happens to (a) lower isodose levels and (b) higher isodose levels?
(a) in higher energy electron beams, low energy isodose levels bulge (b) higher isodose levels laterally constrict for isodose levels >80%
37
For low energy electron beams (4-6 MeV), surface dose generally around what percentage?
80%
38
For higher energy beams, >20 MeV, what is the approximate surface dose % range?
90-100%
39
Explain the reasoning for the change in surface dose between low and high-energy electron beams in terms of fluence (# particles/unit area), scattering angle, and the ratio of surface dose to dose at dmax. (see Figure 14.10 in Khan, 3rd ed.)
As the beam energy increases surface dose increases. The reasons for this are: - fluence at the surface is dictated by relatively forward-directed electrons (i.e. tangential to the surface, electron paths approx parallel). - At low energies, electrons are scattered more easily & through larger angles (e.g. 45 deg) from the original incident direction, within the phantom. Comparatively, higher energy beams experience less lateral scatter and coincidence with smaller scatter angles. - consider the relative fluence of electrons at the surface is equivalent for both high and low energy beam scenarios. At depth, the low-energy beam is going to have a greater fluence (more particles in a set volume) than that of a higher-energy beam. - Hence, ratio of surface dose to dmax dose for lower energy electron beams is less - for higher energy beams, fluence at depth and surface is approx equivalent, hence ratio of surface dose to dose at dmax is greater
40
For electron beam energies (a) less than 4 MeV and (b) less than 20 MeV, what is the relative % contribution (normalised to dmax) of bremmstrahlung contamination in the tail section?
(a) 1% (b) 4%
41
At what depth is flatness and symmetry assessed for electron beams? What document states this and what is the tolerance?
dref (cm)=0.6R50-0.1 1% deviation from baseline is tolerance
42
Energy losses of charged particles are described by what quantity? What are the two components of this quantity?
stopping power consists of two components: (1) mass collision stopping power, resulting from electron-orbital electron interactions (atomic excitations and ionization), (2) mass radiative stopping power resulting from electron-nucleus interactions (bremsstrahlung production)
43
Of the two components of stopping power, which has an important role in radiation dosimetry? Give the equation for dose in terms of this component of stopping power
Collisional stopping power, (S/p)_c, where Dose (D)= e- fluence x (s/p)_c
44
How would you verify the accuracy of the PDI to PDD conversion for an electron scan performed with an ion chamber?
Use a diode and perform measurements at selected positions. Diode measurements will be dose.
45
TG 106: How would you determine the effective SSD of an electron beam energy?
measurements at dmax for various air gaps between electron cone and water surface, to create plot of the sqrt(I0/I) vs gap size. Gradient of line=effective SSD
46
Per TG 106, what are ideal features of a detector used for beam scanning?
high sensitivity, small dimensions, low noise, minimum dose rate and energy dependence
47
What detector type is preferred for relative dosimetry in photon beams?
Ion chambers with small volume
48
What detector type is preferred for relative dosimetry in electron beams?
electron diode
49
What portion of the electron depth dose curve does an electron diode no longer become the preferred detector choice? (TG 106)
Bremsstrahlung portion. For accurate measurement of Bremsstrahlung region, an ion chamber should be used.
50
What does TG 106 (2008) relate to?
Accelerator beam data commissioning equipment and procedures
51
which has greater surface dose, low or high energy electrons?
high