System Operation & Components Flashcards
(40 cards)
- Which system component converts the projection attenuation data into the proper digital form for the array processor?
A. Detector
B. Photodiode
C. Analog to digital converter
D. Host computer
-
- Which of the following is not a function of the PACs system
A. Reconstruction of the image from the raw data
B. Long term data storage
C. Digital distribution of images
D. Image display for the radiologists
-
- Which of the following is not a part of this CT tube?
A. Tungsten target
B. Collimator
C. Rotating anode
D. Filament
-
- Which of the following is false regarding CT systems
A. They are limited to non-oblique transverse scanning
B. They cannot generate a straight coronal or sagittal image
C. They have no moving parts
-
- X-rays are using CT because?
A. They are easily produce but the high frequency generator
B. They are less harmful than a heating effects of a microwave radiation
C. They penetrate the body according to specific attenuation characteristics of each tissue
D. The FDA a imposes no limit on the dose to the patient
-
- Most of the commands from the technologist are received by the?
A. Array processor
B. Operators console
C. Host computer
D. Amplifier
-
- X-rays are produced whenever fast moving electrons collide with any form of matter because?
- The electron splits into two x-rays photons of equal energy upon impact with the matter
- The electron loses kinetic energy which is converted into the emission of an x-ray photon
- The electron converts the electron it collides with into a x-ray photon
Which ones are correct
-
- The mA determines that the?
A. Quantity of x-rays photons
B. Co efficient of attenuation
C. Slice thickness
D. Energy level of the x-ray photons
-
- Which of the following is a component not normally located in the gantry of the modern CT system
A. Pre-patient collimators
B. Solid-state detectors
C. Array processor
D. X-ray tube
-
- Tube interscan delay time refers to
A. The time between the end of one scan and the start of the next scan during which the tube will cool
B. The quality insurance test performed daily by the technologist
C. The time required for the tube to make a complete 360° revolution about the gantry
D. The length of time required for the production of photons to begin after applying the voltage to the tube
-
- Increase ____________ Will increase that likelihood that a given x-ray will penetrate a material?
- kV
- mA
- exposure time
Which ones are correct
-
- Which of the following are not commonly use on today’s commercially available CT scanners
- Xeon detectors
- Fourth generation technology
- Continuous rotation technology
Which ones are correct
-
- Which of the following statements is false regarding occurrences after an x-ray photon penetrates the detector aperture?
- The photon can pass through the detector unmeasured
- The photon is converted completely into a electron
- The measured single is enhanced by amplifier
Which ones are correct
-
- Scatter radiation is caused bye
A. Miss calibrated detector
B. X-rays and generated spontaneously in the ambient atmosphere
C. Deflections from the original trajectory
of an x-ray photons through the patient
D. Improper patient positioning
-
- What controls does the operator have over the x-rays in a CT exam?
A. The temperature and color of the x-rays can be tuned
B. The energy level and the quantity of x-rays can be selected
C. The volume and tone of the x-rays can be dialed
D. The phase and frequency of the x-rays can be adjusted
-
- Which of the following is a standard that’s used to achieve compatibility for image transfer between various imaging modalities viewing stations and printers in a hospital
A. TCP/IP
B. T1
C. PACS
D. DICOM
-
- Enormous heat builds up in a CT tube is caused by the
A. Intensity of the x-rays in emitted from the tube
B. Electron beam between the Cathode and anode in the tube
C. X-ray filter in the tube which absorbs the lower energy photons before they can enter the patient
D. Collision of the electron beam with the tungsten target on the tube anode
-
- ___________________ made helical imaging possible
A. Multi-row detector scanners
B. The introduction of second generation technology
C. The introduction of fourth generation technology
D. Continuous rotation scanners
-
- The patient table:
A. May move continuously during the scan
B. Can never be position automatically by software due to safety precautions
C. Remains stationary throughout the patient exam once it is in the Shiley positioned
D. Only moves after each slice is completed
-
- The main purpose of the detector is to:
A. Provide additional collimation in order to reduce scatter
B. Limit the x-ray dose to the patient
C. Catcher x-rays photons and convert them to a measurable signal
D. Amplify the measured signal
-
- The cathode white filament:
A. Helps determine the size of the focal spot
B. Lights up the tube so it can be serviced by engineer
C. Releases the x-ray photons
D. Focuses x-rays on to the anode target
-
- Which of the following statements is true?
- KV is the voltage potential between the tube Cathode and anode while mA ultimately controls the filament current and thus the temperature of the cathode filament
- Kate V controls the energy level of the x-ray photons and mA controls the number of x-ray photons emitted from the tube
- Although the KV and mA affect the operation of the CT x-ray tube, the two parameters have no bearing on image quality
-
- The number of electrons flow from the cathode to the anode to the tube is controlled by the
A. Android target
B. Collimation
C. mA
D. kV
-
- Which of the following is true regarding solid-state detectors
- X-ray photons cause the detectors to generate a flash of light
- They are the detector type use on most CT scan or’s today
- They can be used in both the third generation and fourth generation CTs systems
-