Impella and VADs Flashcards

(18 cards)

1
Q

What side of the heart does impella help

A

Left
*unless it’s an RP

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What should ACT be in someone with an Impella

A

160-180 seconds

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

What are patient considerations with Impella devices

A

Do not raise the bed over 30
Monitor distal pulses
Check for bleeding/hematoma
Keep leg straight

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

What should preload be with impella

A

<10

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

What are some reasons that can cause inadequate preload

A

Diuretics
Bleeding
CVVH
Right heart dysfunction
ECMO

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

What should MAP be on impella

A

60-80

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

How do you monitor after load on Impella

A

By monitoring MAP and SVR

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What are some reasons for high after load

A

Pressers
VA Ecmo
Physiologic effects

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What is the range of impella flow

A

P0- P9

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

How to tell if the impella is in proper position based on the monitor

A

Will show a pulsatile motor current

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

What are some trouble shooting strategies when there is suctioning present with impella

A

Reduce P by 1-2 levels
Assess volume status (add if cvp <10)
Confirm RV function
Assess position

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

What causes suctioning with impella

A

Inadequate LV filling
Incorrect position in LV
RV failure

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

How to recognize and fix diastolic suction

A

-negative diastolic pressure
- normal systolic
- low diastolic flow

Check filling/volume status

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

What is the purpose of the purge system with impella

A

To prevent blood from entering the device

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

What type of fluid is used for the purge system in impella

A

D5W and heparin

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

What does it mean if a PAPi is less than.9

A

There is RV dysfunction

17
Q

What is the correct positioning for an impella RP

A

Outlet at/above the bifurcation of right/left PA branch

Inlet in the IVC at the level of the diaphragm or apex of the heart

18
Q

How to calculate the cardiac output when impella RP is in use

A

FICK for accurate reading