Machine Safety features Flashcards

1
Q

Oxygen failure alarm

A
Has a distinct tone different from any other alarm.
Cant be disabled.
Cant be silenced.
Audible/Visual display.
Works irregardless of power.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Anti-hypoxic devices

A

Chain link 25, pneumatic, paramagnetic. Nitrous oxide/oxygen mix no less than 25% oxygen delivered.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Mechanical chain link

A

Oxygen and nitrous oxide linked by a chain to their control valves.
Nitrous cannot be turned on without turning oxygen on.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Pnuematic

A

Pressure sensitive diaphragm that measures changing pressures in oxygen and nitrous oxide, and cuts off nitrous oxide if oxygen is too low.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Paramagnetic analyser

A

Continuously measures oxygen concentrations.

Nitrous automatically switches off if oxygen falls below 25%.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly