Clinical Chemistry Flashcards
(498 cards)
A solution that has a transmittance of 1.0 %T would have an absorbance of:
A. 1.0
B. 2.0
C. 1%
D. 99%
B.
Absorbance= 2-log%T
A= 2-log 1
A= 2
Which formula correctly describes the relationship between absorbance and %T ?
A. A = 2 – log %T
B. A = log 1/T
C. A = –log T
D. All of these options
D
In absorption spectrophotometry:
A. Absorbance is directly proportional to transmittance
B. Percent transmittance is directly proportional to concentration
C. Percent transmittance is directly proportional to the light path length
D. Absorbance is directly proportional to concentration
D
Which wavelength would be absorbed strongly by a red-colored solution?
A. 450 nm
B. 585 nm
C. 600 nm
D. 650 nm
A
Red colored solution TRANSMITS light proportional to its wavelength but ABSORBS light of wavelength complementary to its color.
A green-colored solution would show highest transmittance at:
A. 475 nm
B. 525 nm
C. 585 nm
D. 620 nm
B
A technologist is performing an enzyme assay at 340 nm using a visible-range spectrophotometer. After setting the wavelength and adjusting the readout to zero %T with the light path blocked, a cuvette with deionized water is inserted. With the light path fully open and the 100%T control at maximum, the instrument readout will not rise above 90%T. What is the most appropriate first course of action?
A. Replace the source lamp
B. Insert a wider cuvette into the light path
C. Measure the voltage across the lamp terminals
D. Replace the instrument fuse
A
Which type of monochromator produces the purest monochromatic light in the UV range?
A. A diffraction grating and a fixed exit slit
B. A sharp cutoff filter and a variable exit slit
C. Interference filters and a variable exit slit
D. A prism and a variable exit slit
D
Diffraction grating and prisms both produce a continous range of wavelength but using a prism and a variable exit slit can produce a UV light of a very narrow bandpass
Which monochromator specification is required in order to measure the true absorbance of a compound having a natural absorption bandwidth of 30 nm?
A. 50-nm bandpass
B. 25-nm bandpass
C. 15-nm bandpass
D. 5-nm bandpass
D
A bandpass refers to range of wavelengths passing through the sample. The narrower the bandpass the better the resolution. Accurate absorbance measurement requires a bandpass of LESS THAN 1/5 OF THE NATURAL BANDPASS OF CHROMOPHORE.
Which photodetector is most sensitive to low levels of light?
A. Barrier layer cell
B. Photodiode
C. Diode array
D. Photomultiplier tube
D
Which condition is a common cause of stray light?
A. Unstable source lamp voltage
B. Improper wavelength calibration
C. Dispersion from second-order spectra
D. Misaligned source lamp
C
Which type of filter is best for measuring stray light?
A. Wratten
B. Didymium
C. Sharp cutoff
D. Neutral density
C
1Which of the following materials is best suited for verifying the wavelength calibration of a spectrophotometer?
A. Neutral density filters
B. Potassium dichromate solutions traceable to the National Bureau of Standards reference
C. Wratten filters
D. Holmium oxide glass
D
Why do many optical systems in chemistry analyzers utilize a reference light path?
A. To increase the sensitivity of the measurement
B. To minimize error caused by source lamp fluctuation
C. To obviate the need for wavelength adjustment
D. To reduce stray light effects
B
Which component is required in a spectrophotometer in order to produce a spectral absorbance curve?
A. Multiple monochromators
B. A reference optical beam
C. Photodiode array
D. Laser light source
C
The half-band width of a monochromator is defined by:
A. The range of wavelengths passed at 50% maximum transmittance
B. One-half the lowest wavelength of optical purity
C. The wavelength of peak transmittance
D. One-half the wavelength of peak absorbance
A
The reagent blank corrects for absorbance caused by:
A. The color of reagents
B. Sample turbidity
C. Bilirubin and hemolysis
D. All of these options
A
A plasma sample is hemolyzed and turbid. What is required to perform a sample blank in order to correct the measurement for the intrinsic absorbance of the sample when performing a spectrophotometric assay?
A. Substitute deionized water for the sample
B. Dilute the sample 1:2 with a standard of known concentration
C. Substitute saline for the reagent
D. Use a larger volume of the sample
C
Which instrument requires a highly regulated DC power supply?
A. A spectrophotometer with a barrier layer cell
B. A colorimeter with multilayer interference filters
C. A spectrophotometer with a photomultiplier tube
D. A densitometer with a photodiode detector
C
Which statement regarding reflectometry is true?
A. The relation between reflectance density and concentration is linear
B. Single-point calibration can be used to determine concentration
C. 100% reflectance is set with an opaque film called a white reference
D. The diode array is the photodetector of choice
C
Bichromatic measurement of absorbance can correct for interfering substances if:
A. The contribution of the interferent to absorbance is the same at both wavelengths
B. Both wavelengths pass through the sample simultaneously
C. The side band is a harmonic of the primary wavelength
D. The chromogen has the same absorbance at both wavelengths
A
Which instrument requires a primary and secondary monochromator?
A. Spectrophotometer
B. Atomic absorption spectrophotometer
C. Fluorometer
D. Nephelometer
C
Which of the following statements about fluorometry is accurate?
A. Fluorometry is less sensitive than spectrophotometry
B. Fluorometry is less specific than spectrophotometry
C. Unsaturated cyclic molecules are often fluorescent
D. Fluorescence is directly proportional to temperature
C
Which of the following components is not needed in a chemiluminescent immunoassay analyzer?
A. Source lamp
B. Monochromator
C. Photodetector
D. Wash station
A
Which substance is used to generate the light signal in electrochemiluminescence?
A. Acridinium
B. Luminol
C. Dioxetane phosphate
D. Ruthenium
D