Exam 2 - Mapleson Breathing Systems (Quick Tool) Flashcards
(37 cards)
What are the 5 major components of Mapleson system?
- Reservoir bag
- Unilateral Corrugated tubing
- APL valve
- Fresh gas inlet
- Patient connection
What 3 components are missing from Mapleson system?
- CO2 absorber
- Unidirectional Valves
- Separate inspiratory & expiratory limbs (tubing is unilateral)
What are other names for Mapleson circuits (2)?
- Carbon Dioxide washout circuits
- Flow-controlled Breathing Systems
Standard diameter of corrugated breathing tube on Mapleson system?
- Significance?
- What additional function can corrugated tubing potentially have during ventilation because of this?
- 22 mm
- Creates low resistance pathway for gases
- Also acts as potential reservoir for gases
Fill in blank: To minimize FGF requirements for Mapleson system, the volume of the breathing circuit must be _______ or ________ the _______.
- greater than or equal to tidal volume (TV)
Other names for APL valve?
- Pop-Off valve; Pressure Relief valve
Fill in blank: If gas inflow is greater than patient’s uptake and _______ uptake, the high pressure causes the APL to __________.
- circuit
- OPEN
T/F: APL is partially open during spontaneous ventilation
- False
- APL is fully open during spontaneous breathing
APL orientation during assisted ventilation:
- Partially closed
APL orientation @ 0 cmH20:
Valve is open
Pop-Off valve pressure when considered fully closed?
70 cmH20
Which Mapleson component generates positive pressure ventilation?
- Reservoir Bag
Which Mapleson(s) do not contain APL valves?
Mapleson E & Mapleson F
What is the expected effect with no APL valve present?
Less likely for excessive pressure to occur
Which component determines the type of Mapleson performance and classification?
The POSITION of the Fresh Gas Inlet
- Identify where FGF enters circuit
- Identify Mapleson system type & other name
- A: FGF enters near reservoir bag (opposite of pt)
- Mapleson A (MAgill’s)
- Identify location of APL valve on Mapleson A
- D: Located near the patient side
- Magill’s system is most efficient for _________ ventilation.
- When would you most likely us this system?
- Spontaneous
- Transport
Explain why Mapleson A is the worst choice for controlled ventilation?
- Rebreathing will occur during controlled ventilation unless MV greater than or = 20 L/min (FGF very high (3x MV) & unpredictable)
MV = minute ventilation
To prevent rebreathing, ______ must be greater than or equal to MV
- FGF
Rebreathing depends on FGF
Order the direction of fresh gas flow during spontaneous inspiration on Mapleson A system.
- FGF inlet
- circuit
- APL
- patient
- Identify where FGF enters system
- Identify Mapleson system type & other name
- Locate APL
- What components are missing?
- B (FGF enters near patient)
- Mapleson E (also called Ayre’s T-Piece)
- There’s no APL valve
- APL & Reservoir Bag
“Ayre was Extubated to a T-Piece”
In Mapleson E, corrugated tubing attached to the T-piece forms the ______.
Reservoir
Mapleson E indicated for what type of patients?
- Why?
- Spontaneous breathing
- No reservoir bag so unable to ventilate for patient