Nuclear medicine II Flashcards

1
Q

Primary interaction NaI in Tc99 imaging

A

Photoelectric (High Z material)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Energy resolution for Tc 99 m on Gamma camera

A

10 % (14%, because the FHWM)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Energy resolution

A

Photopeak broadening as a percentage of photopeak energy

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

If energy resolution is 10% (10% FWHM) what is the energy window

A

20%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q
A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

What is the sensitivity range of well counter (contamination testing)

A

nCi

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Pulse height analyzer window width

A

20%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What detectors DOES have gas and DOES NOT have NaI

A
Dose calibrators (before injection measure). 
- Provide radioisotope info
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What is ACLG used for QA

A

ACLG is used for dose calibrators

  • A: Accuracy annual
  • C: Constancy daily
  • L: Linearity quaterly
  • G: Geometry Installation

Dose calibrators have a high accuracy (injection within 20%)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Medium energy collimators

A

Ga 67
K 81
In 111

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

High energy collimators

A

I-131

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Accuracy required for dose injected (NRC comission)

A

20%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Low energy collimator

A

Tc99 m and I123

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Intrinsic

A

withou collimator (INdependent)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

extrinsic

A

with collimator (EXtra)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

What is this artifact?

A

Craked crystasl

17
Q

What is this

A

Photomultiplier not working

18
Q

What is this artifact

A

Ununiformity (captured using flood field)

Poor uniformity calibration

19
Q

Whatis this

A

Septal penetration

20
Q

How to decrease resolution of nM

A

Increase parallel hole colimator - patient distance

21
Q

Mottle of 100 counts/pixel in uniformioty flood?

A

10% (Mottle=1/square root of dose to receptor)

22
Q

Extrinsic system resolution of NM gamma camera

A

8 mm

23
Q

What is

A

Change energy window

24
Q

Pinhole

A

Thyroid

25
Q

Transmutation

A