The light on the CU is _____ upon connection to power
Green
LED blinks ____ when charging and turns constantly on when fully charged
Green
Following an AC reset in the DP, a green blinking light means
AC is fully charged and ready to be used for night
Following an AC reset in the DP, a yellow blinking light means
AC is not fully charged and should be charged
Following an AC reset in the DP, a red blinking light means
Malfunction AC, should be replaced
You have ___ mins to connect AC to DP after taking out of CU
5
To log onto the Genio App, go to the IP and then type in password, which is…
GenioXX (last two characters of repeater MACID
What’s the first step in connecting to patients AC?
Pressing magnifying glass to find Bluetooth
What is the AC pairing code when connecting and prompted to put in?
The last 6 digits of the AC serial number
How many minutes do you have to pair to AC once password is asked?
2 mins
What happens if you miss the 2 min pairing window to patients AC during a titration?
Reset AC in CU
When programming duty cycle, it must be less than _____
80%
Name the letters
C = confirmatory stim
D = delay
L - train length
I = train interval
T/F - the RTC (real time clock) should be calibrated at the patients activation titration
TRUE, important for AC log
What is the authentication code for the smart phone app when promoted in initial set up?
6728
How do you locate the AC to connect to when promoted on patients phone?
First two letters of AC SN
How long of a window do you have to pair the AC to the patients phone?
2 mins
T/F - if the patient pauses from the phone app, the delay time is not in effect post pause
FALSE - it is