Meter Flashcards
What does the multi RAE meter look for?
- LEL 10%-20%
- Oxygen (O2) 19.5-23.5
- Hydrogen sulfide (H2S) 10ppm-15ppm
- Carbon monoxide (CO) 35ppm- 100ppm
- Volatile organic compounds (VOC) 50ppm-100ppm
Multi RAE
What contaminants may give you false reading, damage sensors, or damage meter?
Organic silicones
Silicates
- Lead containing compounds
What are VOCs and what is the highest ionization potential (IP) that the meter can detect?
Volatile organic compounds Dash modern industry industrial chemicals such as fuels, coatings, solvents, refrigerant etc. they have a tendency to evaporate easily under normal conditions making them volatile and they have a high flashpoint.
10.6 eV
How do you do a bump test on a multi RAE?
- Once you’re at your home screen press and hold the Y+ & N- keys at the same time the bump screen will appear. Apply the 4 gas and press start.
- Now bump the PID sensor. Apply gas press start
Multi RAE
What do we want to do prior to metering a hazardous area?
Block pump inlet to make certain pump is working. Press reset
Bump test. If it fails bump test then calibrate.
Confirm measurement gas (conversion factor) is set appropriately
How does the multi RAE read VOC’s?
There are three different PID sensors That can be used. One reads ppm, and reads 0.1 to 5000 ppm, and another that reads 1 to 1000 ppm, third one that reads parts per billion 10 PPB - 2000 ppm
What gas do we calibrate LEL, 02, CO, and H2S on? What gas do we calibrate VOC’s on?
Multigas- methane, oxygen, carbon monoxide, hydrogen sulfide
Isobutylene
How do we calibrate Multie RAE?
- Must be in fresh air to perform
- home screen press “O” button and “N-“ at same time.
- Next screen will have six icons
- Select upper left icon that looks like calibration bottle
- Scroll down to “multi sensor span” & select
- Attach 4 gas and select start
- Once done you will see pass or fail the meter will go into PID calibration
- Apply Isobutylene press start
- Once complete select back button to go to the home page
When using multi RAE to meter for combustible liquids with flashpoint above 100°F what area of the meter should be watched?
VOC’s High carbon molecule products may give erroneous low readings in the LEL area
Multi RAE
What are the alarm levels?
10% LEL 02 19.5% and above 23.5% 35 ppm CO 10 ppm H2S 100ppm VOC
Toxi RAE
What three Gases Are supplied for this meter?
(NH4)Ammonia
(Cl2) Chlorine
(HCN) Hydrogen cyanide
Toxi RAE What does (max) indication code mean?
Gas exceeds electronics maximum range
Toxi RAE What does (over) indication code mean?
Gas exceeds measurement range
Toxi RAE What does (HIGH) indication code mean?
Gas exceeds hiGh Alarm limit
Toxi RAE What does (Low) indication code mean?
Gas exceeds lower alarm limit
Toxi RAE What does (STEL) indication code mean?
Gas exceeds STEL limit
Toxi RAE What does (TWA) indication code mean?
Gas exceeds TWA limit
Toxi RAE What does (CAL) indication code mean?
Calibration failed, or needs calibration
Toxi RAE What does (BAT) indication code mean?
Low battery
Name five sensors that can be used on TOXI RAE meter
ACDC
Ammonia Chlorine Nitrogen Dioxide Sulfur Dioxide Hydrogen Cyanide
What does the gas ranger look for?
Methane
What is 5% gas
50,000ppm
Gas ranger
When you were in % LEL when does it alarm?
10% LEL
Gas Ranger
When you are in %Gas When does the meter alarm?
5% Gas