Chapter 36 Flashcards

1
Q

Leakage radiation must be less then ____mR/hr at a distance of 1 m from the protective housing

A

100

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

The SID indicator must be accurate within __% of the indicated SID.

A

2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

The x-ray beam and the light must beam must coincide within ___% of the SID

A

2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

The PBL must be accurate to within ___ of the SID

A

2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

The variation in x-ray intensity should not exceed ___%

A

5

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

The maximum acceptable variation in linearity is ___% from one mA station to an adjacent mA station.

A

10

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

The SSD must not be less than ___cm on stationary fluroscopes and not less then ___ cm on mobile fluroscope.

A

38, 30

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

___ time produces an audible signal when the fluroscopc time has exceeded 5 minutes

A

cumulative

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What device is used to cover the opening on the side of a fluroscopic table when the bucky tray is moved out the way?

A

bucky slot cover

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

A ___ of at least 0.25mm Pb equivalent should be positioned between the fluroscopist and the patient

A

protective curtain

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

____ radiation is the useful beam.

A

primary

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

What are the two types of secondary radiation:

A

scatter and leakage

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

The intensity of scatter radiation 1 m from the patient is approximately __% of the intensity of the useful beam at the patient.

A

0.1

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

When PBL is activated, the collimators are automatically adjusted so that the radiation field matches the ____.

A

film size

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

What are the two type of filtration?

A

inherent and added

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

The glass envelope encasing the

x-ray tube, the insulating oil, and the glass window are examples of what type of filtration?

A

inherent

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

Inherent materials amounts to approximately __mm of aluminum equivalent.

A

0.5

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

A sheet of aluminum or its equivalent of appropriate thickness located outside the glass window of the tube housing above the collimator shutters are examples of what type of filtration?

19
Q

Added filtration is __ mm of aluminum equivalent

20
Q

The inherent and added filtration should combine to equal the required ___ of aluminum equivalent. (fixed tube, above 70kVp)

21
Q

The variation in the x-ray intensity should not exceed _%

22
Q

The maximum acceptable variation in linearity is __% from one mA station to an adjacent mA station

23
Q

The x-ray control panel must show _____.

a. beam on time
b. mAs used
c. positive beam limitation
d. automatic exposure controls

24
Q

What is the minimum requirement for filtration on x-ray equipment operating above 70 kVp?

a. 1.5 mm Al equivalent
b. 2.0 mm Al equivalent
c. 2.5 mm Al equivalent
d. 3.5 mm Al equivalent

25
* The output intensities of reproduced exposures should not vary more than _____% from each other. a. 5 b. 10 c. 15 d. 20
A
26
* Measuring the relationship between the output intensities of adjacent mA stations is a test of _____. a. HVL b. beam limitation c. reproducibility d. linearity
D
27
* The source-to-skin distance must be no less than _____ cm on stationary fluoroscopes. a. 30 b. 38 c. 42 d. 48
B
28
* The _____ is the primary protective barrier for the fluoroscopic x-ray tube. a. ceiling b. floor c. image intensifier d. table top
C
29
A fluoroscopy system with automatic collimation must provide an unexposed border around the image at _____ above the table top. a. all heights b. 35 cm c. 25 cm d. 18 cm
A
30
* The fluoroscopic exposure control switch is always a(n) _____ type. a. rheostat b. automatic c. deadman d. on-off
C
31
* The bucky slot opening in the side of the table must be automatically covered with _____ mm Pb during fluoroscopy. a. 1.25 b. 1.0 c. 0.5 d. 0.25
D
32
* A cumulative timer is designed to _____ of fluoroscopy beam-on time. a. make the radiologist aware b. turn off the radiation every 5 minutes c. prevent more than 10 minutes d. shut down the system after 15 minutes
A
33
Primary radiation is _____. a. leakage radiation b. the scattered beam c. the useful beam d. all of the above
C
34
The intensity of scatter radiation 1 meter from the patient is approximately _____ of the intensity of the incident beam at the patient. a. 0.01% b. 0.1% c. 10% d. 50%
b
35
At the level of the tabletop, the intensity of the fluoroscopic beam should not exceed _____ for each mA of operation at 80 kVp. a. .02 mGy/hr b. 0.2 mGy/hr c. 21 mGy/min d. 0.21 mGy/min
C
36
* A controlled area is one occupied primarily by _____. a. radiology personnel b. patients c. the general public d. both A and B
D
37
* A secondary barrier is designed to absorb _____ radiation a. leakage b. scattered c. primary d. both A and B
D
38
* A quantity that reflects both dose and volume of tissue irradiated is the _____. a. estimated skin exposure b. dose area product c. proportional region d. Rem
B
39
* The design of an uncontrolled area in radiology is based on the NCRP recommended limit of _____ mGy/yr for the public. a. 50 b. 10 c. 1 d. 0.5
C
40
* Devices designed to measure total accumulated radiation intensity are called _____. a. dosimeters b. integrators c. scintillation detectors d. photo cathodes
A
41
* Gas-filled radiation detectors are based on the _____. a. thermoluminescence of crystals b. photoemission of electrons c. ionization of gas d. scintillation of phosphors
C
42
* The _____ is used for precise calibration of the output intensity of diagnostic x-ray tubes. a. G-M counter b. ion chamber c. scintillation detector d. proportional counter
B
43
* A Geiger-Muller counter is a(n) _____. a. scintillation detector b. optically stimulated dosimeter c. gas-filled detector d. thermoluminescent dosimeter
C
44
``` Patient dose can be reduced by using___ I. faster screen II. higher kvp III. increased distance a. I b. II only c. II & III d. I, II, and III ```
a