1
Q

What determines how frequently a particular QC test should be performed

A
  1. Likelihood of the fault occurring
  2. Consequence of a fault if it occurred
  3. Likelihood of detection by other means
  4. Manufacturers recommendations
  5. National guidelines
  6. Legal requirements
  7. The resource cost involved
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What must happen when a QC test exceeds an action level?

A

If QC exceeds an action level then either investigation or remediation should be scheduled but the equipment can still be used

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

What must happen when a QC test exceeds a suspension level?

A

If QC exceeds the suspension level then the equipment must not be used until it has been investigated or recalibrated