Rocuronium 1%
0.6 mg/kg
Midazolam 1 mg/mL
1 – 2 mg
Fentanyl 50 mcg/mL
1 – 2 mcg/kg
Propofol 1% (or 10 mg/mL)
2 mg/kg
Lidocaine 2% (or 20 mg/mL)
1 – 1.5 mg/kg (max of 100 mg)
“S”AMMTIDE
Suction
S”A”MMTIDE
Airway
SA”M”MTIDE
Monitor
SAM”M”TIDE
Machine
SAMM”T”IDE
Tape/Table
SAMMT”I”DE
IV
SAMMTI”D”E
Drugs
SAMMTID”E”
Everything else
Supplies for airway:
2 different laryngoscope blades and handles. 2 ETT - different sizes. 2 oral airways – different sizes. Ambu bag LMAs
Stylet shouldn’t pass..
the murphy eye.
Monitor supplies
EKG + electrodes, pulse oximeter, BP cuff, ETCO2, Nerve stimulator, temp probe
Steps in checking machine:
Turn on/verify red outlet. Check cylinder pressures. Check pipeline pressures. Calibrate O2 analyzer to RA and 100% O2. Preform circuit pressure check. Check ventilator function. Check soda lime. Check fill level of vaporizers.
IV supplies:
IV start kit, IV catheters, etc. IV fluid IV tubing (primary, secondary, blood tubing) Art. line start kit & transducer CVL start kit
Everything else:
stethoscope
Bair hugger
NG/OG tube
IV pole/drape clamps
Cylinder pressures:
What is a full O2 tank?
When does it need to be changed?
1900psi
<500psi
Pipeline pressure should be:
45-55psi
Steps in circuit pressure check:
- flowmeters turned off.
- close adjustable pressure limiting valve (APL)
- occlude breathing circuit
- Pressurize circuit no more than 35cmH2O (O2 flush)
- confirm no pressure change for 10 seconds.
- squeeze bag to determine no leak.
- open/close vaporizers to determine no leak.
- open APL to determine patency of valve.
- squeeze air out of bag to determine patency of scavenging system.
Steps to check ventilator function:
- place reservoir bag on distal end of circuit.
- verify APL is open.
- Turn O2flow to 6-10L/min.
- switch from manual to ventilator mode.
- Watch ventilator bellows.
- Test lung should inflate/deflate.
- check inspiratory and expiratory flutter valves.
- disconnect bag/replace mask
- turn off O2 flow