Tx Planning Final Flashcards
(82 cards)
T/F: Penumbra decreases w/increasing SSD
F
T/F: Penumbra increases with larger source size
T
Electrons can be used to treat tumors to a depth of _________cm deep.
5
T/F: Safety is no accident is published by ASTRO and is a recommendations for radiation therapy departments
T
Treatment planning includes the following steps: (select all that are true)
patient positioning
tumor delineation
beam arrangement
computing dose delivered and eval of treatment plan
transfer of the treatment planning information to delivery sysetm
All
T/F: If you tattoo a patient in SIM at the CT setup point but dosi moves isocenter to another location you will have to shift daily on the machine?
T
How do you verify your shifts from CT isocenter to treatment isocenter?
Subtract the coodinates in the X,Y, and Z direction from CT iso to plan iso
When determining parameters to optimize a patients treatment some of the parameters include: (select all that are true) Target volume Dose-limiting structures Treatment volume dose prescription dose fractionation dose distribution
All
T/F: OARs are contoured on a CT scan by the physician
F
T/F: IMRT is known as inverse planning
F
What modality gives the best spatial resolution?
CT
motion artifact can happen on scan from the following: (Select all that is correct) breathing motion cardiac motion bone motion table motion
breathing motion
cardiac motion
Match the tissue-type to the associated HU:
bone
water
air
bone = +1000 water = 0 air = -1000
At diagnostic energies the most common interaction in tissue is what? (select all that is correct)
photo electric
compton effect
pair production
photo electric
compton effect
Gross tumor + microscopic disease is called
CTV
T/F: Target volume is defined as the primary tumor + local spread and lymphatic spread
T
Commonly used imaging studies to help delineate the tumor is the following: (Select all that apply) CT PET SPECT MRI PMRI
CT
PET
SPECT
MRI
Irradiated volume is defined as the ____________ Isodose line
50%
What’s the energy range?
Co-60, 4MV, 6MV
Low-energy Megavoltage X-ray
What’s the energy range?
10-25MV
Medium or High energy Megavoltage X-ray
T/F: To treat with IMRT you must have MLC or compensator option on your treatment unit
T
T/F: Using low energy megavoltage x-ray beams while treating parallel-opposed beams, the body thickness should not exceed approximately 15cm?
T
Delivering diff dose to diff volumes in 1 course of tx over 33 fxs.
Dose painting
Select below the MLC size that would give you smaller beamlet options
1cm
0.5cm
0.5cm