Vaporizers Flashcards
Pressure exerted by the molecules of vapor is the
saturated vapor pressure
As temperature increases
vapor pressure increased
atmospheric pressure
is the boiling point
Isoflurane boiling point
48.5C (119.3F)
VP 238
Sevoflurane boiling point
58.5C (137.3)
VP 170
Desflurane boiling point
22.8C (73.04F)
VP 669
temperature-compensated
vaporizer that is equipped to self-adjust to maintain a constant output over a wide range of operating temperatures
Factors that influence vaporizer output
Flow rates less than 250 ml per minute or greater than 15 LPM
Temperature – modern vaporizer construction minimizes the effect of temperature changes
Intermittent back pressure – older vaporizer models may be sensitive to use of the flush valve, high respiratory rates or high peak inspiratory pressures
Carrier gas composition – suddenly adding high concentration of N2O may cause transient reduction in vaporizer output because nitrous is more soluble in the volatile agents than O2
Essential parts of the vaporizer
Concentration control dial
Bypass chamber (handles > 80% of the flow)
Vaporizing chamber (handles < 20% of the flow)
Filler port and cap
Fill level gauge
Drain
Modern Vaporizers
- variable bypass
- flow-over
- temperature-compensated
- agent-specific
- out of breathing circuit
Vaporizer Hazards (Older Equipment)
Misfilling (wrong agent) – esp. older equipment
Contamination – fouled agent
Tipping (rare problem)
Overfilling – esp. older equipment
Under filling or dry vaporizer
Simultaneous inhaled anesthetic administration (older equipment without vapor-interlock system)
Leaks – environmental contamination
Desflurane Vaporizer characteristics
Tech 6
High vapor pressure – boils at 22.8° C (73°)
“highly volatile”
Desflurane easier to control if the agent is heated to 39° C which pressurizes it to almost 2 atmospheres (~ 1300 mmHg)
Tec-6 is described as a dual-gas blender rather than a variable bypass vaporizer
only one that has alarms for filling or not working.
require electricity
Altitude impact on vaporizers
Concentration output almost doubles (97%) but PP output only increases by 29%
Ambient pressure changes affect vaporizer output in terms of concentration output
With variable bypass vaporizers, partial pressure ratios change only a small amount over a range of altitudes
does not apply to desflurane and the dual gas blender vaporizer which does not provide a consistent partial pressure at varying altitudes