Lesson 5.1: Central Sterilization Room (CSR) Flashcards

1
Q

Unprocessed biological monitor from the same lot as test monitor and verifies viability of unexposed microorganisms

A

Biological Control

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Bacterial endospore test that assesses if sterilization has occurred. “Biological Indicator” or “Biological Spore” test

A

Biological Monitor

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

A diagnostic test of pre-vacuum; checks sterilizer’s ability to remove air from the chamber and detects air leaks. Not a sterility assurance test

A

Bowie-Dick Test

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Destruction/Inhibition of most viruses and bacteria; doesn’t kill spores and cannot be verified by a monitor

A

Chemical Disinfection

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Continuous elapsed time while sterilizer is operating at preselected sterilizing parameters (temperature and pressure)

A

Exposure Time

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

A process using steam heat under pressure length of time to kill all microorganisms

A

Saturated Steam Sterilization

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Remains visibly wet following a complete sterilization/drying cycle

A

Wet Pack/Wet Load

How well did you know this?
1
Not at all
2
3
4
5
Perfectly