Week 5 Flashcards
(40 cards)
What are the electrical measuring systems we use at Michener
pH meter
Siemens RAPIDPoint 500 (Blood gas) analyzer
Vitros 350
what is polarography
-measuring current that flows through an electrochemical cell and the electrical potential between two electrodes as a constant voltage is applied
-the current is produced by a reduction of analyte in the sample
-amount of current is proportional to the concentration
-part of oxidation reduction reaction
-also known as amperometry
what is electrical potential
force that causes electrons to move in a conductor and is measured in volts
What are the advantages of polarography
-can detect low volume samples
-Pharmaceutical uses measuring Antibiotics, hormones, vitamins,
- rapid
-Cost effective
What is ion selective electrodes ISE
-selection for desired ion measurement
-ONLY ONE TYPE SHOULD BE DETECTED
-other ions are different because their ionic radius is different
-different types of polymeric materials as simple membranes to separate ions or a liquid membrane system for ion complexing
what are the advantages of ISE
-less expensive
-not destructive
-not contaminating
-short response time
-not affected by colored compounds or turbidity
What are glass electrodes made up of
-thin layer of alkali metal silicates acting as sieves for ions while producing chemically active binding sites in the glass
-composition of glass membranes determines what ion it is selective for
example
-H+ sensitive glass electrode (pH meter)
-Na+ sensitive glass electrode
What is a H+ (ph) glass electrode
-contains Ag wire in 0.1 HCl solution with AGCl tip
-a potential difference is created across a membrane that is measured
-pH electrode most common for H ion activity
-needs a glass membrane
What is a combination electrode
example: Galvanic cell
-when the indicator and reference electrodes are in one small probe
-Ag/AgCl internal reference covered in glass with pH sensitive tip. The Ag/AgCl are stable and dont change over time because of temperature
-Loss of electrolyte due to evaporation will not change the potential
-connected to voltmeter
-The outer glass is filled with KCl and has a little pore at the tip
What is Calomel HgCl
-mercury covered with Calomel -HgCl which will come in contact with Kcl
-As long as the Cl doesnt change the potential of the half cell is constant
-the activity of the Cl is set to a fixed value by saturating the solution with chloride salt like KCl
What is a pH meter
-gives H concentration in a solution using
Actual molar concentration of H ions
Dissociation constant of the acid (pKa)
Temperature
Issues with glass electrodes
- buffer standard accuracy
-junction drift must be corrected every 2 hours because of the potential at the pore
-whenever the H concentration is low and Na is high there can be a saline error or alkaline error which gives a low pH.
-temperature must be same during calibration and pH measuring
What is a solid state electrode
-measure of anions
-crystals on the membrane to provide ions
-silver salts common
-membrane of the Cl electrode will have AgCl and silicone
-used for Cl, Br, and I
What are liquid ion exchange electrodes
-has membrane with ion binder (carrier) dissolved in inert solvent (plasticizer) in PVC
-the carrier is called ionophore because it will selectively bind to ion of interest to produce electrical potential
Example:
K+ electrode: valinomycin (antibiotic) binds K+ in the centre of a ring of oxygen atoms
Ca2+ electrode: organo-phosphate salt in PVC
What is a compound gas electrode
-has a SECOND membrane
-the first separates what you want to test from the solution and it goes through a reaction and the second membrane measures the produce of the reaction
what is a gas sensing electrode for PCO2
Severinghaus CO2 electrode
-type of pH electrode good for pt monitoring
-CO2 will go through a Teflon membrane and alters the pH of a bicarb buffer solution and the H concentration is measured at the glass electrode
CO2 + H2O <—–> H2CO3 <—–> H+ + HCO3-
-there is a lower pH which is proportional to the CO2 in specimen - blood gas analyzer
what is a gas sensing electrode for PO2
-called a Clark electrode and is based on polographic techniques
-a platinum cathode surrounded by a tubular silver anode
-the O2 in a sample will diffuse through a gas permeable polypropylene membrane and be reduced at the platinum cathode with a 4 electron gain
O2 + 2 H2O + 4e- —–> 4OH-
Current produced proportional to partial pressure of O2 in sample (PO2)
-blood gas analyzer
What is an enzyme electrode
-measures non ionic compounds like glucose and urea
-the ions generated in an enzyme reaction are proportional to the concentration of analyte
urease immobilized in a membrane over an ammonium (NH4+) sensing electrode
urease
urea ————-> NH4+ + CO2
-NH4+ produced is measured by glass electrode & is proportional to concentration of urea
-NH4 can cause falsely high K
Example: enzyme glucose oxidase is used on an electrode to measure glucose
What are some membrane properties of electrodes
Selectivity
-membrane must have a high degree of selectivty for ion to be tested for example a Vanco membrane is more selective for K then Na
Lifetime
-FINITE life because of contamination, membrane parts loss- machine fatigue
Stability
-the junction potential will be changed because of protein decomposition due to blood and plasma causing varying electrical signal and loss of membrane components
Response time
-as the electrode gets older the response time increases and accuracy/precision decreases
What are some electrode configurations
- Stationary-pH meters
- Flow-Through
-membrane forms part of the wall of the “flowcell” that keeps multiple electrodes
-used on Siemens RAPIDPoint 500 measures pH, pCO2, pO2, Na+, K+ Cl- , Ca++ - Dry-Slide- used on Vitros (potentiometric slide)
measures Na+ K+ Cl-
What is in a Stationary Electrode Configuration
-simple electrode system; e.g. pH meter
-electrode is lowered into the test solution
hand held ones like from sofina
What are temperature and Protein contamination disadvantages for electrodes
Temperature
Nernst equation is temperature dependent
need fixed & constant temp. e.g. pre-heater
Protein contamination
-leads to response delay, changing selectivity and electrode sensitivity
-needs PM to deproteinize and condition electrodes
auto alert for Siemens RAPIDpoint
What are standardization and interfering substances disadvantages for electrodes
Standardization
-a standard should be the same as the unknowns pH, ionic strength, protein content
-must be wiped after use because otherwise there will a protein residue on the membrane causing residual junction potential
Interfering substances
-can cause false reading
Heparin can falsely decrease ionized Ca
What are electrode poisoning and ph change disadvantages for electrodes
Electrode poisoning
-when Electrodes are filled with incorrect filling solutions, deteriorating filling solutions
-when you use Harsh cleaning agents, disinfectants on an electrode it can cause damage to ISE membranes or denature enzymes on membranes
Specimen: pH changes
Some ions very sensitive because of competitive binding to proteins in serum e.g. ionized Ca++