Dosi 1 Flashcards
(156 cards)
Concerning the precision of treatment planning and delivery of treatment in radiation therapy, the greatest source of uncertainty is:
•The histology of the disease.
•Physical response data.
•Patient and organ motion during treatment.
•Linear accelerator calibration.
•Patient and organ motion during treatment.
An unsealed chamber used as a secondary dosimeter gives a reading of 101.0 for a 100 monitor unit exposure at a field size of 10 cm2. If the ambient temperature is 25 degrees C and the atmospheric pressure is 770 mm Hg, the corrected reading per monitor unit is:
A. 0.923.
B. 0.997.
C. 1.007.
D. 1.013
C. 1.007
HgP std =760mmHg
Pactual =770mmHg
Tstd=0°C =273.15 K
T actual=25°C=298.15
CorrectionFactor= (770/760)×( 273.15/298.15)
A patient is treated by using parällel opposed equally weighted fields at 80 cm source-to-surface distance (SSD) on a 4 MV linear accelerator to a total midplane dose of 60 Gy. The patient diameter is 20 cm. The field size is 15 x 15 cm2. The total dose at a depth of 1.2 cm is:
Depth (cm). 15 x 15 cm
1.2. 100.0
5. 83.7.
10. 63.6
15. 46.9
20. 34.1
25.0. 25
A. 62.6 Gy.
B. 64.7 Gy.
C. 66.5 Gy.
D. 68.1 Gy.
B. 64.7 Gy
AP=30 ( 100/63.6)
PA=30(37.1/63.6)
The correct source-to-tray distance on the therapy machine is 50 cm. If a block is constructed with the tray distance set at 40 cm, which of the following will result?
A. The blocked area will be too small.
B. The block could be in the wrong position on the tray.
C. The block divergence will not match the beam divergence.
D. The block will be too far from the patient.
A. The blocked area will be too small.
For a 4-field box technique, which of the following will deliver to femoral heads?
A.100 SSD, equal max dose to all fields
B.100 SSD, equal target dose to all fields
C.100 SAD, equal air dose to all fields
D.100 SAD, equal target dose to all fields
E.100 SAD, equal monitor unit setting for all fields
A.100 SSD, equal max dose to all fields
An individual receives a dose of 0.1 cGy from a gamma source and 0.05 cGy from neutrons.
The dose equivalent from both radiation sources IS:
A. 1.1 mSv
B. 1.5 mSv
C. 8.0 mSv
D. 11.0 mSv
11mSV
Gamma dose equivalent:
0.001Gy × 1= 0.001 Sv
Neutron dose equivalent: 0.0005Gy×20=0.01Sv
Step 3: Add the dose equivalents.
Total dose equivalent = 0.001 Sv+0.01Sv=
0.011Sv=11mSv
0.001Sv+0.01Sv=0.011Sv
The accelerator component that injects electrons into the linear accelerator is:
klystron
electron gun
high voltage power supply bending magnet
thyratron
electron gun
Which of the following will change the buildup in a high energy beam?
(1) use of large field sizes on a linear accelerator with resulting electron contamination
(2) placement of the diaphragm or treatment cone greater than 20 cm from the skin surface
(3) placement of bolus material on the skin surface.
(A) (1), (2), and (3) only are correct.
(B) (1) and (3) only are correct.
(C) (2) and (4) only are correct.
(D) (4) only is correct.
(E) All are correct.
(A) (1), (2), and (3) only are correct.
Tissue-air ratio (TAR) is dependent on:
(1) beam energy
(2) depth
(3) field size
(4) source-surface distance
(A) (1), (2), and (3) only are correct.
(B) (1) and (3) only are correct.
(C) (2) and (4) only are correct.
(D) (4) only is correct.
(E) All are correct.
(A) (1), (2), and (3) only are correct.
When the equivalent square of a rectangular field is calculated, the result is less accurate when the:
(A) sides of the rectangular field are not vastly different.
(B) areas of the two fields are similar.
(C) field has an elongation ratio of 2
(D) beam energy is greater than 6 MV.
(E) perimeter of the two fields is the same.
C field has an elongation ratio of 2
Parallel opposing fields are used to deliver 5000 cGy in 25 fractions to the isocenter, located at mid-depth of 10 cm. The TAR at zero field size at 10 cm depth is 0.534 and the average SAR after irregular field calculation is 0.195. The tray factor is 0.939, air dose rate is 219.2 cGy/min and the timer correction is +0.02 min. Calculate the timer setting for each field.
(A) 0.56 min
(B) 0.65 min
(C) 0.69 min
(D) 1.12 min
(E) 1.35 min
C) 0.69 min
A 4 MV linear accelerator is calibrated to deliver 1 cGy per monitor unit in a phantom at depth with 80 cm SSD and 10 x 10 cm field size. [SEE TABLES] The dose rate to a
(A) 0.990 cGy per monitor unit.
(B) 0.996 cGy per monitor unit.
(C) 1.000 cGy per monitor unit.
(D) 1.005 cGy per monitor unit.
(E) 1.020 cGy per monitor unit.
(D) 1.005 cGy per monitor unit.
An orthovoltage beam is used to deliver 350 cGy to a depth of 5 cm. Calculate the timer setting for an SSD of 52 cm given the following information:
Exposure rate = 128 R/min at 50 cm SSD “cGy/R” factor = 0.95 cGy/R
Backscatter factor = 1.17
% depth dose at 5 cm = 63%
Timer error = 0
A.3.61
B.3.71
C 4.06
D.4.22
E.4.34
D. 4.22
A patient is prescribed for 70.20 Gy in 39 fractions, using a bilateral arc technique. The output factor equals 1.00 cGy/MU. Using the data provided below, the monitor unit setting for the left arc is:
RIGHT ARC Gantry
Angle TMR
210. .748
230 .672
250 .587
270 .563
290 .591
310 .672
330 .728
LEFT ARC Gantry
Angle TMR
30 .752
50 .679
70 .591
90 .563
110 .587
130 .672
150 .748
(A) 107
(B) 137
(C) 141
(D) 274
(E) 282.
(B) 137
Which of the following represents the ratio of dose at a given point in a phantom to the dose at the same point at depth of maximum in the phantom?
(A) scatter-air ratio
(B) off-axis ratio
(C) tissue-phantom ratio
(D) tissue-air ratio
(E) tissue-maximum ratio
(E) tissue-maximum ratio
Which of the following represents the ratio of dose at a given point in a phantom to the dose at the same point at depth of maximum in the phantom?
(A) scatter-air ratio
(B) off-axis ratio
(C) tissue-phantom ratio
(D) tissue-air ratio
(E) tissue-maximum ratio
(E) tissue-maximum ratio
A patient is supine for treatment of the thoracic spine from a posterior field 7 cm wide, 25 cm long at 100 cm SAD, 94 cm SSD. This field must be matched at the isocenter to a previously treated posterior spine field 7 cm wide and 15 cm long also at 100 cm SAD, 94 cm SSD. The patient’s anterior posterior separation throughout the treatment region is 23 cm. The gap between the fields measured on the patient’s anterior skin is:
(A) 1.2 cm
(B) 2.4 cm
(C) 3.4 cm
(D) 4.0 cm
(E) 6.8 cm
A. 1.2
Gap formula
(L1/2 * d/SAD) + ( L2/2 * d/SAD)
(25/2 * 6/100)+( 15/2*6/100)
0.75+0.45= 1.2
Which of the following represents the ratio of the scatter dose at a given point in the phantom to a dose in free space at the same point?
(A) scatter-air ratio
(B) off-axis ratio
(C) tissue-phantom ratio
(D) tissue-air ratio
(E) tissue-maximum ratio
(A) scatter-air ratio
The most radiosensitive phase for mental retardation in the fetus from RT accidental exposure is:
A. 0-7 weeks.
B. 8-15 weeks.
C. 16-24 weeks.
D. more than 25 weeks.
B. 8-15 weeks.
The most radiosensitive phase for mental retardation in the fetus from RT accidental exposure is:
A. 0-7 weeks.
B. 8-15 weeks.
C. 16-24 weeks.
D. more than 25 weeks.
B. 8-15 weeks.
A radioactive source is considered to be isotropic if it exhibits which of the following characteristics?
A. Summation of the total dose is proportional to the point source.
B. Dose at depth is variable in all directions based on the placement of the source.
C. Intensity around the source is equal in all directions.
D. The total dose is a function of the size of the source calculated based on kerma rate.
C. Intensity around the source is equal in all directions
For which of the following malignancies has hypofractionation radiation therapy been shown to significantly improve local control compared with standard fractionation?
A. low-grade sarcomas
B. adult high-grade gliomas
C. head and neck cancer
D. inflammatory breast cancer
C. head and neck cancer
As per the American Brachytherapy Society guidelines, the dose required to treat uveal melanomas with Iodine-125 plaque brachytherapy is:
A. 50Gy prescribed to 0.5cm depth.
B. 65Gy prescribed to apex.
C. 85Gy prescribed to apex.
D. 100 Gy prescribed to 0.5cm depth.
C. 85Gy prescribed to apex.
Which of the following statements is true for clinical electron beams?
A. The mean energy decreases as a quadratic function of depth.
B. The maximum energy is the same as the most probable energy.
C. The surface dose is inversely proportional to the electron beam energy.
D. The most probable energy is a function of the practical range (Rp).
D. The most probable energy is a function of the practical range (Rp).