Equipment operation and quality control Flashcards
(282 cards)
During Bremsstrahlung
(Brems) radiation production
An electron approaching a positive nuclear
charge changes direction and loses energy.
Design characteristics of x-ray tube targets that
determine heat capacity include
- the rotation of the anode
- the diameter of the anode
- the size of the Focal spot
If exposure Factors of 85 kV, 400mA, and 12ms yield an output exposure of 150 mR, what is the milliroentgens per milliampere-seconds(mR/mAs)?
(A) 0.32
(B) 3.1
(C) 17.6
(D) 31
If the output or 4.8mAs was 150mR, then 1mAs is equal to 31.25mR (150mR ÷ 4.8mAs =31.25mR/mAs).
If the output or 4.8mAs was 150mR, then 1mAs is equal to 31.25mR (150mR ÷ 4.8mAs =31.25mR/mAs).
Vs/Vp=Ns/Np
To determine secondary I
Ns/Np=Ip/Is
When using the smaller field in a dual- field image
intensifier?
the image is magnified
Light-sensitive AEC devices are known as?
phototimers
AEC devices has 2 types
1 type = ionization chamber
2 type = phototimer
AEC devices = ionization chamber function as
Beneath the table top above the IR, for centering the par centered to it sensor and radiographed.
When a predetermined quantity o ionization
has occurred , the exposure terminates.
AEC devices = phototimer
Small fluorescent screen, is positioned beneath the cassette. When remnant radiation emerging from the patient exposes the IR and
exits, the fluorescent screen emits light. Once a
predetermined amount o fluorescent light has
been “seen” by the photocell sensor, the exposure is
terminated automatically.
The low product of milliampere-seconds and kilovoltage determines in single phase anode
heat units.
The filament of the cathode
assembly is made of
thoriated tungsten
System function to
compensate or changing patient/part thicknesses
during fluoroscopic procedures?
Automatic brightness control
Increased total brightness gain does not affect
spatial resolution
Delivery of large exposures to a cold anode for the use of exposures exceeding tube limitation can result in
-cracking to the anode
-rotor-bearing damage
The high-voltage, or step-up, transformer functions to
increase voltage to the necessary kilovoltage.
Circuit devices that permit electrons to flow in
only one direction are
solid-state diodes
The slit camera is considered the standard for (annual) measurement of the
-effective focal-spot size.
-spatial resolution
It cause pitting, for many small surface melts, of the anode’s focal track?
Repeated, frequent overloading
There are three types of beam restrictors—aperture diaphragms.
cones, cylinders, and collimators.
The most practical and efficient type is the collimator.
The type of x-ray tube designed to turn on and off rapidly, providing multiple short, precise exposures, is
grid-controlled
T e device that receives the remnant beam, converts it into light, and then increases the brightness of that light is the
image intensifier
A high-speed electron is decelerated as is attracted to a tungsten atom nucleus. T is results in
Bremsstrahlung radiation
If the positive nucleus of a tungsten atom attracts the electron, changing its course, a certain amount of energy is released during the “braking” action. T is energy is given up in the form of an x-ray photon called
Bremsstrahlung (“braking”) radiation.