Pacemakers Flashcards
(43 cards)
Indications for temporary pacemakers
- Symptomatic bradycardia unresponsive to medical therapy
– 2nd Degree Type II and 3rd Degree Heart Block
– Permanent pacemaker failure
– Support cardiac output after cardiac surgery
– Tachy dysrhythmias (overdrive pacing)
– Diagnostic studies: Electrophysiology studies (EPS)
Permanent pacemaker indications
– Sick Sinus Node Syndrome
– Symptomatic AV block
– Pan-conduction disease (often age related conduction tissue changes)
– Hereditary, idiopathic, or drug-related QT interval prolongation
Permanent ventricular lead placement
The ventricular lead is usually placed in the apex of the right ventricle and held in
with tines at the end to lead.
Permanent Atrial Lead Placement
The atrial lead is harder to keep in place due to the smooth surface, therefore often a lead is used with a screw on the end to fix it within the atria
Some leads emit steroids to reduce inflammation and scarring
Pacemaker function: Pace
Electrical stimulus to the ventricle or atria
– Capture is the effective depolarization of the heart by the artificial pacemaker
– Capture threshold is the minimum pacemaker output setting required to pace the heart
Pacemaker function: Sense
The pacemaker wire senses electrical activity in the chamber in which it is placed
Pacemaker function: Pacing triggered
The pacemaker fires an electrical stimulus if it senses no electrical activity
Pacemaker function: Pacing inhibited
The pacemaker inhibits electrical stimulus due to sensed intrinsic rhythm
Pacemakers deliver a pacing stimulus only when…?
The hearts intrinsic pacemaker fails to function
VVI
- The ventricular lead paces the ventricle and inhibits itself if it senses an intrinsic rhythm
- The patient loses atrial filling and atrial kick (loss of CO)
DDD
– 1st D = both chambers are paced
– 2nd D = both chambers are sensed
– 3rd D = inhibits both atrial and ventricular pacing when sensing intrinsic rhythm; and triggers response to sensed atrial activity to allow for rate responsive pacing
DDD: Atrial sensed with ventricular paced rhythm
– The SA node is intact and pacing normally
– AV synchrony remains intact
DDD: Atrial Paced with intrinsic ventricular rhythm
- This strip shows loss of intrinsic SA node activity, the atrial lead then paces the atria
- The impulse travels normally through the AV node
- There is intrinsic BBB
Limitations for DDD pacing
Atrial pacing cannot be achieved with chronic atrial fib
If a patient with a DDD pacer goes into Afib there are rate limitations set for the atrial and ventricular responses
Pacemaker Codes
- Chamber Paced
- Chamber Sensed
- Response to Sensing
- Rate ResponsiveP
Common occurrence: Fusion Beats
Causes
– A paced and an intrinsic ventricular beat occur simultaneously
– Resulting QRS a mix between intrinsic and paced
Intervention : none
Common occurrences: PVC
Causes
– electrode causing irritable ventricular focus
– Note: PVC’s normally occur in the first 24h after lead placement
Intervention
– May need to administer antiarrhythmic drugs – May need reposition of lead wire
Complication: Battery Failure
Applying a magnet over the pacer will illicit fixed (asynchronous) pacing
If the battery is failing, the pacer will pace at a different rate than what’s programmed
If the battery of a DDD pacer begins to fail it will convert to VVI to conserve battery time
Complication: Loss of output
Possible Causes – Battery depletion – Pacemakeroff
– Oversensing
– Faulty cable connection
– Dislodged/fracturedlead
Complication: Failure to capture
The pacemaker spikes fail to initiate the depolarization of the ventricle
– EKG pacemaker spikes not followed by QRS complexes
S/S Failure to capture
Signs and Symptoms – bradycardia
– hypotension
– fatigue
Failure to capture: Potential causes
– Displacement of the pacing lead wire (most common) – mA set too low – fracture of the lead – battery failure – perforation of the myocardium by the lead wire – edema or scar tissue formation at electrode tip – fibrin formation at the lead tip
Failure to capture interventions
Intervention:
– Notify physician immediately
– The physician may need to reprogram the pacer, replace or reposition the pacing wire
– May reposition patient onto left side
Less Common Causes of Noncapture May Include:
Twiddler’s syndrome Electrolyte abnormalities – e.g.,
hyperkalemia Myocardial infarction Drug therapy Battery depletion Exit block