L 21: SRS, GK, PSI Flashcards
(33 cards)
Winston-Lutz Test
checks with the small SRS cone for
* gantry isocentricity
* Laser alignment to the isocenter
Picket fence test
Garden fence test
Checks the MLC accuracy
Homogenity Index
Max PTV Dose / Prescription Dose
Conformity Index
- Volume encompassed by prescription / target volume
- Ideally between 1.0-2.0
Gradient Index
volume of 50% prescription dose / Volume of total prescription dose
* also known as the dose fall off outside the target
SRS elements
- Planning with high spatial resolution (MRI 1mm slices)
- Using narrow multiple beams, multi non-coplaner isocentric arc (improves dose conformality)
- Multiple isocenters, dyanmically shaping the field during arc roatation with mini MLCs
- Rigid immobilization
- Optimization is very important
- Highly accurate target (+/- 1mm)
- Isocenter displacment within +/- 0.1- 0.3 mm.
Trigeminal Neuralgia
dose gradient can be high of 16Gy/mm
Focused Cones:
- Reduces Penumbra
- long, narrow metal cones placed below the jaws to bring collimator closer to the patient surface: extends SDD so reduces teh penumbra
- Isocenter remains centered within the cone
- Not used now with mini MLCs
Pitch
Tilt
Yaw
Roll
SBRT CT-Sim
Include 15 cms above and below
SRS
- Single fraction
- Intracranial
- Narrow multiple beams, non co-planerIsocentric arcs
- High degree of dose conformality
- Multiple isocenters
- Mechanical accuracy +/- 1mm error acceptable
SRT
- Multiple fractions
- Intracranial
Types of frame
Not important for exam
- Leksell: Used here for GK
- Richard-mundinger
- Todd-wells
- Brown-Robert-wells
- Gill- Thomas- Cosman: Uses bite block
CO sources in GK
192-201 sources
6000Ci
Gamma Knife
- More than 1 isocenter
- multiple arcs
- can be with frame or frameless
- Dosimetry: %DD, TMR, Cross beam profiles, output factors
GK Questions
- Source: Co-60
- Half life: 5.26 years
- What energies: 1.17Mev -1.33 Mev avg 1.25 MeV
- What collimator sizes: 4/8/14/18: old machine; 4/8/16 mm new machine
- How many sealed sources: 192
- Source to focus distance: 40..3 cms
- Mechanical precision at isocenter: +/- 0.3mm
- Plugs: Made of tungsten alloy, 6cm thick, old machine, external
- Co decays to Ni and gives out gamma-rays
PSI (Prostate seed Implant)
- Can be used for early stage , confined to prostate gland, no EPE or SVI
- some patients might need PSI + EBRT
- Transperineal approach under the guidance of US.
- Inter-Seed placing: 1cm
- Inter-Needle spacing: 1cm
- Abstain from sexual activity for 2 weeks/use condoms
- Vendor certificate: +/- 10%
- Sample check: +/- 5%
- Basic requirement to release a permanent implant patient from the hospital does not exceed ** 0.5 rem (500mrem)**
- No contact with preg women or young children for 2 months
- No sexual intercourse in 2 weeks, use condoms
Temporary PSI
- Radio isotopes with relatively longer half life
- sufficient dose rate to deliver prescribed dose in 3-4 days
Permanent PSI
More popular
- Radio isotopes with shorter half life
- Photons of low energy
- I-125 or Pd-103
Prostate Nomogram
- Prostate volume: using Ultrasound, 5mm intervals
- # of seeds
- Strength of the seed
- Interseed spacing 1cm
- interneedle spacing 1cm
Seed Strength
- I-125: 0.3 mCi &
Pd-103: 1.7 mCi