Follow-up: Device assessment Flashcards
Hones pattern recognition of IEGMs for various complications and arrhythmias commonly encountered in-clinic and discusses appropriate resolution. Currently weighted 10% in the CCDS exam.
True / False
Hyperkalemia typically causes loss of atrial capture before loss ventricular capture.
True.
If atrial loss of capture has occurred due to significantly increased thresholds, suspect hyperkalemia.
If a patient presents with an unusually wide QRS, what metabolic state should you consider?
Hyperkalemia.
True / False
Hyperkalemia decreases stimulation threshold.
False - Hyperkalemia increases stimulation threshold.
When looking at the pacing state histogram.
If total % of Ap state (shown as ApVp + ApVs) is greater than all other states, what can be inferred?
A large amount of pacing is likely at the base rate. Suggestive of SSS or Atrial incompetence.
When looking at pacing state histogram.
If total % of As state (shown as AsVp + AsVs) is greater than all other states, what can be inferred?
- If AsVp is large then the condition is likely CHB or the AVD is programmed too short.
- If AsVs is large then Sinus Rhythm is prevalent and pacemaker is on standby most of the time.
When looking at pacing state histogram.
If total % of Vp state (shown as AsVp + ApVp) is greater than all other states, what can be inferred?
Condition is likely CHB or AVD is programmed too short.
When looking at pacing state histogram.
If total % of Vs state (shown as AsVs + ApVs) is greater than all other states, what can be inferred.
AV conduction is present in this patient.
True / False
A normal P-wave amplitude histogram follows a bell shaped curve distribution.
True.
When viewing a P-wave amplitude histogram, there a two distinct groupings. One grouping is larger in amplitude, while another is smaller in amplitude. There is a void of data in-between the two groupings, what is most likely?
- Signals recorded in the larger amplitude group are likely sinus beats
- Signals recorded in the smaller amplitude group are likely some form of atrial tachycardia
CRT pacing shows dominant RV pacing on an ECG. The patient has CHB, RV and LV thresholds are appropriate and V-V delay is 0ms. What could be occurring at the LV site?
LV exit block or LV conduction delay due to substrate in vicinity of LV electrode.
Describe the rhythm on the IEGM.
Atrial Fibrillation with rapid ventricular response.
R-R variability is distinctive of AF. It is impossible to know 100% however as this is a single chamber device. A dual chamber device would allow true discrimination by visualising atrial activity.
Describe the rhythm on the IEGM.
Atrial tachycardia with 1:1 conduction.
Describe the rhythm on the IEGM.
2:1 AV Block
Describe the rhythm on the IEGM.
Atrial Tachycardia.
Does this patient likely have CHB or not?
Patient likely has CHB.
Note how there is no conduction of P-waves and Vp throughout the strip. This remains true even when an atrial tachycardia starts.
Is this a short duration A-Tach or A-Fib episode?
Short duration A-Tach.
Note how R-R interval is regular. The rate is also too slow to be AF.
Describe the rhythm on the IEGM.
Sinus Rhythm.
There is atrial lead noise respondent to lead damage. Unlikely to be EMI interferrence as this would display on both leads.
You ask a patient to perform provocation manouvers in clinic and witness this. Both leads are programmed Bipolar. Describe what is happening.
Atrial lead noise likely respondent to lead insulation failure.
Source of noise is likely myopotentials. Unipolar sensing would also give rise to myopotential.
Is the device detecting this bipolar lead noise? How best to program around it?
- The atrial lead noise is being detected by the device. This would inhibt output, which could be catastrophic if this was the ventricular lead in a CHB patient.
- Try programming the lead to the unipolar. Retest provocation manouvers in clinic.
- Ultimately a lead revision is necessary.
Which conductor has likely failed in the following biploar leads?
Atrial ring conductor has likely failed.
Rarer for the tip conductor to fracture first. This can be tested by programming the lead unipolar tip to can and asessing the IEGM.
What is the atrial threshold?
Atrial threshold is 1.5V.
Note how a P-wave is evident on the ECG immediately following the 1.5V stimulus but not evident immediately following the 1.4V stimulus. Note also that an intrinsic P-wave breaks through following the 1.4V stimulus, also indicating non capture.
A reversion to 2.4V inbetween is characteristic of Biotronik threshold testing and should be ignored for the purpose of threshold determination.
Describe the rhythm on the IEGM.
Atrial Flutter with variable conduction.
Rhythm is too regular to be AF.
Describe the rhythm on the IEGM.
Flutter with 2:1 conduction.
What can be said about this rate trend?
Highlights some form of sustained atrial arrhythmia at varying rates.
Patient also likely has CHB as there is little ventricular rate deviation during Tach/Sinus.
Yes / No
Is this patient likely to have experienced atrial arrhythmia? If so, which type?
Yes - Atrial Fibrillation.
The top left graph shows atrial rights across the board from 60-400bpm. Due to an even spread its most likely the patient is ecperiencing AF. Flutter or Tach would give bunched As events at a certain rate.
True / False
This patient likely has intrinsic conduction - explain.
False.
Top right graph shows almost all V events are Vp events (white), highly suggestive of CHB. The few Vs events (Black) are likely ventricular ectopics.
The following patient has intrinsic conduction. Whats wrong and how will you correct this?
Patient is almost 100% Vp - extend AV delay or program RVp reduction algorithm.
Not only will this drain battery unnecessarily but will also put patient at risk of pacemaker syndrome.
What is the most likely arrhythmia experienced here?
Most likely AF.
Note how atrial rate is spread out across the whole rate range - thus R-R interval is highly irregular. Also atrial rates above 400bpm are almost always indicative of chaotic firing.
Why are these labeled PVCs when they’re clearly organised at regular intervals?
A device will label any V event a PVC if there is no A event preceeding it.
These could very well be conducted Sinus Beats with undetected P-waves by the device or Junctional beats with no preceeding P-wave. The device wont differentiate and instead labels them PVCs.
What is the likely arrhythmia and does this patient likely have intrinsic conduction?
Afib or Aflutter - No intrinsic conduction.
As this is a daily mean average its difficult to know whether the flutter rate was 250bpm or if the highly variable AF rate averaged to 250bpm.
Look at the V-rate throughout. It is unchanged during/after arrhythmia. Highly suspicious of CHB.
What is the RV threshold?
0.9V.
The V sensed events after 1.2V & 1.1V are ectopic beats. Clear capture can be shown at lower voltages. The stimulus of 0.8V elicits no resonse.
The device registers sinus tachycardia - is this correct?
Yes.
Narrow farfield complex, rates ~150bpm and 1:1 A:V relationship are indicitave of Sinus Tachycardia.
Onset would be helpful here as gradual onset = Sinus // Sudden onset = Atrial Tachycardia.
What is the RV threshold?
0.6V.
V sensed events following 0.9V stim are likely ectopic beats, clear capture is evidenced at lower voltages.
What is the rhythm shown?
Sinus Rhythm with 1:1 conduction.
The ‘extra beat’ seen on the atrial channel is infact farfield V sensing. It is likely the Atrial lead is placed close to the tricuspid valve and thus ‘sees’ ventricular activity.
Why are some of the atrial events labeled with ARs (FFP) and others not?
ARs (Far Field Protection) = PVARP
Those without markers = PVAB
Simply those without markers have fallen closer to the preceeding V event and fallen within the blanking period, not the refractory period. Remember the device sees things in the refractory (but doesnt act upon it) and is blind to everything within the blanking period.
Yes / No
The device is labeling the second atrial ‘Ars’ event with an amplitude - is it only measuring this impluse and missing the first?
No.
The device is not measuring the far field atrial signal, its just the number is formatted to appear in that part of the screen and the far field signal coincidenally lines up (check the V amplitude). The first atrial event is being sensed (and measured) appropriately as seen at the top by the As marker.
What is the likely arrhythmia shown by the IEGM.
VT.
More V than A events, Sudden onset of fast V rate with atrial dissociation, Farfield morphology change.
True / False
The following shows Ventricular Fibrillation.
False.
Ventricular R-R interval is stable, Ventricular amplitude is stable, Farfield morphology change
No atrial channel so this could (unlikely) be atrial tachycardia with 1:1 conduction
True / False
The following shows Atrial Tachycardia.
False - Most likely Ventricular Tachycardia.
The atrial rate is stable at approximately 90-100bpm
V rate is ~150bpm, R-R interval is regular, Broad farfield morphology
What arrhythmia is starting here?
Most likely Atrial Fibrillation
Unstable A-A intervals and varying amplitudes are suggestive of AF
Why was the circled Ap delivered despite there being multiple atrial events preceeding it?
The preceeding 3 atrial events were not counted due to either falling within the blanking or refractory periods of the atrial channel.
Yes / No
Did the circled Ap event capture, why?
No - fell too soon after a preceeding A-event.
Most likely within the refractory period of the atria, thus unable to initiate a contraction.
What does the IEGM show?
Atrial Tachycardia with 1:1 conduction.
You see this in clinic - what is it and what would you do?
Atrial lead noise giving rise to oversensing & inappropriate Mswitch. Immediate doctor consutation.
What is the Atrial threshold?
THR = 0.8V. Note how there are P-waves in lead II and II at 0.8V stimuation, but not post 0.7V.
Also see how there is intrinsic atrial breakthrough following the 0.7V stimulation, further alluding to non-capture. The IEGM signal post 0.8V looks different as it is likely a fusion beat.
What behaviour is inappropriate here?
Mswitch has incorrectly resolved as atrial arrhythmia hasn’t terminated.
This gives rise to the increase in ventricular rate seen until the Mswitch is again activated - at which point the ventricular rate drops back to the base rate.
What does the IEGM show?
Dropped p-waves giving rise to wenckebach behaviour
P-waves are dropped due to undersensing. Note how dropped p-waves are significantly smaller in amplitude (size on IEGM) than those appropriately sensed and tracked.
True / False
This shows successful ATP.
False.
Note how tachycardia continues afterwards. Note also how farfield morphology and cycle length is unchanged - thus ATP did not enter the tachycardia circuit and alter the arrhythmia in any way.
What rhythm does the IEGM show?
Sinus Rhythm with long 1st degree AVB
Why are the V-events labelled ‘PVC’ when there are clearly A events preceeding them?
The AV delay is grossly extended (480msec)
Thus the device no longer counts the Atrial event as associated to the Ventricular event.
What is the LV threshold?
1.3V
Which is the affected lead and what is the likely diagnosis?
LV lead fracture - lead displays make/break connectivity.
High impedance = more likely fracture.
Low impedance = more likely insulation failure.
True / False
The IEGM shows atrial flutter and CHB - normal function.
False - RV lead shows inappropriate sensing due to noise, thus abnormal function.
The rhythm is indeed atrial flutter with CHB.
What rhythm is shown here, anything unusual?
Likely Sinus Rhythm (regular R-R interval) - RV shows sensed lead noise.
Note how this is counting towards VF and VT counters. If left untreated this could lead to inappropriate therapy.
What type of ATP is this? Was it successful?
Successful burst ATP.
Note rate and far-field electrogram morphology changes pre/post ATP.
True / False
The patients underlying rhythm is atrial fibrillation with CHB
False - Atrial Flutter (regular P-P interval & amplitude) with CHB
What is the most likley rhythm displayed here?
Predominantly Atrial flutter (80%) with periods of Sinus p-waves (20%) with CHB throughout.
As grouping on left represents sinus activity, grouping on right represents flutter. Unlikely atrial fibrillation as the grouping is too tight (210-290bpm). Atrial fibrillation would be much wider across all rate ranges.
CHB as patient is 100% Vp.
Why does the IEGM show Ars (FFP) after every paced beat?
Retrograde P-waves are being conducted and sensed within the PVAB.
What is the likely arrhythmia shown here?
Likely AF
Note how the atrial rate is spread across all rate ranges, highlighting irregularity. Flutter would be far more concentrated to a few rate ranges.
What is the likely atrial arrhythmia? Is this conducted to the ventricles?
AF with CHB thus no intrinsic conduction.
Note 100% Vp thus most likely complete heart block patient
What was the most likley indication for this patient?
Chronotropic incompetence - majority of atrial events are paced.
Note how all ventricular events are sensed - intrinsic conduction is present.
Note also how there is evidence of atrial flutter with 20% of Atrial beats >250bpm.
True / False
The most likely rhythm is Atrial Flutter with CHB.
True.
100% Vpacing indicates no intrinsic conduction.
Atrial events are grouped tightly (210-290bpm) suggestive of flutter. Atrial Fibrillation would be across the whole rate range.
What was the shocked rhythm and was the shock successful?
VF - Chaotic & irregular. Successfully terminated with 550V shock.
True / False
This patient has intact VA conduction.
True.
Note Ars (FFP) marker post V-events. This is a retrograde P-wave being sensed by the device and falling within the PVAB.
True / False
The rhythm is most likely atrial flutter.
False.
Irregular P-P intervals and alternating p-wave amplitudes alludes to this being Atrial Fibrillation.
True / False
The rhythm is most likely Sinus Tachycardia.
True.
Rhythm shows 1:1 relationship of A and V events. AV and VA intervals remain fixed.
VT / AF with RVR
What rhythm is the most likely diagnosis?
AF with Rapid Ventricular Response.
Atrial fibrillation is clearly shown on the atrial channel. Ventricular response is both rapid and irregular - most likely in response to chaotic atrial intervals being conducted. VT would show far greater R-R stability.
True / False
The following patient doesn’t exhibit ventricular conduciton.
False.
Approximately 55% of all ventricular activity is sensed - thus intrinsic conduction must be present, most of the time.