Task group for commissioning TPS
Tg-157
Task group for proton including tps
Tg185
Explain step for commissioning linac
You want to do absolute dosimeter with ion chambers. They need to be calibrated by accredited national lab then you need to measure all the appropriate factors of correction
Name the formaulate for IC to get the right charge measurement fully corrected
M = Mraw Pion Pelec Ppol Ptp
Pion for the recombination of ion,with 2 bias voltage
[1-(Vh/Vlow)^2 ]/ [Mhigh/Mlow - (Vh/Vlow)^2 ]
Pelec : given by the accredited lab if the ic and electrometer were not calibrated together
Ppol:effect of polarization , measure with absolute of (M+ - M-)/(2M+)
Ptp: 273.2+T/273.2+22 x 101.3/P