Modes of Mechanical Ventilation Flashcards
2 types of pressure measurements
peak inspiratory pressure
plateau pressure
Peak inspiratory pressure
total pressure required to distend lungs and airways
pressure used to calculate dynamic compliance (intrinsic and extrinsic factors)
plateau pressure
distending pressure to expand only lungs (intrinsic)
measures redistribution of air flow through lungs
Pplat is used to calculate static compliance
when can you measure Pplat
only in volume control during inspiratory hold
4 parts of each breath
start of inspiration (trigger variable) inspiration itself (limit or target variable) end of inspiration (cycling variable, transition to expiration) expiration (baseline variable, end expiration, PEEP)
3 components of a breath
ti=inspiratory time
te=expiratory time
TCT=total cycle time
4 types of trigger variables
pressure
volume
flow
time
pressure trigger variable
decrease in circuit stimulates ventilator to deliver breath
volume
volume change in circuit can stimulate ventilator to deliver breath
flow
change of flow in circuit stimulates ventilator to deliver
time
set time interval triggers ventilator to deliver breath (set by RR). this occurs independent of patient effort
limit variable
controls how an inspiratory breath is maintained once threshold is reached variable will not exceed set limit. this does NOT cause termination of inspiration
pressure as limit variable
sets upper pressure limit that cannot be exceeded
volume as limit variable
set upper volume limit that cannot be exceeded
flow as limit variable
sets maximum airflow that cannot be exceeded
cycling variable
transition from inspiration to expiration. based on either volume, pressure, flow, or time
volume cycling variable
ventilator delivers flow until set volume achieved. if inspiratory pause set (typically 10-20%) this variable changes to a time based cycling variable
pressure cycling variable
once pressure is achieved, flow will transition to expiration
flow cycling variable
once inspiratory flow drops below set threshold (default at 25%), ventilator will transition to expiration. noted in pressure support ventilation mode
time cycling variable
ventilator terminated inspiratory breath after predetermined inspiratory time has been delivered
intrinsic peep
secondary to incomplete expiration, referred to as auto peep (related to air trapping, MV or RR maybe too high)
extrinsic peep
provided by a mechanical ventilator, referred to as applied peep
auto peep causes
high MV, expiratory flow limitation, expiratory resistance
volume control ventilation
trigger variable
limit variable
cycling variable
delivers set TV at set RR
time
volume
flow