Previous Exam Guides Flashcards
(136 cards)
QA program monitors
ppl
QC program is for
film processors, CT scanners, XR tubes
most QC testing on rad equipment is done
annually
all general purpose rad units must have a min. of _____ mm Al filtration
2.5
filtration is measured w _______ measurements
HVL
focal spot size is measured w
pinhole/star/slit camera
it is acceptable to perform a ______ test instead of measuring the focal spot size
line pair
the measured kVp should be +/- ____ of the kVp indicated on the console
10%
what is the testing device used to measure cassette screen-film contact?
wire-mesh
comparing the mR output ay 50 mA @ 0.2 sec, 100 mA @ 0.1 sec, & 200 mA @ 0.05 sec is part of a test for _______
linearity
when a light in a viewbox is out, then
bulb must be replaced, all bulbs in viewboq must be replaced, all bulbs must match adjacent viewboxes
processor crossover racks should be cleaned out on a _______ basis
daily
processors should be monitored every day for
densitometric measurements, water/chem temps, & replenishment tank levels
an artifact can be
an object that is not part of the XRed anatomy, pt motion, improper grid, or warped cassette
a ________ is an example of an exposure artifact
foreign object
a static mark artifact is created during ______ of the film
handling
what sort of artifact is a guide-shoe mark?
processing artifact
the largest # of repeat exams is generally caused by ________ types of artifacts
exposure
misalignment of the processor turnaround assembly can cause?
guide-shoe marks
thin lines on the leading/trailing edge of a film are
guide-shoe marks
a bi-color chem stain on an XR
curtain effect, or dichroic stain
low humidity in the darkroom can cause
tree static
a streak of high optical density on the finished XR can be caused by
light leaks in the cassette/darkroom
when a QC test for exposure linearity is performed, adjacent mA stations must be w/in this amount of one another
10%