3. Equipment and Physics Flashcards

(112 cards)

1
Q

Who first studied nerve stim

Who first used it to perform block

How were nerves blocked prior to this

A

Electrical nerve stimulation was first studied
by French physiologist Louis Lapicque in 1909.

It was first used to perform nerve blocks by
Von Perthes in 1912.

Before this, nerves were blocked by
direct instillation of local anaesthetics
(by dissection and exposure of nerve plexus) or paresthesia techniques.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Nerve stim

How does the technique work

A

The technique of electrical nerve stimulation

is based on the premise

that a current of sufficient amplitude
applied for a sufficient time will
depolarise a nerve.

In the case of nerve blocks,
this means either motor response
or sensory stimulation (since most nerves are
mixed).

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Which is more commonly stimulated Motor or sensory

What does the Cathode do

What does the Anode do

A

However, it was also noted that stimulating

motor fibres was easier

than sensory fibres,

and more importantly,

application of a cathode
depolarised the nerve,

while an anode hyperpolarised the nerve.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

What does the Cathode do

What does the Anode do

A

application of a cathode

depolarised the nerve,

while an anode
hyperpolarised the nerve.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

What is the gold standard

A

At present,

ultrasound guidance is becoming more popular,

but electrical nerve stimulation is
still the commonest method employed.

However, no method of nerve blockade is described as gold standard.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

What is Rheobase

A

Rheobase

is the minimum current of

indefinite duration required

to depolarise a nerve.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

How do we calculate the total charge required to depol nerve

A

The total charge (Q) required to depolarise a nerve is

the product of the current intensity (I) 
\+
the duration (t) for which it is applied.
Q = I × t
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What current intensity is required for depolarisation

A

In turn,
the current intensity
required to produce depolarisation

is given by the following equation

(where Ir is the rheobase
and C is the chronaxie):

I = Ir × (1 + C/t)

t = infinity, we get I = Ir, and so Q = Ir.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

what is Chronaxie

A

Chronaxie is the minimum duration

of current twice the rheobase required

to stimulate a nerve

(as shown in the previous answer).

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Chronaxae related how to fibre size

A

It is inversely proportional to
fibre size and
hence ease of stimulation.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

What is the Chronaxiae of Aa fibres

A

Aα (motor) has a
chronaxie of 0.05–0.1 millisecond,

Aδ (sensory) is 0.15 millisecond
C (unmyelinated sensory) 0.4 millisecond

Hence, stimulating motor nerve
requires shorter pulses than sensory fibres.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Desirable properties of electrical
nerve stimulation are

x 5

  • the most important
A
  1. Short pulse width:
  2. Square-wave current
  3. Cathodal stimulation
  4. Constant current generator * most important
  5. Frequency: 2hz
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Short pulse width

Refers to

Why is this advantageous

A

pulse width refers to the time duration
for which the current is applied.
Shorter pulse width has two advantages:

1 Since the motor fibres have 
a smaller chronaxie, 
shorter pulse width
stimulates them 
but not the sensory fibres. 

This results in motor responses
but not painful paresthesia,
which is undesirable anyway.

2 Shorter pulse width may be superior
to longer in estimating needle to- nerve distance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Square-wave current

A

Slow rising current allows

for accommodation

(resulting in difficulty in nerve stimulation)

of nerve fibres.

This can be avoided by the
square-wave form of applying
current (abrupt rise and abrupt fall).

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Cathodal stimulation

A

Cathodal stimulation:

it is preferable to stimulate
the nerve with needle as cathode,

since this then depolarises it,

whereas needle as anode
hyperpolarises the nerve

(necessitating application of higher
current for stimulation).

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Constant current generator

A

Constant current generator (not fixed):

a peripheral nerve
stimulator (PNS) should 
deliver the same current 
despite changing
impedance applied. 

This is the most important property of the
peripheral nervous system (PNS).

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

Frequency:

A

Frequency:
a stimulation frequency of
2 Hz is better than 1 Hz,

since it allows
faster manipulation of needle.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

Describe important things during PNS

Negative

Positive

Distance between

Current

A

During nerve stimulation,
the following things are vital:

  1. Negative (cathode) to needle.
  2. Positive (anode) to patient.
  3. It was considered that the anode
    site should be at least 20 cm away
    from the needle site to reduce direct muscle stimulation, but this has
    been found to be unnecessary.
4. 
Acceptable current is between 0.2 and 0.5 mA. 
Above 0.5 mA, the needle 
may be further away from the nerve, 
and such injections may
not be successful. 

Below 0.2 mA, injection may be intraneural.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

components of a peripheral nerve stimulator

A

Microcontroller

Constant current generator
(most important)

Oscillator

Clock reference

LCD display

Controls

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

components of a peripheral nerve stimulator

Explained

A
  1. Microcontroller
    Brain of the peripheral nervous system:
    processes variable, like
    current, pulse width, frequency
  2. Constant current generator
    (most important)
    Generates the same current despite changing impedance

3 Oscillator
Generates the desired frequency

4.
Clock reference
Synchronises the current with the frequency

  1. LCD display
    For current amplitude, frequency and the pulse width selected

6Controls
For selecting parameters

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

appropriate settings of a PNS

for performing a nerve block include

A
  1. negative lead to needle

2.
positive lead to patient

3.
a square-wave impulse (to prevent accommodation)

  1. pulse duration 0.1 millisecond (for stimulating motor nerve fibres preferably)

5.
frequency of 2 Hz (better than 1 Hz)

  1. an initial current of 1–2 mA

7.
a final current of 0.2–0.5 mA
(> 0.5 mA, the needle may be further
away from the nerve,
and such injections may not be successful;
< 0.2 mA, the injection may be intraneural)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

What is the law that governs the

principle of nerve stimulation

A

The current required is
inversely proportional
to the square of the distance
between the needle and the nerve

Coulombs

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

Coulombs Law

A

The inverse-square law (Coulomb’s Law)
dictates that the current required

(I) to stimulate a nerve,
is proportional to the minimal current (i),
and
inversely to the square of the distance (r)

from the nerve (k is a constant)
.
I = k(i/r2)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q
How may nerve stimulation be altered in 
elderly, 
diabetics or 
those with
neurological diseases,
A
Usually, 
a motor response between 0.2 and 0.5 mA 
is sought and considered appropriate. 
However, 
in elderly, 
diabetics 
or those with neurological diseases, 

higher currents may be needed due to slower
nerve velocities and lower motor amplitudes.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
What is the Raj test How is it performed What does it confirm Explain the mechanism
The disappearance of the motor response induced by a low current (0.5 mA) following injection of local anaesthetics or normal saline (conducting solutions), confirms the proximity of needle to the nerve and constitutes the Raj test. This does not result due to the physical displacement of the nerve but due to the dissipation of current density near the nerve.
26
What is the Tsui test. How is it performed What does it confirm
``` The exaggeration of motor response induced by a low current (0.5 mA) following injection of 5% dextrose (non-conducting solutions), ``` confirms the needle-to-nerve proximity as well and constitutes the Tsui test
27
Is a sensory response able to elicit a motor response
a lack of motor response does not rule out the possibility of sensory nerve contact by the injection needle
28
Peripheral Nerve Stimulators 1. Optimal Range
Optimal range for a PNS is 0–5 mA. This is because some patients may need higher current for stimulation (diabetics, elderly, neurologic disease). Newer devices may have higher ranges (0–10 mA) used for epidural stimulation. ``` Higher ranges (0–80 mA) are used in neuromuscular monitors. ```
29
Peripheral Nerve Stimulators Percutaneous nerve stimulation
Percutaneous nerve stimulation is a new technique involving the stimulation of nerves non-invasively. ``` The current needed for this is higher than invasive stimulation, but offers the identification of insertion points in especially difficult cases (obese). ```
30
What range should the PNS be checked in.
Biomedical engineering departments have measured the accuracy of PNS in the higher current ranges (> 1 mA) in the past. ``` It was subsequently argued that since the current used for performing nerve blocks is in the range of 0.2–0.5 mA, it is prudent to check the accuracy in this range. This has been adopted by some manufacturers ```
31
Which are more accurate insulated or non
Non-insulated needles were the first to be used. Both the tip and the shaft were conductive, causing current dispersion and lower accuracy.
32
What is an issue with non insuated needles
They also caused local muscle stimulation through the shaft of thenneedle.
33
What are insulated needles coated in Why are they beneficial
The development of Teflon-coated insulated needles resulted in better precision. This is because only the tip is conductive, and hence the current is not dispersed.
34
What type of needle tips are available
Various needle-tip designs are prevalent. Among the sharp needles, the tip may have a long (standard, 15°) bevel or short (30° or 45°) bevel
35
Which bevel cuts nerves Which bevel causes blunt trauma
The long-bevel needles may cause sharp cuts on nerves, while the short bevel leads to blunt nerve damage.
36
Which type of needle bevel is more frequently associated with injury Which is more severe
Although nerve injury is more frequent with long-bevel needles, it may be more severe if it occurs u sing a short-bevel needle.
37
Which bevel is used more frequently these days
``` Blunt-bevel needles offer more resistance as they pass through tissue planes and thus give a better feel. ``` Hence they are most commonly used nowadays
38
Needle gauge is an important consideration while performing blocks Superficial injections
Needle gauge is an important consideration while performing blocks. Superficial injections are best given using 25/26-G needles.
39
Needle gauge is an important consideration while performing blocks Single shot injections
The 21/22-G needles are best for single-shot injections,
40
Needle gauge is an important consideration while performing blocks Continuous catheter injections Catheter size
18/19-G Tuohy-tip needles are best suited for continuous catheter techniques. In such cases, 20-G catheters are used.
41
Needle length is an
consideration when doing nerve blocks. Shorter needle may not be sufficient, while longer needles may have potential for tissue damage if introduced further than needed •
42
25 mm what block
Interscalene
43
50 mm what block
Cervical plexus Supraclavicular Axillary Femoral Popliteal (posterior)
44
100 mm
Infraclavicular Popliteal (lateral) Paravertebral Lumbar plexus Sciatic (posterior)
45
150 mm
Sciatic (anterior)
46
An in-line pressure-monitoring provides
An in-line pressure-monitoring device measures pressure while injecting local anaesthetic. It provides an objective assessment of pressures rather than subjective feel. The latter can vary between individuals and devices. I
47
Intraneural injection pressures Perineural
Intraneural injection pressures are >20 psi while those made perineurally have lower pressures of 5–20 psi. Therefore, the device guides placement of the tip according to injection pressures.
48
The continuous catheter technique involves 2 elements and sizes
The continuous catheter technique involves the use of larger needles (Tuohy-, Sprotte- or facet-tipped 18/19 G) and fine stimulating catheters (20 G). Once a nerve is stimulated using the needle, the catheter is threaded through the needle.
49
In the case of non-stimulating catheters, how is this done
In the case of non-stimulating catheters, the perineural space is dilated by injecting saline and threading the catheter 3–5 cm beyond the needle tip.
50
Benefit of Catheters How were they limited
Although the use of catheters helped to prolong postoperative pain relief, they were limited by secondary block failures (primary block refers to the block following initial injection, while secondary block is one following continuous infusion).
51
How was secondary failure improved
This was improved by stimulating catheters. They are threaded along the nerve, and their position confirmed using electrostimulation in real time. This is followed by initial bolus and continuous infusion, both through the catheter. This has reduced secondary failures.
52
‘catheter over needle’ and ‘catheter through needle’ Which is more frequently used which has a problem with leakage why
``` Systems with ‘catheter over needle’ and ‘catheter through needle’ are available. ``` ``` Though the latter are more prevalent, their use is often plagued by leakage through the injection site (because the hole made by the needle is larger than the catheter size). ```
53
Tunnelling the catheter
reduces the chances of dislodgement and helps maintain the | catheter for a longer time.
54
Ultrasound waves are Beyond what which is Clinical U/S are what freq
(> 20 KHz) are waves beyond the audible frequency range of audible sound (20–20 000 Hz). Clinically used ultrasound is in the 1–20-MHz frequency range.
55
Describe how an US Probe works
Ultrasound waves are generated by applying an electric field to piezoelectric crystals to produce a series of pressure waves. ``` The pressure waves are transmitted from the probe head and reflected back dependent upon the tissue type. ``` The returning pressure waves are detected and generate an electric current that is converted into a two-dimensional image. This interpretation assumes the speed of sound in soft tissues to be 1540 m/second.
56
The speed of sound in soft tissues
The speed of sound in soft tissues to be 1540 m/second.
57
The various modes of ultrasound in use are:
A Mode B Mode M Mode Doppler mode
58
A-mode:
A-mode: the simplest type of ultrasound. A single transducer scans a line through the body with the echoes plotted on screen as a function of depth.
59
B-mode:
B-mode: the commonest mode. A linear array of transducers simultaneously scans a plane through the body that can be viewed as a two-dimensional image on screen.
60
M-mode:
M-mode: M stands for ‘motion’. Ultrasound pulses are emitted in quick succession, recording a video in ultrasound. This can be used to determine the velocity of specific organ structures such as cardiac valves and jets
61
Doppler mode:
Doppler mode: This mode makes use of the Doppler effect ``` (change in frequency of a wave for an observer moving relative to the source of the wave) ``` in measuring and visualising blood flow.
62
What type of waves are US waves What are their properties
Ultrasound waves are sound waves Wavelength (λ) Amplitude (A) Frequency (f) Period (τ) Velocity (c):
63
Wavelength (λ):
Distance between two consecutive corresponding | points of the same phase.
64
Amplitude (A)
maximum height of the wave.
65
Frequency (f)
number of complete cycles per second.
66
Period (τ):
time taken for one complete wave cycle to occur.
67
Velocity (c): Calculted how
speed at which sound waves pass through a medium. It may be calculated using the equation: c = λ × f
68
What is relationship between wavelength and frequency why
wavelength and frequency bear an inverse relationship. Since the velocity within a medium is constant,
69
How does frequency affect image quality high
Higher-frequency beams experience more attenuation (directly proportional) and hence have lesser penetration. ``` They are used for performing superficial blocks (interscalene and supraclavicular). ```
70
How does frequency affect image quality Low
Low-frequency beams have better penetration and allow visualisation of deep structures (infraclavicular and sciatic nerve blocks).
71
Linear probes generate Curved generate
Linear transducers generate high-frequency ultrasound, while curved array probes low-frequency ultrasound.
72
Linear array ``` freq ax res attenuation depth image best ```
Linear array ``` High frequency (6–13 MHz) Greatest axial resolution More attenuation Limited depth of penetration Rectangular images Best for superficial structures (e.g. brachial plexus) ```
73
Curved array ``` freq ax res attenuation depth image best ```
Curved array ``` Low frequency (2–5 MHz) Decreased axial resolution Less attenuation Deeper penetration Sector-shaped image Best for large or deep structures (e.g. sciatic nerve) ```
74
Phased array ``` freq ax res attenuation depth image best ```
Phased array ``` Consists of many small ultrasonic elements High-resolution beam Characteristic image is sector-shaped Used for echocardiography ```
75
J-shaped
J-shaped (hockey-stick footprint) probes are linear array probes which are small in size hence ideally suited for paediatric usage.
76
Axial resolution
Axial resolution is the ability of the system to display small structures along the axis of the beam as separate from each other. It is directly proportional to the frequency of the beam.
77
Lateral resolution
is the ability of the system to display small structures side by side (same depth) as separate from each other.
78
Attenuation is What does it cause
``` The sum total of reflection, refraction, scattering and absorption ``` It leads to loss of clarity of the image.
79
How can attenuation be corrected
It can be corrected by time-gain compensation (also called depth-gain compensation). Attenuation is directly proportional to the frequency. So higher frequency ultrasound beams undergo higher attenuation and allow the best visualisation of superficial structures.
80
What is an artefact What are the diff types artefact x4
An artefact is an image, or part of it, that does not correspond to the anatomy of the structure being examined. Shadowing Post-cystic enhancement Reverberation Anisotropy
81
Shadowing
Shadowing: when the ultrasound beam cannot pass through a structure (e.g. bone), the beam is reflected back and the tissues immediately behind the structure appear dark.
82
Post-cystic enhancement
Post-cystic enhancement: when the ultrasound beam passes through a fluid-filled structure such as the urinary bladder, cysts or blood vessels, very little is reflected, and therefore the tissues behind the fluid appear bright.
83
Reverberation
Reverberation: occurs when ultrasound is repeatedly reflected between two highly reflective surfaces.
84
Anisotropy
Anisotropy: is the property of tendons, nerves and muscles to vary in their ultrasound appearance depending on the angle of insonation of the incident ultrasound beam.
85
what is echogenicity
On returning to the transducer, the amplitude of an echo is represented by the degree of brightness (i.e. echogenicity) of a dot on the display. Each tissue displays a different echogenicity, allowing identification of structures.
86
Anechoic:
Anechoic veins/arteries offer no reflection and appear black.
87
Hypoechoic
Hypoechoic: muscle and central nerve plexus offer weak reflections and appear dark.
88
Hyperechoic
Hyperechoic: bone and peripheral nerves offer strong reflections and appear bright.
89
Factors determining needle visualisation under ultrasound ``` improves technique size of needle angle of insertion depth echogenicity ```
Technique deteriorates improves Technique Out of plane In plane Size of needle Smaller (22 G) Larger (17-G Tuohy) Angle of insertion Steep Shallow Depth of insertion Deep Shallow Echogenicity Non-echogenic Echogenic (cornerstone reflectors)
90
Cornerstone reflectors how
introduced by some companies (Pajunk), these are intended to improve needle visibility under ultrasound, even at steep angles. They do so by reflecting all ultrasound waves without losses.
91
real-time spatial compound imaging
In real-time spatial compound imaging, a transducer array is used to rapidly acquire several overlapping scans of an object from different angles. These scans are averaged to form a compound image that shows improved image quality because of reduction of artefacts.
92
technologies involve either one of the following:
Needle-guidance systems: Sonic GPS (Ultrasonics) and eTrAX needle systems (CiVCO). Newer imaging modalities: Multibeam (Sonosite), Cross XBeam (GE) and Flexi Focus (BK Medical).
93
How does U/S Aid RA Can detect what demonstrate reduces risk of
Ultrasound provides real-time visualisation of the nerve and surrounding structures during regional anaesthesia. It allows detection of anatomical variations and demonstrates spread of local anaesthetic during injection. Intraneural or intravascular injection can be detected by ultrasound.
94
How does it improve block What does the evidence not demonstrate
``` Evidence shows ultrasound-guided peripheral nerve blocks can be more successful than peripheral nerve stimulator techniques and have a faster onset time. ``` However, evidence to clearly demonstrate that ultrasound-guided regional anaesthesia is safer (in terms of neural injury) than using a peripheral nerve stimulator is still not available.
95
In the short-axis view
In the short-axis view tubular structures such as nerves and blood vessels appear as though they have been sliced across their diameter, like discs of salami.
96
In the long-axis view,
In the long-axis view, tubular structures are sliced longitudinally along the length of the tube.
97
The needle approach is described In plane
The needle approach is described as in-plane if the needle remains parallel to the ultrasound beam, allowing visualisation of the tip and shaft.
98
The needle approach is described Out plane
In the out-of-plane approach, the needle is inserted more perpendicular to the ultrasound beam and can only be visualised as a dot when the needle crosses the beam.
99
newer applications of electrical nerve stimulation include the following.
Percutaneous electrode guidance: Sequential electrical nerve stimulation (SENS): Epidural stimulation (Tsui test)
100
Percutaneous electrode guidance
involves percutaneous stimulation of peripheral nerves to identify a nerve before skin puncture. This reduces the number of unsuccessful painful insertion and helps identify the best insertion point.
101
Sequential electrical nerve stimulation (SENS):
delivers current at 3 Hz with sequential pulses of 0.1, 0.3 and 1 msecond to improve motor response.
102
Epidural stimulation (Tsui test)
Epidural stimulation (Tsui test): the placement of wired epidural catheters threaded into epidural space can be confirmed by electrical stimulation between 1 and 10 mA. The motor responses elicited direct toward the level of the tip of the catheter. Stimulation at currents < 1 mA indicate intrathecal placement.
103
Non-electrical methods to confirm the placement of epidural | catheters
Non-electrical methods to confirm the placement of epidural catheters include epidural pressure waveform guidance and electrocardiographic guidance
104
Multistimulation technique
Multistimulation technique involves seeking specific component bundle motor responses separately and then blocking the component nerves individually. For example, for the axillary block, the radial, ulnar, median and musculocutaneous responses are sought individually
105
Multistimulation technique Advantages
``` Higher success rate Lower total volume of local anaesthetic needed Shortening of onset times of nerve blocks Lower potential of local-anaesthetic toxicity because of lower doses ```
106
Multistimulation technique Disadvantages
Increased patient discomfort because of multiple needle redirections Time for the entire procedure is increased Theoretically, the risk of nerve damage may be higher because of repeated needle passage the incidence of nerve injury with multistimulation technique has been found to be similar to that of conventional techniques.
107
Multistimulation technique useful for
Multistimulation technique has been found to be useful for interscalene, axillary, infraclavicular, mid-humeral, femoral and sciatic nerve blocks.
108
Multistimulation too risky for
However, frequent redirections in the supraclavicular area may increase the risk of arterial puncture and pneumothorax, and are not advised.
109
As the current needed to stimulate nerve calculation what current is best used initially Then what improves specificity
As the current needed to stimulate a nerve is inversely proportional to the square of the distance from the needle, high currents help in finding the nerve initially. Hence they increase the likelihood of finding the nerve. Subsequently, lowering the current as the needle approaches the nerve helps to improve the specificity of the response.
110
The resistance encountered by a stimulating electrode depends directly on inversely on
The resistance encountered by a stimulating electrode depends directly on the tissue resistance and inversely on the conductive area of the electrode.
111
How can nerve stimulation electrode be improved
Stimulating a nerve using a microtip electrode (i.e. despite high resistance) improves the specificity of the stimulation.
112
What is water lipid ratio of nerves How does this affect stimulation
Water–lipid ratio of tissues is proportional to conductance, and since nerves have a high water–lipid ratio, they have higher conductance than skin, muscle, fat or bone; hence they are stimulated in preference upon application of a current.