Chapter 8 Q & A Flashcards

1
Q

given 5 subtances: Backing material, matching layer, PZT, gel and skin, whose impedences have been emasured. The lables have fallen off. Lables are 0.8Mrayls, 1.9Mrayls, 1.6Mrayls, 2.0Mrayls and 1.0Mrayls.

If the impedence of the backing is 1.9Mryals what are the others?

Why

A

PZT 2.0 Mryals

matching layer 1.6 Mrayls

gel 1.0 Mrayls

skin 0.8 Mrayls

because the impdedences along the path decrease the closer you get to the skin

PZT>matching>backing>gel>skin

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Pulse length is (directly/incersely) related to pulse duration

A

directly

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Q-factor is (directly/inverserly) related to bandwidth

A

inversely

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

pulse duration is (directly/inversely) related to bandwidth

A

inversely

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

the sensitivity of transducers that create short duration pulses it likely to be (greather/less/equal to)that of transducers that create long pulses

A

less than

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

all correctly describe an imaging transducer except

A: high sensitivity

B low Q

C wide bandwidth

D damped

A

A does not describe imaging transducers

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

shorter duration events (such as damped pulses) are more likely to be wide bandwidth

True/False

A

true

short duratoin events are likely to be wide bandwidth

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

what occurs when a PZT crystal’s temp is raised above the Curie point

A

PZt is depolarized and the piezoelectric properties are lost

transducer is ruined

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

the accoustic impedence of the matching layer is approx the same as the accoustic impedance of skin

True/False

A

False

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

imaging transducers are usually of high rather than low bandwidth

A

True

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

A very high Q factor transducers is used more often in diagnostic imaging than a low Q factor

true/false

A

false

imaging transducers are low Q

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

the damping material in a transducer increases the sensitivy

true/false

A

false

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

damping material in a transducer increases the pulse length

true/false

A

false

damping shortens the pulst length

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

damping material in a transducer decreases the pulse duratoin

true./false

A

true

damping reduces pulse duration

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

damping material in a transducer improves the systems lateral resolutoin

true/false

A

false

damping does not affect lat res

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

damping material in a transducer improves the systems longitudinal resolution

true/false

A

true

damping improves longitudinal res

17
Q

damping material in a transducer decreases bandwidth

true/false

A

false

damping increases bandwidth

18
Q

damping material in a transducer decreases the quality factor

true/false

A

true

damping decreases the Q factor

19
Q

if the freq of the electrical excitation voltage of a pulsed wave transducer is 6 MHz, then the operating freq of the transducer is 6 MHz

true/false

A

false

pulsed freq is NOT determined by excitation freq

20
Q

if the pulse repitition freq of a transducer is increased, then the freq of sound produced by the transducer remains the same

true/false

A

true

Freq and pulse rep are not related

21
Q

the diameter of the active element of a transducer helps to determine the freq of the sound that the transducer creates

true/false

A

false

22
Q

freq of the electrical excitation voltage of a continuous wave transducer is 6 MHz then the operating freq is 6 MHz

True/false

A

true

Only CW transducers

23
Q

two piezoelectric crystals are made from the same material. The thicker crystal will make a continuous wave transducer with a lower freq

true/false

A

false

with CW element thickness does not determine sound freq

24
Q

the normal propagatoin speed in piezoelectric material is about 3 to 5 times greater than that is soft tissue

true/false

A

true

25
Q

the impedance of a transducer active element is 1.9 Mrayls and the impedance of the skin is 1.4 Mrayls

what is an acceptable impedance for the matching layer?

A 1.2 Mrayls

B 1.4 Mrayls

C 1.726 Mrayls

D 1.95 M rayls

A

C

the impedance of the matching layer is between that of the active element and the skin

26
Q

which of the following crystals will produce sound with the lowest freq?

A thin with low speed

B thin with a high speed

C thick with a low speed

D thick with a high speed

A

C

lowest freq is produced by thick and slow active element

27
Q

WHich tyupe of transducer has a great Q-factor, therapeutic or imaging

A

therapeutic

28
Q

which type of transducer has a greater bandwidth, continuous wave or imaging

A

imaging

29
Q

which type of transducer has more backing material, therapeutic or imaging

A

imaging

30
Q

in an imaging transducer, what is the purpose of attaching the backing material to the PZT?

A increase bandwidth

B decrease the Q-factor

C improve image quality

D decrease the transducers sensitivity

A

C

improve image quality

31
Q

a pulsed wave transducer has a resonant freq of 5 MHz. The lowest freq in the pulse is 2 MHz and the highest is 8 MHz. What is the bandwidth?

A 5 MHz

B 8 Mhz

C 3 MHz

D 6 MHz

E 2 MHz

A

D

bandwidth is the range of freq found in a pulse

8 MHZ-2 MHz=6 MHz

32
Q

for the transducer described in # 25, what is the main freq?

(resonant 5 MHz, low 2 MHz, high 8 MHz)

A 5 MHz

B 8 Mhz

C 3 MHz

D 6 MHz

E 2 MHz

A

A

the resonant , main or center freq is 5 MHz

33
Q

for the transducer described in # 25, what is the Q-factor

(resonant 5 MHz, low 2 MHz, high 8 MHz)

A 0.8

B 3 MHz

C 5 MHz

D 1.5

E 2 MHz

A

A

Q-factor is a resonant freq divided by the bandwidth

5 MHz/6 MHz=.8

Band width has no units.