Art Of Imaging Critique/ Quality Managent Flashcards

(69 cards)

1
Q

What are the three steps in an effective image critique?

A
  1. Classification of the image
  2. Determination of the cause of the problem
  3. Recommendation of the corrective action
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What is the first step when looking at an image? The radiographer needs to determine if the image is

A

Acceptable or unacceptable

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

When determining the cause of the problem, there are three categories to classify an issue. What are those three major categories?

A

1.Technical problem
2. Procedural factor problem
3. Equipment malfunction problem

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Factors that influence the cause of the problem:
Technical Factors

A

IR exposure
Exposure Indicator
Contrast
Detail
Distortion

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Factors that influence the cause of the problem:
Procedural Factors

A

Patient Preparation
Patient Position

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Factors that influence the cause of the problem:
Equipment Malfunction

A

Processing
Radiographic tube/table
Fluoroscopic tube/table

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Quality assurance includes each of the following…

A
  1. Equipment maintenance
  2. Radiology reports accuracy
  3. Scheduling
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

The larges hospital accreditation agency is the…

A

The Joint Commission

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

The Joint Commission approval is linked to ____ reimbursements.

A
  1. Federal
  2. State
  3. Insurance company
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Input for the purchase of new radiographic equipment should include the…

A
  1. Radiologists
  2. Radiographers
  3. Administrators
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Installation and acceptance testing of equipment is the responsibility of the:

A

Radiology equipment vendor

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Routine quality management procedures should be preformed by the:

A

Qualified radiographer

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

What quality control should be done on a regular, ongoing basis?

A

Image repeat analysis

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Each of the following monitoring tests should be done on a semiannual basis:

A

Focal spot size estimation
Collimator accuracy
Kilovoltage Accuracy

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Test tools for focal spot size estimation include each of the following

A

Pinhole camera
Star test pattern
Line pair resolution

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Focal spots measuring 0.3 mm or less should be measured with a

A

Start test pattern

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

Focal spots measuring 0.3 mm or more should be measured with a

A

Pinhole camera

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

Which type of equipment is used to measure half-value layers?

A

DUKE chest phantom

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

When testing collimator accuracy, an SID error of ____ is allowed.

A

2%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

The centering mark should be within ____ of the light field central ray.

A

1%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

Distance indicators are checked with a

A

Tape measure

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

A generator must be recalibrated when the kVp drifts beyond ____ kVp of the labeled setting.

A

5

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

The accuracy of single-phase generator time settings can be measured with a

A

Computerize dosimeter

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

A thorough reject analysis program would include data collection on images produced

A

By each technologist
In each radiographic room
By service engineers

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
Exposure time settings should be maintained within ____ of the labeled setting.
+ or - 5 percent
26
The source-to-skin distance limit for a stationary fluoroscopic unit is ____ inches.
15
27
Tomographic system uniformity and completeness of motion are tested with a
lead mask with a pinhole.
28
Tomographic systems should be tested for each of the following
Completeness of motion Section depth indicator accuracy Section thickness accuracy
29
Most AECs utilize a ____ ion chamber combination.
Three
30
Computerized storage and transmission systems for digitized images of all types are called
PACS
31
The accuracy of mA stations on a radiographic generator:
1. Should be measured after the timer and kVp accuracy have been confirmed 2. tend to drift as the tube ages 3. Are measured using mR/ mAs calculations across all mA stations
32
You have just taken a CR radiograph of a patient’s knee and the S# is 225. Unfortunately, the image has some degree of motion distortion and needs to be repeated using the same exposure settings. In repeating the image, the new image shows no motion but the S# is now 400. No changes were made to collimation, patient position, exposure techniques, etc. Based upon this new S# value,
Exposure reproducibility is suspect
33
A qualified x-ray service engineer should be used to evaluate and correct the following radiographic/fluoroscopic imaging features:
automatic brightness control fluoroscopic contrast image display monitor brightness/contrast AEC reproducibility
34
Digital detectors with a higher DQE will
produce better images of small objects with low subject contrast
35
A common device to measure display monitor quality and characteristics is a/an
luminance meter
36
An effective method to assess department image quality is to perform a repeat image analysis. This procedure will demonstrate which of the following department traits?
percent of repeat images per room percent of repeat images per radiographer
37
True or false: The art of image critique is the application of scientific knowledge to the analysis of an image.
True
38
Image critique is particularly challenging with current technology due to the problems centering around
the sophistication of the digital technologies used.
39
Generally speaking, the process of image analysis should start with
a review of the entire image
40
With film-screen radiography, the largest percentage of image quality problems center around
automatic processing.
41
As medical imaging continues to adopt digital imaging technologies, one of the greatest challenges for technologists regarding image analysis will be
clearly identifying the problem as a photographic vs. geometric problem with detail. identifying possible causes that may be electronic in nature. determining if postprocessing algorithms are accurate.
42
All of the following are related EXCEPT: exposure timer error. technical factor problem. equipment malfunction problem. procedural factor problem.
exposure timer error.
43
All of the following would be classified as technical factor problems EXCEPT
the presence of residual barium in the patient’s colon scheduled for an IVU exam.
44
Images that are grossly underexposed
will possess poor visibility of detail
45
Problems with radiographic/fluoroscopic equipment image quality can be kept to a minimum by
a quality control program
46
Static marks on a film radiographic image represents
a visibility of detail problem.
47
The presence of the Moire effect on an image indicates a problem with
a CR cassette and grid.
48
A photographic problem with visibility of detail would be likely due to
inadequate kVp
49
If the electronic timer of a high-frequency generator is suspect,
a qualified service engineer should be called.
50
A film-screen image that demonstrates too little contrast would represent
a technical factor problem.
51
An image is presented to the department quality control coordinator by a disgruntled radiologist that shows an asymmetrically narrowed ankle mortis on an AP projection. His dissatisfaction would be most likely classified as a
procedural factor patient positioning problem
52
Geometric problems dealing with distortion could be due to
x-ray tube alignment
53
Technical factor problems can include both
photographic and geometric problems.
54
An optimal diagnostic quality radiograph
is always within acceptable limits.
55
The three alignment factors include the
x-ray tube, part, and receptor.
56
True or False: Equipment malfunction involves radiographic/fluoroscopic equipment only.
false
57
You have just completed an AP projection of the knee using a DR detector. On review of the image, the image demonstrates a series of several, thin white lines across the bottom third of the image. One of your colleagues immediately states that the image problem is due to grid cut-off and tells you to repeat it without the use of the grid. Your response should be to
analyze the image further to look for nongrid technical issues.
58
An effective and reliable assessment of DR image quality, from an exposure intensity standpoint, is to
look at the Exposure Index(EI) number and reference the EI target numbers.
59
As you are performing a KUB on a 375 lb. male patient using AEC, the exposure terminates within 1–2 milliseconds. The DR image displayed following exposure demonstrates objectionable quantum noise. The likely cause can be classified as a(n)
equipment malfunction problem procedural problem technical problem
60
The presence of two opaque metallic objects in the region of C1–C2 on a lateral cervical spine projection is an example of
a procedural issue
61
A radiologist who complains about an inability to assess heart size on a PA chest radiograph because of patient rotation, is likely upset about
a procedural problem.
62
63
possible ethical ALARA violation
an S# of 50
64
AEC hits back-up mAs setting
barium still remaining in small bowel
65
possible explanation of image quality cues
hypothesis
66
equipment malfunction problem
presence of grid lines on image
67
poor visibility of bony trabeculae
small focal spot not selected
68
residual feces in bowel
improper patient preparation
69
image receptor alignment problem
image cut-off along the long dimension of a cassette