Machine Safety features Flashcards
(5 cards)
1
Q
Oxygen failure alarm
A
Has a distinct tone different from any other alarm. Cant be disabled. Cant be silenced. Audible/Visual display. Works irregardless of power.
2
Q
Anti-hypoxic devices
A
Chain link 25, pneumatic, paramagnetic. Nitrous oxide/oxygen mix no less than 25% oxygen delivered.
3
Q
Mechanical chain link
A
Oxygen and nitrous oxide linked by a chain to their control valves.
Nitrous cannot be turned on without turning oxygen on.
4
Q
Pnuematic
A
Pressure sensitive diaphragm that measures changing pressures in oxygen and nitrous oxide, and cuts off nitrous oxide if oxygen is too low.
5
Q
Paramagnetic analyser
A
Continuously measures oxygen concentrations.
Nitrous automatically switches off if oxygen falls below 25%.