Pressure Pathology Flashcards
(89 cards)
What is a characteristic of a normal pressure waveform in a fluid-filled monitoring system?
A. Rounded contours with low-frequency oscillations
B. Sharp waveforms with visible high-frequency oscillations at low pressure
C. Overshooting waveforms at high pressure
D. Blunted upstroke with delayed deflection
B. Sharp waveforms with visible high-frequency oscillations at low pressure
What waveform feature suggests underdamping in a pressure monitoring system?
A. Dull, rounded waveform
B. Blunted signal with delayed peak
C. Narrow spikes or exaggerated overshoot
D. Absence of oscillations
C. Narrow spikes or exaggerated overshoot
What typically causes overdamping in a pressure transducer system?
A. A hypersensitive transducer
B. Increased arterial pressure
C. High-frequency filter settings
D. Problems in the fluid path or incorrect calibration
D. Problems in the fluid path or incorrect calibration
What happens when a transducer is too sensitive in a fluid-filled pressure system?
A. The signal becomes flat and unresponsive
B. A small pressure wave causes a large deflection
C. The system becomes non-functional
D. The waveform becomes more accurate
B. A small pressure wave causes a large deflection
Which of the following most likely indicates overdamping in a pressure tracing?
A. Rapid upstroke with multiple oscillations
B. Tall, narrow waveform peaks
C. Blunted, rounded waveform
D. High-frequency noise in the baseline
C. Blunted, rounded waveform
What effect does an air bubble in a left ventricular (LV) pressure line typically have on the pressure tracing?
A. It flattens the waveform
B. It causes delayed waveform peaks
C. It produces exaggerated systolic and diastolic overshoot
D. It eliminates the pressure signal
C. It produces exaggerated systolic and diastolic overshoot
What type of damping is associated with the presence of an air bubble in a pressure line?
A. Overdamping
B. Normal damping
C. Underdamping
D. Hyperdamping
C. Underdamping
What is one method used to correct an underdamped waveform caused by an air bubble?
A. Replacing the catheter
B. Increasing ECG sensitivity
C. Flushing the line and instilling diluted contrast media
D. Decreasing the flush rate
C. Flushing the line and instilling diluted contrast media
What waveform characteristic indicates a normal pressure tracing after flushing an air bubble from the line?
A. Tall and spiky waveform with excessive oscillations
B. Sharp, crisp upstroke with little systolic overshoot
C. Broad, flattened waveform with minimal signal change
D. High-frequency noise in the baseline
B. Sharp, crisp upstroke with little systolic overshoot
What does a “noisy” pulmonary capillary wedge pressure (PCWP) waveform typically suggest?
A. Overdamping
B. Catheter obstruction
C. Electrical interference
D. Underdamping
D. Underdamping
What is a common cause of an overdamped hemodynamic waveform?
A. Tubing that is too stiff
B. A hyperdynamic circulatory state
C. A bubble or clot in the transducer
D. Short tubing length
C. A bubble or clot in the transducer
Which of the following can reduce overdamping in a pressure monitoring system?
A. Using softer tubing
B. Increasing the length of the tubing
C. Replacing with stiffer tubing
D. Introducing a small air bubble
C. Replacing with stiffer tubing
What corrective action is recommended for underdamping due to a hyperdynamic state?
A. Shorten tubing
B. Increase filter on amplifier or introduce contrast media
C. Use stiffer tubing
D. Tighten catheter connection
B. Increase filter on amplifier or introduce contrast media
What should be done if the catheter is producing an underdamped waveform due to being placed in a turbulent jet?
A. Flush the system
B. Reposition the catheter
C. Replace the amplifier
D. Lengthen the tubing
B. Reposition the catheter
If there is a complete loss of signal in the monitoring system, which of the following is NOT a potential cause?
A. Bad cable
B. Catheter obstruction
C. Overdamping
D. Disconnected catheter
C. Overdamping
What should be done if pressures do not return to zero when expected?
A. Introduce a small bubble into the system
B. Recalibrate and check zero at mid chest
C. Switch to a softer tubing
D. Increase the internal diameter of the tubing
B. Recalibrate and check zero at mid chest
What can result from partial withdrawal of the pigtail catheter out of the LV during pressure measurement?
A. Increased LV-Ao gradient
B. Loss of all waveform data
C. Contamination of LV pressure by Ao pressure
D. Normal diastolic waveform with decreased upslope
C. Contamination of LV pressure by Ao pressure
What artifact is observed in the diastolic waveform when the pigtail catheter is malpositioned?
A. Diastolic plateau with no change
B. Continued decline (downslope) in diastolic pressure
C. Early systolic notch
D. Pulsus alternans
B. Continued decline (downslope) in diastolic pressure
What is the correct catheter position for accurate measurement of LV and Ao pressures across the aortic valve?
A. Just at the aortic root
B. With side holes above the Ao valve
C. All side holes under the Ao valve
D. In the descending aorta
C. All side holes under the Ao valve
What diagnostic error can result from failing to recognize catheter malposition during LV-Ao gradient measurement?
A. Overestimation of aortic valve disease
B. False diagnosis of ventricular septal defect
C. Underestimation of aortic valve stenosis
D. Missed detection of mitral regurgitation
C. Underestimation of aortic valve stenosis
How is the true LV-Ao gradient revealed after catheter malposition is corrected?
A. By decreasing amplifier sensitivity
B. By advancing the catheter slightly into the LV
C. By switching to a different transducer
D. By flushing the catheter with contrast
B. By advancing the catheter slightly into the LV
What is a common mechanical cause of loss of pressure in a hemodynamic monitoring system?
A. Catheter tip in turbulent flow
B. Loose connections among tubing, catheter, manifold, or transducer
C. Small internal diameter of the tubing
D. Use of stiff tubing in a hyperdynamic patient
B. Loose connections among tubing, catheter, manifold, or transducer
What hemodynamic artifact may result from loose connections in the pressure monitoring system?
A. LV pressure appearing higher than Ao pressure
B. Complete loss of signal
C. LV pressure falsely appearing lower than Ao pressure
D. Pulsus paradoxus waveform
C. LV pressure falsely appearing lower than Ao pressure
Which condition must be ruled out before concluding that low LV pressure is due to a technical issue like loose connections?
A. Septic shock
B. Cardiac tamponade
C. Heterotopic heart transplant or extracardiac hemodynamic support
D. Pulmonary embolism
C. Heterotopic heart transplant or extracardiac hemodynamic support