PHY 2 Final Prep Flashcards
(94 cards)
If the relative output power of an ultrasound instrument is calibrated in decibels and the operator increases the output by 20 dB, what would the beam intensity be increased by?
One hundred times.
(A 10 dB increase = ×10 increase in intensity
A 20 dB increase = ×10 ×10 = ×100 increase in intensity)
Refraction is a change in what parameter of the ultrasound beam?
Direction
A real-time instrument produces images at 30 frames per second. The maximum imaging depth is 15 cm.
The line density is constant. To increase the maximum imaging depth to 30 cm, you would change the frame rate to how many frames per second?
15 fps
(The system shows 30 images per second when looking 15 cm deep.
If you want to look twice as deep (30 cm), the system needs more time for the signal to go down and come back. Because of that, it can only show half as many images per second so it isn’t as much work on the machine to look deeper into the body.)
What is the relationship between pulse repetition frequency and imaging depth?
Inversely related
A 7.5 MHz transducer produces approximately what size wavelength in soft tissue?
0.2 mm
Wavelength =
Speed of sound in soft tissue ≈ 1540 m/s
Frequency = 7.5 MHz = 7,500,000 Hz
1540 / 7500000 =0.000205
When comparing gray scale and bistable imaging systems, what does a gray scale imaging system display that is not seen with a bistable system?
A range of reflection amplitudes.
(Bistable imaging shows only two brightness levels: black or white (on/off).
Gray scale imaging shows many levels of brightness, representing a range of echo strengths (amplitudes).
This allows gray scale systems to provide more detailed images compared to bistable systems.)
In general, a Doppler frequency shift could result from…
- A moving sound source and a stationary listener.
- A stationary sound source and a moving listener
- A moving sound source and a moving listener
(Note: could NOT result in a stationary sound source and a stationary listener)
How many shades of gray is a four bit digital scan converter memory capable of storing?
16
What technique is used to fill in empty data between scan lines?
Interpolation
What are three factors which could cause the presence of FALSE sludge/debris in a cystic mass?
- Reverberation artifact
- Scanning at high gain
- Side lobe artifact
(Note: NOT Speed of sound artifact)
A linear array transducer scans by generating lines of insonation from:
Different points, each line traveling parallel to each other
(A linear array transducer works by originating from different points along the transducer surface and
Traveling parallel (not angled) (straight line))
You are scanning the liver and the beam travels through a mass in the right lobe that has a propagation speed of 1.62 millimeter/microsecond. Where will the mass be placed on the screen?
Too anterior from its actual position.
(Since the sound travels faster than expected, the system thinks it took less time to reach the structure and return. But it doesn’t know the speed is faster—it assumes 1.54 mm/μs—so it underestimates the depth and the mass will be displayed closer to the surface (i.e., too shallow) on the screen than it actually is.)
There are several differences between pulsed and continuous wave Doppler. What is one true statement about continuous wave Doppler?
Continuous wave Doppler:
Simultaneously transmits and receives ultrasound.
What do interfaces that produce echoes depend on?
change in acoustic impedance
(The greater the mismatch in impedance between two tissues, the stronger the reflection (echo))
Aliasing occurs when the Doppler shift exceeds the:
Nyquist limit
Using color flow Doppler, if you steer the color window to the right or left, the frame rate will:
Not be changed.
You are scanning an interface at normal incidence in which the acoustic impedance is unchanged from one media into the other, but the acoustic velocity is 2 times higher in media one than it is in media 2. What likely will occur?
There will be no reflected sound.
What does Huygens Principle refer to?
Constructive and destructive interference of sound waved from multiple sources.
You are experiencing blossoming while scanning a patients carotid artery. This is making the vessel look as though it is larger than what it actually is. How do you eliminate the color Doppler artifact?
Turn the color Doppler gain down
In diagnostic medical sonography, the wavelength of a 2.25 MHz ultrasound beam is:
0.7 mm
(Wavelength = Propagation speed / Frequency)
(Frequency must be in Hz)
(2.25 MHz x 10^6 = 2,250,000 Hz)
(1540 m/s / 2250000 Hz = 0.7 mm)
Doppler power mode will display:
1.Doppler signal strength
2. Tissue perfusion
3. Presence of flow within small vessels
(Note: does NOT display flow velocity)
Spectral broadening will cause what change in a spectral analysis tracing?
Vertical thickening
The cosine of the angle in the Doppler forums refers to the angle between…
The beam transmitted and the direction of flow
During pulse-wave Doppler you send a 5MHz beam and then you received back a 4.89 MHZ beam. Which direction will the blood be moving?
Away from the probe
(Since the received frequency (4.89 MHz) is lower than the transmitted frequency (5 MHz), it indicates a negative Doppler shift, which means the blood is moving away from the probe.)