Topic 19: dose measurement + QA in radiotherapy Flashcards
(23 cards)
Why is dose distribution by X-ray within tissue important?
- Used to determine dose at specific patient volume
How is dose distribution measured?
- Using tissue equivalent phantom:
> Has same atomic number as tissue
> Same e- density as tissue - Same density as tissue
Why is water used inside the phantom?
- Must be possible to move ionization chambers within phantom
Describe isodose charts
- Dose distribution plotted in form of isodose chart
- Isodose lines = line of constant dose
- Expressed as % of max dose
Describe effect of depth to dose
- Dose will fall with depth = radiation weaker > deeper beam goes = some energy absorbed + spread via inverse square law
Describe the effect of low/energy beams
- Low energy = edges diffuse = blurry due to compton scattering = hard to target tumors
- High energy = edges better defined = good focus for tumors
Explain the skin-sparing effect at higher energy
- High-energy beams don’t give max dose at skin surface = they go deeper = skin gets less damage = better for the patient
- At higher energy = 100% dose line below surface of phantom = spot where the full treatment dose hits = inside the body = protects the skin + shallow tissues
What is the relationship between depth of peak + photon energy?
- Depth at which the dose is highest = peak = is about ¼ of the energy in megavolts
What is the main advantage of high energy therapy?
- Surface dose only 20-50%
- Protects skin + targets tumors deeper
Describe treatment plan
- Graphically describes dose distribution when 1+ beams converge on treatment volume
- Treatment volume = volume that has to be treated
- Treated volume = volume actually treated
> Planning target volume
> Clinical tumor volume
> Gross tumor volume
Give the criteria for good dose distribution
1) Dose throughout treatment volume uniform ±5%
2) Treated volume should be as close to PTV
3) Dose to treated volume = exceed dose by at least 20%
4) Dose to sensitive sites e.g. eyes/spine = be below tolerance dose
How can the criteria be satisfied?
- By using several beams
- Small as possible
- Enter patient as close to treatment volume as possible
Describe the steps for LINAC planning
- CT for image guided radiotherapy taken
- Targets defined
- Treatment planned by medical physicist
- Physician verify plan + approve for use
Give the actions planned by medical physicist
- Choose radiation beam direction
- Choose beam type + energy
- Conform radiation for each beam
- Execute treatment plan simulation
- Verify plan
Describe dose distribution of skin
- Skin = curved = beam enter skin at not 90◦ = alter dose distribution
- Solutions:
1) Curvature corrected
2) Distribution corrected - At low energy = no skin-sparing effect = skin can be built up = using tissue-equivalent bolus = give surface at right angles
- At high energy = bolus not used = mathematical corrections applied
Explain the use for tissue compensator
- Placed in beam remote from skin
- Suitably shaped attenuator makes up for missing attenuation of tissue
- If large volume at surface needs to be treated = distribution altered via wedge-shaped compensator
- Wedge = attenuate beam at thick end = tilt isodose curve
Why are beam intensities optimized?
- Achieve best coverage of tumor + save healthy tissue
What is the use of collimators?
- Spare sensitive organs
Where are isodose curves plotted?
- On the anatomy = MRI/CT images
Describe IMRT
- Targets + sensitive organs defined = computer finds best plan
- IMRT = delivers radiation beams in multiple arcs
- Uses = sophisticated inverse planning software + multileaf collimators = shape radiation beam + modulate beam intensity
- Delivers optimum dose
Describe the simulation
- Diagnostic X-ray set = same mounting + couch movements as treatment accelerator
- Radiographs taken = provide info on treatment area for planning = check for accuracy
- Can be used for localization of sealed sources = brachytherapy
Describe positioning
- No point in making accurate plan if patient positioning cannot be guaranteed
- Patient = placed in correct position in relation to beam = remain stationary during treatment
- Patient shell may be used to position = used in 30% patients = head + neck
Describe the aspects of QA
- Commissioning of new equipment
- Interface between various equipment
- QA = daily/weekly/monthly/yearly
DAILY: - Constancy
- Ionization chamber
WEEKLY: - Profiler
MONTHLY: - Profiler
- Ionization chamber
YEARLY: - Phantom
- Water