Mid Term Exam Flashcards

(256 cards)

1
Q

T/F: All contrast media is potentially harmful and should be used knowledgeably.

A

True

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

T/F: Intravascular contrast media administration is performed under the supervision of a licensed physician.

A

True

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

T/F: Technical and nursing personnel injecting contrast media must be knowledgeable in the recognition of, and able to assist in the treatment of any adverse reaction.

A

True

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

T/F: Literature in the MRI community and manufacturer recommendations lists the dosage of intravenous gadolinium chelate contrast to be 0.1 mmol/kg of body weight.

A

True

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

T/F: Any and all adverse events, incidents or “near incidents” that occur pertaining to MR safety are to be reported to the MRI Medical Director, the MRI Department Supervisor or the Radiology Administrator immediately.

A

True

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

T/F: A student may review and sign the MRI screening sheet prior to an individual entering the scan room.

A

False

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

T/F: The Medical Director will be involved with the introduction of new policies and procedures as needed with the advancement of technologies and procedures in the MRI environment.

A

True

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

T/F: Zone III – This area is synonymous with the MR scanner room itself.

A

False

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

T/F: Lenz’s Law is an important consideration with non-ferromagnetic objects.

A

True

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

T/F: Having undergone a previous MRI successfully with an implanted device indicates that the device is MRI safe.

A

False

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

T/F: The potential for translational forces (“projectile forces”) exists at any field strength.

A

True

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

T/F: After a quench the patient and personnel should evacuate the area immediately.

A

True

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Which of the following describe T1 Recovery? (Choose all that apply)

Spin lattice relaxation

The time it takes 63% of transverse magnetization to decay in the tissue.

Caused by the nuclei giving up their energy to the surrounding environment or lattice.

Spin-spin relaxation that results in decay or loss of coherent transverse magnetization.

A

Spin lattice relaxation

Caused by the nuclei giving up their energy to the surrounding environment or lattice.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

What MUST be done prior to anyone entering the scan room? (Select the BEST answer)

Pat the patient down for metal objects.

A screening sheet must be filled out and in the patient’s chart.

The patient must be scanned with the metal detector.

The completed screening sheet must be signed by a level II staff member and the patient must be scanned with the metal detector, prior to admittance to the MRI scan room.

A

The completed screening sheet must be signed by a level II staff member and the patient must be scanned with the metal detector, prior to admittance to the MRI scan room.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

What happens to the alignment of hydrogen protons when placed in an external magnetic field? (Check all that apply)

Nothing until an RF pulse is applied.

The magnetic moments of the nuclei align with or against the main magnetic field.

More nuclei oppose the main magnetic field than line up with the main magnetic field.

Slightly more nuclei line up with the main magnetic field than against it.

A

The magnetic moments of the nuclei align with or against the main magnetic field.

Slightly more nuclei line up with the main magnetic field than against it.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Where do we look up the safety of an implanted device? (Check all that apply)

www.MRISafety.com

Shellock’s reference manual

The manufacturer’s website

If we remember that it is a safe implant, there is no need to look it up again.

A

www.MRISafety.com

Shellock’s reference manual

The manufacturer’s website

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

What effect does the magneto-hemodynamic effect have? Is this a permanent effect? (Check all that apply)

Increase in the amplitude of the “R” wave on EKG gating.

Increase in the amplitude of the “T” wave on EKG gating.

Reversible effect– EKG tracing returns to normal when the patient is removed from the static magnetic field.

Permanent effect– EKG tracing continues to show an elevated wave form after the patient has had an MRI exam.

A

Increase in the amplitude of the “T” wave on EKG gating.

Reversible effect– EKG tracing returns to normal when the patient is removed from the static magnetic field.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

What is the gyromagnetic ratio of hydrogen?

  1. 28 MHz/T
  2. 8 MHz/T
  3. 57 MHz/T
  4. 57 MHz/G
A

42.57 MHz/T

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

Select the three components of the magnetic field from those listed below.

Static Magnetic Field

Faraday’s Field

Radiofrequency Fields

Time Varying Magnetic Fields

Precessional Frequency Field

A

Static Magnetic Field

Radiofrequency Fields

Time Varying Magnetic Fields

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

What are the guidelines for pregnant employees? (Choose all that apply)

No changes to pregnant employees work environment in MRI. Everyone knows that MRI doesn’t use radiation.

Pregnant employees may enter the scan room, but should leave before scanning begins.

Pregnant employees are exempt from scanning and may only do paperwork.

Pregnant employees MUST transfer to x-ray once they’ve announced their pregnancy to the MRI department.

A

Pregnant employees may enter the scan room, but should leave before scanning begins.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

T/F: The gyromagnetic ratio expresses the relationship between the angular momentum and the magnetic moment of each MR active nucleus.

A

True

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

Identify which of the following describe a “fringe field.” (Select all that apply)

The stray magnetic field outside the bore of the magnet.

An area on the outskirts of the magnet– generally two feet long by three feet wide.

Active or passive shielding is used to reduce the size of the fringe field.

The fringe field is not confined by walls, floors, or ceilings.

A

The stray magnetic field outside the bore of the magnet.

Active or passive shielding is used to reduce the size of the fringe field.

The fringe field is not confined by walls, floors, or ceilings.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

What would you do in the event of a CODE BLUE in MRI? (Select all that apply)

Stay with the patient in the scan room until help arrives.

Begin CPR.

Activate the Code Blue response process (push the Code Blue button)

Get the patient out of the scanner and into Zone III; locking the scan room door and monitoring it when other Code Team members arrive

A

Begin CPR.

Activate the Code Blue response process (push the Code Blue button)

Get the patient out of the scanner and into Zone III; locking the scan room door and monitoring it when other Code Team members arrive

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

What are the two results of resonance? (Choose from the list below)

The NMV moves out of alignment and away from B0.

The NMV moves into a repeating pattern of /- 100 MHZ.

The magnetic momements of hydrogen nuclei move out of phase with each other.

The magnetic moments of hydrogen nuclei move into phase with each other.

A

The NMV moves out of alignment and away from B0.

The magnetic moments of hydrogen nuclei move into phase with each other.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
T/F: The "warm zone" in MRI is the scan room itself.
False
26
Which of the following best describe spin-lattice relaxation? (Choose all that apply) T2 Decay T1 Recovery Time it takes 88% of the longitudinal magnetization to recover in tissue. Nuclei give up energy to the surrounding environment, resulting in longitudinal recovery.
T1 Recovery Nuclei give up energy to the surrounding environment, resulting in longitudinal recovery.
27
What makes a nucleus an MR active nucleus? (Select all that apply) Odd mass numbers Atoms with a balanced mass number Nuclei with a net spin or angular momentum None of the above
Odd mass numbers Nuclei with a net spin or angular momentum
28
T/F: We ALWAYS scan pregnant patients with contrast.
False
29
T/F: For resonance of hydrogen to occur, RF at EXACTLY the Larmor frequency of hydrogen must be applied.
True
30
What are the three types of motion present in an atom? (Select all that apply) Electrons spinning on their own axis. Nuclei spinning around each other. Electrons orbiting the nucleus. The nucleus itself spinning about its own axis.
Electrons spinning on their own axis. Electrons orbiting the nucleus. The nucleus itself spinning about its own axis.
31
Identify the two factors that determine which hydrogen nuclei align parallel and which align anti-parallel with the main magnetic field. The strength of the external magnetic field How many nuclei there are in each patient's body The thermal energy level of the nuclei The orientation of the patient in the scanner (head first vs. feet first)
The strength of the external magnetic field The thermal energy level of the nuclei
32
Which way does the main magnetic field run in permanent magnets? Horizontally Vertically Side to Side Down the bore of the magnet
Vertically
33
What is the terminal velocity of a paperclip at 1.5T? 64 mph 55mph 40mph None- not attracted
40mph
34
Which of the following describe the purpose of gradients in MRI? (Choose all that apply) Alter the main magnetic field in a linear fashion Change thermal equilibrium Spatially encode data Cause electrons to spin on their own axis
Alter the main magnetic field in a linear fashion Spatially encode data
35
What does SAR stand for? Special Atomic Rate Start Atoms Rolling Specific Absorption Rate Specific Atomic Rate
Specific Absorption Rate
36
What does the net magnetization vector (NMV) represent? The type of MR active nuclei present The relative number of electrons to protons The atomic number of the nuclei present The relative balance of spin-up to spin-down nuclei
The relative balance of spin-up to spin-down nuclei
37
Which way does the main magnetic field run in superconducting magnets? Along the bore of the magnet (horizontal) Along the Y-axis Vertically Side to side
Along the bore of the magnet (horizontal)
38
What is the best device to use when monitoring a patient's status in MRI? Blood pressure cuff Pulse ox EKG Auditory monitoring
Pulse ox
39
What does the Larmor Equation calculate? The NMV The precessional frequency The gyromagnetic ratio The atomic mass
The precessional frequency
40
What is the T2 relaxation time of tissue? [Hint: Think about which axis T2 relaxation is occurring in (longitudinal vs. transverse) and whether T2 relaxation indicates recovery vs. decay.] The time it takes 43% of the longitudinal magnetization to recover in the tissue. The time it takes 63% of the transverse magnetization to recover in the tissue. The time it takes 63% of the longitudinal magnetization to decay in the tissue. The time it takes 63% of the transverse magnetization to decay in the tissue.
The time it takes 63% of the transverse magnetization to decay in the tissue.
41
What is another term for T2 Decay? Spin-lattice relaxation Spin-spin relaxation T1 recovery NMV
Spin-spin relaxation
42
How do gradients affect the main magnetic field? (Select all that apply) They add to the main magnetic field They subtract from the main magnetic field They alter the main magnetic field in a linear fashion They give the image its shades of gray
They add to the main magnetic field They subtract from the main magnetic field They alter the main magnetic field in a linear fashion
43
T/F: Rewinders are gradients that rephase.
True
44
Which of the following controls T2* weighting in gradient echo? TE TR Flip angle TR and flip angle
TE
45
What controls the polarity of a gradient? The steepness of the slice select gradient The sharpness of the frequency readout The direction of the current through the gradient coil The strength of the main magnetic field
The direction of the current through the gradient coil
46
T/F: TR controls T2 weighting.
False
47
Which of the following is known as the basic MRI contrast? T1 weighting T2 weighting Gray imaging Proton density weighting
Proton density weighting
48
# Choose extrinsic contrast parameters from the list below. ``` T1 Recovery Time TR Proton Density Flip Angle Flow TE b Value ```
TR Flip Angle TE b Value
49
What are the two extremes of contrast in MRI? T1 Recovery Fat Air Water
Fat | Water
50
T/F: The TE is twice the TAU.
True
51
T2* decay is faster than T2 decay since it is a combination of which two of the following effects? T1 decay with spin echo T2 decay itself Dephasing due to magnetic field inhomogeneities *weighting
T2 decay itself Dephasing due to magnetic field inhomogeneities
52
Which of the following best describes a T1 weighted image? (Select all that apply) Bright water Bright fat Dark Water Dark Fat
Bright fat Dark Water
53
Which of the following statements is true: Spin-spin interactions drive T1 contrast Slow tumbling molecules lead to decreased spin-spin dephasing Fast tumbling molecules lead to increased spin-spin dephasing Spin-spin interactions drive T2 contrast
Spin-spin interactions drive T2 contrast
54
What is the approximate T1 relaxation time of water at 1T? 200 ms 500 ms 2000 ms 2500 ms
2500 ms
55
Which of the following items are T1 and T2 relaxation times dependent on? (hint: there are three) How well the molecular tumbling rate matches the Larmor frequency of hydrogen Spin-spin interactions of oxygen atoms How closely packed the molecules are The inherent energy of the tissue The number of electrons orbiting the nucleus
How well the molecular tumbling rate matches the Larmor frequency of hydrogen How closely packed the molecules are The inherent energy of the tissue
56
T/F: Intrinsic parameters are those that cannot be changed because they are inherent to the body's tissues.
True
57
To achieve T1 weighting in gradient echo sequences, the flip angle should be ______. Large Small No flip angle should be used
Large
58
T/F: Spin echo sequences do not compensate for magnetic field inhomogeneities, while gradient echo sequences do compensate for magnetic field inhomogeneities.
False
59
___ controls the amount of T1 weighting. TR TE TI PD
TR
60
For PD weighting, the TR must be ______. Long Short Intermediate T1
Long
61
T/F: Generally speaking, gradient echo sequences are faster than spin echo sequences.
True
62
For T2 weighting, the TE must be _____. short long intermediate PD
long
63
Gradients that dephase are called __________. rewinders weighted spoilers variable flip angles
spoilers
64
The spin echo is created by: The application of a long TE The application of a rewinding gradient A 180 degree RF pulse A rapid TI
A 180 degree RF pulse
65
Gradients are used to rephase spins in gradient echo sequences because: The resulting images have less artifact than spin echo images They are faster to apply than an RF pulse They rephase spins more completely than an RF pulse They result in images with better contrast resolution
They are faster to apply than an RF pulse
66
Which of the following controls T1 weighting in gradient echo? TI TE Gradient amplitude TR and flip angle
TR and flip angle
67
Saturation results in T1 weighting because: The flip angle used was small so vectors are not pushed beyond the transverse plane Only occurs when the TR is long as this allows for full recovery Vectors are pushed beyond the transverse plane because they were completely recovered before the RF excitation pulse Vectors are pushed beyond the transverse plane as they were incompletely recovered before the RF excitation pulse
Vectors are pushed beyond the transverse plane as they were incompletely recovered before the RF excitation pulse
68
T/F: nhomogeneities are areas within the magnetic field that do not exactly match the external magnetic field strength.
True
69
What does the phase encoding gradient do? Encode data along the LONG axis of the anatomy Encode data along the DIAGONAL axis of the anatomy Encode data along the SHORT axis of the anatomy Both the DIAGONAL and SHORT axes
Encode data along the SHORT axis of the anatomy
70
Which gradient performs slice selection for coronal slices? X gradient Y gradient Z gradient Any combination of gradients
Y gradient
71
Which gradient performs phase encoding for coronal slices? X gradient Y gradient Z gradient Any combination of gradients
X gradient
72
When is the slice select gradient switched on during a spin echo pulse sequence? During the rephasing pulse only During the 90 and 180 degree pulses During the excitation pulse only It is always on during the scan
During the 90 and 180 degree pulses
73
To achieve thin slices, do we need a STEEP or SHALLOW slice select slope? STEEP SHALLOW
STEEP
74
To achieve THIN slices would we use a narrow or broad transmit bandwidth? Narrow Broad
Narrow
75
When is the phase encoding gradient switched on in a pulse sequence? Before the 90 degree excitation pulse Usually just before the application of the 180 degree rephasing pulse Between TRs During the spin echo
Usually just before the application of the 180 degree rephasing pulse
76
What does the slope of the frequency encoding gradient determine? (Select all that apply) The size of the anatomy covered along the frequency encoding axis The 180 degree rephasing pulse The FOV The size of the phase matrix
The size of the anatomy covered along the frequency encoding axis The FOV
77
What is another term for the duration of the readout gradient? X gradient Sampling time Z gradient Window of opportunity
Sampling time
78
How often must we sample frequencies, according to the Nyquist Theorem? Once per cycle Twice per cycle Three times per cycle Less than once per cycle
Twice per cycle
79
What artifact occurs if we sample frequencies less than once per cycle? Zipper Cross talk Cross excitation Aliasing
Aliasing
80
What is the unit of K space? Inches Radians per meter Radians per centimeter Degrees
Radians per centimeter
81
T/F: The phase axis of K space is vertical and is centered in the middle of K space perpendicular to the frequency axis.
True
82
Which part of K space do steep phase gradients fill? Outer lines of K space, positive only Outer lines of K space, both positive and negative Central lines of K space, positive only Central lines of K space, both positive and negative
Outer lines of K space, both positive and negative
83
Which part of K space do shallow phase gradients fill? Outer lines of K space, positive only Outer lines of K space, both positive and negative Central lines of K space, positive only Central lines of K space, both positive and negative
Central lines of K space, both positive and negative
84
T/F: K space is not the image
True
85
Which gradient locates signal along the LONG axis of the anatomy? Phase encoding gradient Frequency encoding gradient Slice select gradient Isocenter gradient
Frequency encoding gradient
86
T/F: The scan time is the time to fill K space.
True
87
T/F: Data in K space are symmetrical.
True
88
What kind of information is contributed by data collected in the central lines of K space? (Select all that apply) Resolution Signal Contrast *weighting
Signal | Contrast
89
What kind of information is contributed by data collected in the outer lines of K space? Resolution Signal Contrast *weighting
Resolution
90
What controls the polarity of a gradient? The steepness of the slice select gradient The sharpness of the frequency readout The direction of the current through the gradient coil The strength of the main magnetic field
The direction of the current through the gradient coil
91
T/F: Phase encoding gradient positive fills the bottom half of K space.
False
92
Which gradient performs slice selection for sagital slices? X gradient Y gradient Z gradient Any combination of gradients
X gradient
93
What is another name for the frequency encoding gradient? Phase encoding gradient Short gradient Readout gradient Rephasing gradient
Readout gradient
94
Where are the data points collected by the system stored? FFT K space Frequency domain 42.57 MHz
K space
95
During a typical spin echo sequence, there are three gradient applications. In chronological order they are: Phase encoding, Slice select, Frequency encoding Slice select, Phase encoding, Frequency encoding Frequency encoding, Phase encoding, Slice select Slice select, Frequency encoding, Phase encoding
Slice select, Phase encoding, Frequency encoding
96
Which gradient lies along the horizontal axis of a typical superconducting MRI system? X gradient Y gradient Z gradient W gradient
X gradient
97
If a 512 phase matrix is selected, which k-space lines are likely to be filled in a normal acquisition? 0 to 512 256 to -256 0 to -512 256, 0 to -255
256 to -256
98
Which of the following statements is true? A data point contains frequency encoding information only A data point contains phase encoding information only A data point contains both phase and frequency information A data point represents a specific pixel on the image
A data point contains both phase and frequency information
99
T/F: A thicker slice has a better SNR than a thinner slice.
True
100
Decreasing receive bandwidth by half increases the SNR by how much? 60% 40% 1/2 .22
40%
101
T/F: A fine matrix is one with a low number of frequency encodings and/or phase encodings, and results in a small number of pixels in the FOV.
False
102
# Choose three factors that affect the SNR from the list below (there are three factors). Magnetic field strength of the system Proton density of the area under examination The patient's height and weight Receive bandwidth
Magnetic field strength of the system Proton density of the area under examination Receive bandwidth
103
T/F: A long TE increases SNR and a short TE decreases SNR.
False
104
T/F: TR controls the amount of T2 weighting.
False
105
# Choose the term that best describes "a unit volume of patient tissue" from the list below. Pixel Grayscale value Voxel Slice thickness
Voxel
106
Which of the following factors affect scan time during a sequential acquisition? (choose all that apply) ``` TR TE Phase Matrix NEX Frequency Matrix ```
TR Phase Matrix NEX
107
Which of the following factors affect voxel size (choose all that apply)? Slice thickness Number of pixels or matrix Size of the flip angle FOV
Slice thickness Number of pixels or matrix FOV
108
T/F: Square pixels ALWAYS provide better spatial resolution than rectangular pixels.
True
109
If we double the NEX, what does that do to our scan time? Reduces the scan time by 1/2 Doubles the scan time Reduces the scan time by 1.4 Increases the scan time by 1.4
Doubles the scan time
110
Decreasing the phase matrix does which of the following? Decreases SNR Increases spatial resolution Makes the scan shorter Has no effect on scanning
Makes the scan shorter
111
Of the following parameters, which would give the best spatial resolution? 256 × 256, 3 mm slice thickness, 12 cm FOV, 1 NEX 256 × 128, 8 mm slice thickness, 40 cm FOV, 4 NEX 512 × 256, 6 mm slice thickness, 14 cm FOV, 2 NEX None of the above would give good spatial resolution.
256 × 256, 3 mm slice thickness, 12 cm FOV, 1 NEX
112
Of the following parameters, which would give the highest SNR? 256 × 256, 3 mm slice thickness, 12 cm FOV, 1 NEX 256 × 128, 8 mm slice thickness, 40 cm FOV, 4 NEX 512 × 256, 6 mm slice thickness, 14 cm FOV, 2 NEX None of the above.
256 × 128, 8 mm slice thickness, 40 cm FOV, 4 NEX
113
Increasing the field strength increases SNR because: The flip angle increases The NMV increases in size because the high energy population of spins increases The precessional frequency increases The NMV increases in size because the low energy population of spins increases
The NMV increases in size because the low energy population of spins increases
114
Halving a square FOV reduces the SNR by: 1/2 1/4 3/4 1/8
3/4
115
Which of the following does NOT occur when the receive bandwidth is reduced? SNR increases Chemical shift artifact improves Number of slices available decreases TE increases
Chemical shift artifact improves
116
Doubling the number of signal averages changes the SNR by how much? × 2 1/2 22 √2
√2
117
Which of these is a consequence of decreasing the TR from 2500 ms to 500 ms? Scan time decreases T1 weighting decreases Spatial resolution decreases Slice number available increases
Scan time decreases
118
Do we achieve greater SNR with a long or a short TR? long short
long
119
Pixel area = __ / matrix size. SNR Resolution FOV GNR
FOV
120
Which sequence is considered the gold standard for MR imaging? SSFP Spin Echo Gradient Echo Coherent Gradient Echo
Spin Echo
121
What is the approximate T1 relaxation time of CSF at 1T? 2000 ms 150 ms 500 ms 3000 ms
2000 ms
122
What is the approximate T1 relaxation time of fat at 1T? 2000 ms 200 ms 500 ms 3000 ms
200 ms
123
What do T1 images typically best demonstrate? Anatomy Pathology Bright fluid Everything
Anatomy
124
What does STIR stand for? Short ___ Inversion Recovery.
Tau tau TAU
125
When would we be most likely to use a STIR sequence? CSF flow studies MSK studies (i.e. knee, wrist) Head or Spine studies (to look for periventricular lesions) All studies must include a STIR sequence
MSK studies (i.e. knee, wrist)
126
T/F: FLAIR stands for Fluid Low Attenuation Inversion Recovery
False
127
When would we be most likely to use a FLAIR sequence? MSK studies (i.e. knee, wrist) Head or spine studies (to look for periventricular lesions) All studies must include a FLAIR sequence To image the great vessels
Head or spine studies (to look for periventricular lesions)
128
How is time reduced in fast spin echo? (Choose all that apply) The operator must prescribe fewer slices Use a turbo factor Use thinner slices Use an echo train
Use a turbo factor Use an echo train
129
What artifact can occur as a result of using long echo trains? Cross talk Truncation Image Blurring Aliasing
Image Blurring
130
An inversion recovery pulse sequence begins with which of the following? 90 degree RF pulse 180 degree inverting pulse 180 rephasing pulse Gradient pulse (variable flip angle)
180 degree inverting pulse
131
Which Fast Inversion Recovery sequence(s) do we often use? STIR SSFP FLAIR Incoherent Gradient Echo
STIR | FLAIR
132
What is Double IR prep used to accomplish? Used to null blood for cardiac imaging Flips the NMV back to zero Imaging large patients FSE sequences
Used to null blood for cardiac imaging
133
A coherent gradient echo sequence samples which of the following? FID Only Stimulated Echo Only Both the FID and Stimulated Echo T2 Information Only
Both the FID and Stimulated Echo
134
T/F: During EPI all the lines of K space are filled during one repetition.
True
135
How is K space filled in Fast Gradient Echo Sequences? (choose all that apply) Elliptical filling Centric filling Keyhole filling Outside to inside
Centric filling | Keyhole filling
136
In which of the following sequences is all the transverse magnetization rewound at the end of each TR period? Coherent gradient echo Incoherent gradient echo EPI
Coherent gradient echo
137
Which of the following statements is true in an incoherent gradient echo sequence? Residual transverse magnetization has the same phase as the magnetization most recently created The flip angle is always 10° The system discriminates between residual and the most recently created transverse magnetization because it has a different phase value It is not a steady state sequence
The system discriminates between residual and the most recently created transverse magnetization because it has a different phase value
138
What type of echoes are sampled during SS-FSE? Spin echoes Gradient echoes Incoherent Gradient echoes
Spin echoes
139
What are the advantages of propeller K Space filling? Better detail Low SNR but high resolution Bright signal Rapid imaging, increased SNR and CNR, reduced motion artifact.
Rapid imaging, increased SNR and CNR, reduced motion artifact.
140
T/F: Using a long echo train decreases the SAR.
False
141
T/F: STIR sequences should NOT be used in conjunction with contrast enhancement.
True
142
T/F: Spin echo pulse sequences are classified according to whether the residual transverse magnetization is in phase (coherent) or out of phase (incoherent).
False
143
SSFP sequences usually produce images that are ___ weighted. T1 T2 T2* T1 or T2*
T2
144
T/F: Parallel imaging is usually used to increase resolution or decrease scan time.
True
145
Which artifact is produced as a result of motion of flowing nuclei? Zipper effect Aliasing Phase ghosting Truncation
Phase ghosting
146
Which type of motion can we compensate for in MRI? First order laminar flow Second order motion acceleration Third order motion jerk All of those listed can be compensated for
First order laminar flow
147
T/F: TOF phenomenon results in a dark vessel.
True
148
Which of the following effects does TOF depend on? (Select all that apply) Velocity of flow TE Slice Thickness TR
Velocity of flow TE Slice Thickness
149
# Choose the two main uses of spatial pre-saturation from the list below. (Select all that apply) Chemical Pre-saturation STIR FLAIR SPIR
Chemical Pre-saturation | SPIR
150
T/F: Counter current flow occurs when flowing nuclei are traveling in the same direction as slice selection.
False
151
Which of the following is/are result(s) of time of flight? (Select all that apply) Flow-related enhancement Entry slice phenomenon High velocity signal void Zipper effect
Flow-related enhancement High velocity signal void
152
In a dual echo sequence, which of the following combinations would result in reduced flow artifact in the second echo? TE 20 and 50 ms TE 20 and 70 ms TE 40 and 80 ms TE 40 and 95 ms
TE 40 and 80 ms
153
What is the chemical shift at 1.5 T? 2.2 Hz 22 Hz 220 Hz 220KHz
220Hz
154
T/F: STIR is most useful in post contrast imaging.
False
155
T/F: Even echo rephasing reduces intra-voxel dephasing.
True
156
T/F: Intra-voxel dephasing results in an increase in total signal amplitude from the voxel.
False
157
Stationary nuclei always receive both excitation and rephasing pulses, but flowing nuclei present in the slice for the excitation may have exited the slice before rephasing. This is called ______________. Time of Flight Phenomenon Entry Slice Phenomenon Intra-Voxel Dephasing Turbulent Flow Effects
Time of Flight Phenomenon
158
T/F: As velocity of flow decreases, the time of flight effect decreases.
True
159
T/F: Another term for time of flight is inflow effect.
False
160
Determine which of the following parameters would result in the most entry slice phenomenon (i.e. entry slice phenomenon would increase compared to the other options). Slice 1 of 20, 3000 TR, .7 mm slice thickness Slice 1 out of 20, 400 TR, 1.2 mm slice thickness Slice 10 out of 20, 3000 TR, 1.2 mm slice thickness Slice 10 out of 20, 400 TR, .7 mm slice thickness
Slice 1 of 20, 3000 TR, .7 mm slice thickness
161
T/F: Flow affects image quality.
True
162
The frequency difference between fat and water is called ___________ and can be used to specifically null the signal from either fat or water. Chemical Pre-Saturation Chemical Shift Precessional Frequency Even Echo Rephasing
Chemical Shift
163
T/F: Any tissue can be nulled as long as an RF pulse matching its precessional frequency is applied to the imaging volume before excitation.
True
164
T/F: Pre-saturation techniques increase the RF deposition to the patient.
True
165
What is another term for phase ghosting? Phase wrap Phase mismapping Chemical misregistraion Aliasing
Phase mismapping
166
What is another term for phase ghosting? Phase wrap Phase mismapping Chemical misregistraion Aliasing
Phase mismapping
167
In which direction does phase ghosting occur? (Choose all applicable) Phase Axis Frequency Axis Slice Select Axis Anterior to Posterior
Phase Axis
168
Which of the following are methods of reducing phase ghosting? (Choose all that apply) Using respiratory compensation techniques Decrease the TE Carefully explain the procedure to the patient Swap phase and frequency Use pre-saturation pulses Use "no phase wrap" software
Using respiratory compensation techniques Carefully explain the procedure to the patient Swap phase and frequency Use pre-saturation pulses
169
In which direction does aliasing occur? (Choose all that apply) Phase Axis Frequency Axis Slice Select Axis Anterior to Posterior
Phase Axis Frequency Axis
170
Which of the following is a remedy for aliasing? Decreasing the slice thickness Decreasing the size of the FOV Decreasing the TE Using "no phase wrap" software
Using "no phase wrap" software
171
Which of the following best describes the appearance of chemical shift artifact on an image? Extra information wraps into the image The cancellation of fat and water signals when they are out of phase. Banding artifact at the interfaces of high and low signal. Produces a dark edge at the interface between fat and water.
Produces a dark edge at the interface between fat and water.
172
In which direction does chemical shift artifact occur? (Choose all that apply) Phase Axis Frequency Axis Slice Select Axis Anterior to Posterior
Frequency Axis
173
What unit is used to express the amount of chemical shift? MHz Hz ppm gyromagnetic ratio
ppm
174
What is another name for a truncation artifact in a sagital cervical spine? Truncation Gyromagnetic Artifact Gibbs Faraday
Gibbs
175
How can we reduce truncation? Increase the TE Increase the TR Increase the number of phase encoding steps Increase the number of frequency encoding steps
Increase the number of phase encoding steps
176
In which direction does truncation occur? (Choose all that apply) Phase Axis Frequency Axis Slice Select Axis Anterior to Posterior
Phase Axis
177
Which best describes the appearance of a zipper artifact? Distortion of the image together with large signal voids. Anatomy folded into the FOV A dense line on the image at a specific point. Tissues return a high signal intensity
A dense line on the image at a specific point.
178
Which best describes the appearance of a zipper artifact? Distortion of the image together with large signal voids. Anatomy folded into the FOV A dense line on the image at a specific point. Tissues return a high signal intensity
A dense line on the image at a specific point.
179
Which of the following is a possible cause of shading artifact? A hemorrhage A leak in the RF shielding of the scan room Spin echo sequences Inhomogeneities in the main magnetic field
Inhomogeneities in the main magnetic field
180
Which of the following best describes Moire artifact? Black and white banding artifact on the edge of the FOV Fat and water are out of phase and their signals cancel each other out. Different chemical environments of fat and water produce a dark edge at the interface between fat and water. Anatomy from outside the FOV are wrapped into the image.
Black and white banding artifact on the edge of the FOV
181
T/F: To reduce chemical shift artifact, ALWAYS use the widest receive bandwidth in keeping with good SNR and the smallest FOV possible.
True
182
T/F: To reduce chemical shift artifact, ALWAYS use the widest receive bandwidth in keeping with good SNR and the smallest FOV possible.
True
183
In spin echo sequences using a pre-saturation pulse, what would persistent bright signal within the vessel indicate? Slow Flowing Blood Fast Flowing Blood Blood Clot Vascular Occlusion
Slow Flowing Blood Blood Clot Vascular Occlusion
184
Which of the following best describes what the "P" wave represents? Atrial Systole (Contraction) Ventricular Systole Ventricular Diastole None of the above
Atrial Systole (Contraction)
185
Which of the following best describes what the "QRS" complex represents? Atrial Systole (Contraction) Ventricular Systole Ventricular Diastole (Relaxation) None of the above
Ventricular Systole
186
Which of the following best describes what the "T" wave represents? Atrial Systole (Contraction) Ventricular Systole Ventricular Diastole (Relaxation) None of the above
Ventricular Diastole (Relaxation)
187
Which of the following sequence(s)/technique(s) can be used to make vessels appear bright? Gradient Echo Imaging Gradient Moment Rephasing Contrast Enhancement All of the above
All of the above
188
Which of the following is used to trigger each pulse sequence when using ECG gating? P wave Q wave R wave S wave
R wave
189
How much does a typical patient's heart rate vary during an exam? 10- 20 % 20- 30% 40- 50% 60- 70%
10- 20 %
190
If the R to R interval is 500 ms, the trigger window is 10%, and the trigger delay is 50 ms, what is the time available to acquire data? 310 ms 400 ms 460 ms 500 ms
400 ms
191
If the R to R interval is 500 ms, the trigger window is 10%, and the trigger delay is 50 ms, what is the time available to acquire data? 310 ms 400 ms 460 ms 500 ms
400 ms
192
What determines the length of the R to R interval? The technologist performing the exam The radiologist who establishes the protocol The patient's heart rate The type of sequence
The patient's heart rate
193
What is the waiting period before each R wave called? Trigger Window Trigger Delay R to R Interval Available Imaging Time
Trigger Window
194
T/F: ECG gating is prospective.
True
195
When using MRA to evaluate intracranial vascularity, flow within smaller vessels can best be demonstrated by: 2D Time of Flight MRA 3D Time of Flight MRA 3D Phase Contrast MRA All of the Above
3D Time of Flight MRA
196
When using MRA to evaluate intracranial vascularity, flow within smaller vessels can best be demonstrated by: 2D Time of Flight MRA 3D Time of Flight MRA 3D Phase Contrast MRA All of the Above
3D Time of Flight MRA
197
MRA techniques include all of the following EXCEPT: TOF-MRA PC-MRA CE-MRA CT-MRA
CT-MRA
198
Which of the following does NOT reduce flow motion artifact? SAT ECG gating GMN Subtraction
Subtraction
199
Which of the following can be used to improve coverage and at the same time maintain the signal from blood flowing within the larger volume? MC-TOF PC-MRA MOTSA Phase images
MOTSA
200
The vascular signal, produced on time of flight MRA (TOF-MRA), relies on: MOTA PC Flow related enhancement Velocity-induced phase shifts
Flow related enhancement
201
The vascular signal, produced on PC-MRA, relies on: MOTA PC Flow related enhancement Velocity-induced phase shifts
Velocity-induced phase shifts
202
Contrast enhanced MRAs are acquired with all of the following EXCEPT: Test bolus Bolus tracking VENC settings Fluoro triggering
VENC settings
203
Contrast enhanced MRAs are acquired with all of the following EXCEPT: Test bolus Bolus tracking VENC settings Fluoro triggering
VENC settings
204
Multiphase images, acquired with additional modulation of magnetization, are known as: TOF-MRA PC-MRA VENC SPAMM
SPAMM
205
What is the available imaging time using these parameters? R to R 800 ms, trigger window 10%, delay after trigger 4 ms. 688 ms 700 ms 716 ms 800 ms
716 ms
206
When using MRA to evaluate peripheral vascular flow, such as that within the arteries of the legs, saturation pulses are: Placed superior to the acquired slices Placed in the acquired slices Placed inferior to the acquired slices Not necessary
Placed inferior to the acquired slices
207
When using MRA to evaluate extracranial vascular flow, such as that within common carotid arteries, a recommended technique is: 2D time of flight MRA 3D time of flight MRA 3D phase contrast MRA All of the above
2D time of flight MRA
208
When using MRA to evaluate extracranial vascular flow, such as that within common carotid arteries, a recommended technique is: 2D time of flight MRA 3D time of flight MRA 3D phase contrast MRA All of the above
2D time of flight MRA
209
Magnetic field strength can be expressed in units of Tesla (T) or gauss (G), where: 1 T = 1 G 1. 5 T = 1500 G 3. 0 T = 0.3 G 1 T = 10 000 G
1 T = 10 000 G
210
The homogeneity of a magnetic field is expressed in units of: ppm (parts per million) T (Tesla) G (gauss) MHz (megaHertz)
ppm (parts per million)
211
The homogeneity of a magnetic field is expressed in units of: ppm (parts per million) T (Tesla) G (gauss) MHz (megaHertz)
ppm (parts per million)
212
Gradient strength is expressed in units of all of the following EXCEPT: w/kg mT/m G/cm T/m/s
w/kg
213
Gradient strength is expressed in units of all of the following EXCEPT: w/kg mT/m G/cm T/m/s
w/kg
214
Passive magnetic shielding can be accomplished by lining the scan room walls with: Copper Lead Steel Shim plates
Steel
215
Fourier transform (FT) is a mathematical process that is used to convert MR signal from a FID into a spectrum. FT occurs in the: Image processor Array processor RF coil Gradient amplifier
Array processor
216
What rule determines the direction of current flow through a coil, and the direction of the magnetic field that the current produces? The size of the magnetic field Right hand thumb rule MRI Fringe Field Rule Whether you are using an open or closed configuration
Right hand thumb rule
217
What is the maximum field strength allowed for clinical imaging of adults in the USA? 3 T 4 T 8 T 10 T
8 T
218
T/F: The purpose of magnetic field shimming is to reduce the size of the fringe field.
False
219
What is the most common material used to make permanent magnets? Titanium Helium Iron Alnico
Alnico
220
Ohm's law states: B = k (l) V = I (R) B = H (1 x) 1 T = 10 kG
V = I (R)
221
T/F: The law of electromagnetic induction indicates that whenever a charge particle such as an electron exhibits motion, a corresponding magnetic field is induced.
True
222
Which cryogen is commonly used to super cool wires in MRI? Liquid helium Niobium Alnico Hydrogen
Liquid helium
223
Which material is most commonly used to make superconducting wires? Liquid helium Niobium-titanium alloy Alnico Hydrogen
Niobium-titanium alloy
224
Why do we use shimming? To apply a magnetic field To make the field even or homogenous To make the magnetic field smaller To induce a current
To make the field even or homogenous
225
Select all of the following that determine the strength of an electromagnet: The amount of current passed through the windings The number of windings in the coil The distance or spacing between the windings The diameter of the wire used in the windings
All of the above
226
In a typical superconducting magnet, the direction of the magnetic field is: Vertical Horizontal Around the flux lines Hard to measure
Horizontal
227
T/F: The SNR is enhanced when using local coils.
True
228
T/F: The SNR is enhanced when using local coils.
True
229
What does BOLD stand for? Blood oxygenation level dependent Blood oxygen line developed Body oriented level dependent Type of CT scanning
Blood oxygenation level dependent
230
Functional imaging techniques include which of the following (check all that apply)? DWI fMRI Spectroscopy Perfusion imaging
All of the above
231
T/F: Functional imaging techniques allow MRI to be used to assess function and physiology as opposed to merely conventional structural imaging.
True
232
What term is used to describe the net displacement of diffusion spins across a tissue? SWI fMRI ADC MRS
ADC
233
What is the extrinsic parameter that controls the amount of diffusion seen in an image? TR TE flip angle b factor
b factor
234
What signal intensity is an area of abnormality likely to have on an ADC map? Hypointense Hyperintense Isointense
Hypointense
235
What signal intensity is an area of abnormality likely to have on a trace image? Hypointense Hyperintense Isointense
Hyperintense
236
T/F: Perfusion is the regional blood flow in tissues and is defined as the volume of blood that flows into one gram of tissue.
True
237
Which of the following best describes a use for MR Spectroscopy (MRS)? To show ankle anatomy To aid in brain tumor staging To demonstrate perfusion
To aid in brain tumor staging
238
T/F: Interventional MRI includes performing biopsies in the MR suite.
True
239
Chemical shift is measured in which of the following units? mm/s ppm T/m G/cm
ppm
240
T/F: Perfusion is measured using MRI by tagging the fat in arterial blood during image acquisition.
False
241
The anatomy of white matter tracts can be mapped using strong multi-directional gradients in which of the following? SWI diffusion tensor imaging (DTI) continuous arterial spin labeling (CASL) CBV
diffusion tensor imaging (DTI)
242
T/F: MR microscopy (MRM) is useful as a research tool when studying detail in a very small area of tissue.
True
243
T/F: MR microscopy (MRM) is useful as a research tool when studying detail in a very small area of tissue.
True
244
T/F: Gadolinium chelates are ferromagnetic.
False
245
After gadolinium administration, it is recommended that which images are acquired? One T1 weighted image One T2 weighted image Multiple T1 weighted images Multiple T2 weighted images
Multiple T1 weighted images
246
Enhanced MR images are evaluated for sensitivity and specificity, where sensitivity is ________ and specificity is ________. The ability to detect the lesion/the ability to classify the lesion The ability to classify the lesion/the ability to detect the lesion TE used to evaluate the lesion/TR used to evaluate the lesion TR used to evaluate the lesion/TE used to evaluate the lesion
The ability to detect the lesion/the ability to classify the lesion
247
T/F: All gadolinium contrast agents have the same ‘relaxivity’.
False
248
# Choose all of the following that enhancement agents can be useful in detecting. Infection Tumors Infarction Inflammation
All of the above
249
On a T1 weighted pre-contrast image, fat appears _____ and water appears _____. Bright, Dark Bright, Bright Dark, Bright Dark, Dark
Bright, Dark
250
T/F: Scar tissue enhances in the lumbar spine shortly after the injection of contrast.
True
251
T/F: Scar tissue enhances in the lumbar spine shortly after the injection of contrast.
True
252
Which sequence would be used to best visualize metastatic lesions in the bone? FLAIR with contrast Fat saturation techniques with contrast T2 weighted sequences Non-contrast imaging
Fat saturation techniques with contrast
253
T/F: Gadolinium cannot cross the intact blood-brain barrier (BBB).
True
254
Gadolinium element is ________ and gadolinium chelate contrast media is ________. Ferromagnetic/paramagnetic Ferromagnetic/ferromagnetic Paramagnetic/ferromagnetic Paramagnetic/paramagnetic
Ferromagnetic/paramagnetic
255
T/F: When using gadolinium, the T1 relaxation time of the water is reduced so enhancing lesions (such as tumors associated with free water) appear bright on a T1 weighted image.
True
256
Select all of the following that were discussed as possible MR contrast agents in the text. Gadolinium Blueberry juice Mango juice Dilute barium solutions
All of the above