Equipment Flashcards
(40 cards)
Infrared absorption analyzer
- What gases are measured & what test is it most commonly used for?
- What other 4 terms could be used to refer to this analyzer?
- Carbon monoxide
-carbon dioxide
-methane.
Most commonly used for DLCO. - chopper motor
chopper blade
infrared source
Radiation beam
INFRARED ABSORPTION ANALYZER
- What are the gas concentrations used for calibration?
- What is the process for calibration?
- What should the non-linear display be?
- What is an acceptable phase delay?
- Low end 0%. High end 0.3%.
Use room air for low end calibration. - Zero
Adjust the gain to match the known concentration.
Re-zero - 0.5% or less
- 0.4 seconds
INFRARED ABSORPTION ANALYZER
- How does it work?
- What are 4 key features?
- A small motor rotates an interrupter or chopper.
Which interprets the radiation beam in rhythmic pattern.
Which is proportional to the percentage of gas. - Allows for breath to breath analysis.
-Very fast.
-Not affected by partial pressure (altitude).
-Need to use a scrubber in line before the gas reaches the analyzer so CO2 does not affect the results.
INFRARED ABSORPTION ANALYZER TROUBLESHOOTING
- What are 2 causes of inconsistent readings?
- What tubing should be changed that will stabilize water vapor?
- What would a chattering noise indicate?
- What could a long sample line cause?
- Moisture issues. Moisture affects the radiation beam.
-Failing chopper motor and bearings. - Nafion tubing.
- Sticky demand valve.
- Delayed response time.
GAS CHROMATOGRAPHY
(ABSORPTION TYPE ANALYZER)
- What gases does it measure?
- How does it work?
- What type of analyzer measures the gas concentrations as they exit?
- Neon
-oxygen
-Nitrogen
-Carbon monoxide - The sample gas moves through a separator column
-each gas molecule has differing sizes
-the molecules for the individual gases exit through the column at different rates. - Thermal conductive type analyzer.
-measured as a gas concentration
GAS CHROMATOGRAPHY
(ABSORPTION TYPE)
- What are advantages of this analyzer?
- What are disadvantages?
- Does it require the use of a scrubber and Nafion tubing and why?
- What 3 factors does the analysis time depend upon?
- It can analyze many gases at one time.
- Requires more time for analysis
-cannot measure breath to breath. -Not fast. - Yes, it needs a scrubber and Nafion tubing as it is affected by water and CO2.
- Carrier gas flow.
-column length
-column temperature
MASS SPECTROMETRY
- What gases does it measure?
- How does it work?
- What are other terms that can refer to this analyzer?
- Helium.
-Nitrogen
-oxygen
-carbon dioxide - Sample is drawn in to an ionization chamber by a vacuum.
-the individual gases are separated and analyzed
-measured as a gas concentration - Ionization chamber.
-vacuum pump
MASS SPECTROMETRY
- What are the advantages?
- What are the disadvantages?
- Capable of breath by breath analysis
-allows for multiple gas analysis
-Fast - Very expensive.
-Large in size
-Requires a high degree of maintenanc
-can be affected by condensation
-used mainly in research
MASS SPECTROMETRY
(TROUBLESHOOTING)
- When used during stress testing, what causes a delayed response time for oxygen?
- If the sum of all gases analyzed is less than 100%, what must be done?
- Obstructed sample line.
- The overall game must be adjusted until the sun reaches 100%.
HELIUM ANALYZER
- What are other names for this analyzer?
- What testing is this used for?
- How do you calibrate and what gas percentages?
- Do not use with flammable gases.
- Thermal conductivity analyzer
-Wheatstone Bridge. - Helium dilution.
- Low 0%. Room air
-High 10% helium.
- zero, allow helium to clear, then re-zero
NITROGEN ANALYZER
- What is the name of this analyzer?
- What test is this analyzer used for?
- Other names used that would refer to this analyzer?
- Geissler Tube Ionizer
- FRC via N2 washout
- Ionization chamber
-phototube
-light emitted
-linearizer amplifier
NITROGEN ANALYZER
(GEISSLER TUBE IONIZER)
- How is it calibrated?
- How does it work?
- Low = 0%
- Use 100% O2 (has 0% N2)
-High= known N2 concentration
(Use room air (room air has 79% N2)
- Sample is pumped into an ionization chamber,
-a light is emitted &
-monitored by a phototube.
-Intensity of light is directly proportional to the percent of N2
-measures gas %
NITROGEN ANALYZER
(GEISSLER TUBE IONIZER
- What is a positive feature about this analyzer?
- Is a scrubber or desiccant needed?
Troubleshooting:
- What would cause sporadic or erratic zero readings which return to the expected value?
- How do you calculate % error?
- Fast. Breath by breath
- No. Not rebreathing.
- Aspiration of water through the needle valve.(moisture)
- Expected Value - Measured X 100
/ Expected Value
BLOOD GAS ANALYZERS
- What values are directly measured on a blood gas analyzer?
- What is the priority order of gases and what do they represent?
- PCO2, PO2, PH
- PCO2= ventilation.
(partial pressure of carbon dioxide)
—PO2= oxygenation
(partial pressure of oxygen)
—PH acid-base balance
BLOOD GAS ANALYZERS
- What is the name of the PCO2 electrode?
- What our terms that might be used to refer to this electrode?
- Severinghaus Electrode
- Nylon spacer.
— Teflon membrane
— smooth membrane
BLOOD GAS ANALYZERS
(PO2 Electrode)
- What is the name of the PO2 electrode?
- What features describe this electrode that are terms that might be used to refer to this electrode?
- Clark Electrode
- Membrane is not smooth
—Platinum cathode
— silver-silver chloride anode
—Polypropylene membrane (gas permeable)
— requires a polarizing voltage (battery or plugged in)
BLOOD GAS ANALYZERS
(PO2 Electrode)
- What is the maintenance required for this electrode?
- What are features of this electrode in respect to maintenance?
- If given answer option of “MYLAR MEMBRANE” = It IS NOT the Answer. Why?
- Periodic / routine cleaning of the tip with pumice
- Rough surface attracts protein causing blood clots.
- Does not allow sufficient diffusion of oxygen.
—
BLOOD GAS ANALYZERS
(PO2 Electrode)
Clark Electrode
Troubleshooting:
- What causes ERRATIC measurements & how it is fixed?
- What substances are pumice?
- What can cause an “unacceptable drift”?
- What will tears in the membrane cause?
- Membrane can attract protein and form clots.
—periodic & routine cleansing of the electrode tip with pumice. - Volcanic powder or dust.
- Depletion of buffers in the electrode.
- Inaccurate data
BLOOD GAS ANALYZERS
(PH Electrode)
- What is the name of this electrode?
- What are the two electrodes for PH measurement?
- Sanz Electrode
- Reference Electrode
— Measurement Electrode
BLOOD GAS ANALYZERS
(PH Electrode)
- Describe the Reference Electrode?
- Describe the Measurement Electrode?
- How does the ph electrodes work?
- Reference Electrode
—contains KCL Buffer Solution with a known pH
—it is sealed and has a ceramic plug - Contains buffer
—a silver-silver chloride wire
— a pH sensitive glass capillary tube - Introduction of a sample
—causes an electrical potential to develop.
—the voltage difference between the two electrodes
—is proportional to the pH difference.
CALIBRATION OF BLOOD GAS ANALYZERS
PH
- pH Calibration limits?
- Describe the calibration process?
- Balance (Low) 6.838
—Slope (high) 7.384 - If displayed pH solution value differs
—the balance is adjusted to the low value
—the slope is adjusted to the high value.
CALIBRATION OF BLOOD GAS ANALYZERS
PH
Troubleshooting:
- If the pH electrode cannot be calibrated, what would this indicate?
- A noisy erratic, unstable reading from any electrode, may indicate?
- What should be done if a clot forms?
- What should be checked first before changing the membrane?
- Inability to maintain temperature.
—must have constant temperature. - Re-membraning may be necessary.
- Run protein remover f/b 2 rinse cycles.
- Gas cylinder PSI & solution levels.
CALIBRATION OF BLOOD GAS ANALYZERS PCO2/PO2
- PCO2 calibration limits?
2 PO2 calibration limits?
- What unit of measure are the gases PO2 & PCO2 displayed as?
- What unit of measure are the calibration tanks in?
Note: Question may be:
When calibrating an electrode
- Convert % to mm Hg (Formula)?
- Shortcut for above formula?
- PCO2
—Balance (low)= 5%
—Slope (high) = 10% - PO2
—Balance (low)= 0%
—Slope (high) = 12% or 20% - mm Hg
- Percent
- (PB - PH2O) X Fractional Concentration of gas (gas % as decimal)
(760 - 47) X 0.12=85.56 - Gas % X 7
12 X 7=84
Pick closest answer.
OXYGEN ANALYZERS
- 2 Types of O2 analyzers?
- How measures?
- Which analyzer is like a Clark Electrode?
- Fuel Cell:
—Galvanic Fuel Cell
—Zirconium Fuel Cell
Polarographic - Creates a current
—as a result of oxidation/reduction of oxygen. - Polarographic (has electrolytes in electrode).