Final material E4-7 Flashcards
(51 cards)
The activities of chromatography adsorbents (stationary phases) such as silica and alumina are greatly affected by their water content. The most active adsorbents (silica gel/aluminum) are those that contain the least amount of moisture. Why is this the case?
water’s strong interaction w/ the polar stationary phase reduces the availability of adsorption sites for other polar compounds, diminishing the separation efficiency
essentially, silica is deactivated
You are told to mark the TLC plate with a pencil, NOT with a pen. Why should you NOT use a pen to mark the TLC plate?
what is ink soluble in?
the ink, given its soluble in organic solvents, will travel up the TLC plate w/ the solvent, interfering w/ the results
As silica would be deactivated by water, what real-life experiences tell you that it is best to keep
silica TLC plates stored in a desiccator? That it, where in real life have you purchased a product that
was stored with silica? What is the use of silica gel in those products?
Silica gel packets can be found in shoe boxes to absorb moisture aka water.
The melting points of the two constitutional isomers you used in Part A of the experiment are as
follows: ortho hydroxyacetophenone: 4-6 °C; para-hydroxyacetophenone: 109-111 °C. Explain the large difference in melting points exhibited by these two compounds.
inter vs intra, which is whuch
In ortho, an INTRAmolecular Hbond, can be formed making it less prone to react w/ nearby molecules thus, its melting point is lower
In para, can form INTERmolecular Hbond w/ nearby molecules thus, requiring higher energy + a higher melting point
Your TLC results for ortho-hydroxyacetophenone versus para-hydroxyacteophenone will indicate
that one compound is significantly less polar than the other. Based on an analysis of polarity (using
only dipole moments and electronegativities), which compound would you expect to be more polar? Draw structures of the two compounds and indicate the polarity using the cross-arrow to explain your answer (remember that dipoles are vector values).
Para has a larger dipole moment meaning its more polar than ortho meaning more acidic meaning higher bp
Now, when you perform the experiment, you will find that the ortho-hydroxyacetophenone is much LESS polar than para-hydroxyacetophenone. Discuss why the observed polarity of the two isomers in the TLC experiment is different from your analysis above. Think of your analysis used in question (4) above. Think about how the TLC result can be explained by the expected interaction between each isomer and the silica gel. Think of a qualitative approach of determining polarity by analyzing the expected distribution of charges within a molecule
what does the o isomer react w/ instead: mobile or stationary?
o < p in polarity
this is bc o interacted w/ the mobile phase instead of the stationery given the solvent used for the mobile phase was nonpolar
When looking at the structure of the o-isomer, the OH group has the ability to form an INTRAmolecular Hbond w/ the carbonyl group, making it LESS prone to interacting w/ the molecules of the silica plate. This allows it to pass through the plate faster due to less INTERmolecular forces occurring, resulting in a larger Rf value
In column chromatography, the particle size of the stationary phase makes a significant difference in separation efficiency. Why is this?
what does it increase?
Smaller particles in the stationary phase increases surface area for separation
What is the disadvantage to smaller particle size of the stationary phase?
2 points: increased what?
Smaller sample capacity and increased backpressure
backpressure = the resistance or force opposing the flow of the mobile phase (solvent or gas) through the column
aka, flow of the column is restricted
You have a mixture that is soluble only in 100% MTBE. You want to run your column in 5% MTBE/hexanes. Which method of column loading should you use in this case and why?
dry vs wet loading? is hexane non/polar
dry loading - use when mixture is only soluble in solvents that are MORE polar than the eluent of choice for the column
–> MTBE is MORE polar than hexane
Dry Loading is used when the solvent you dissolve your powdered mixture in is MORE POLAR than the mobile phase
Wet Loading is used when the solvent you dissolve your powdered mixture in is LESS POLAR than the mobile phase
You are performing a gradient elution and are instructed to begin with a low polarity solvent
(petroleum ether) and continue to increase solvent polarity w/ MTBE. Why are the mobile phases
added in order of increasing polarity rather than in the order of decreasing polarity?
2 points
achieve better separation of the components in your sample.
the compound of lowest polarity is eluted first and the most polar is eluted last
After running the column and separating the three components, you can evaporate the solvent to isolate the compounds, then add up the total mass recovered and compare to the total mass you began with. Why is this not 100%
over + below 100%
below 100% –> some of the solvent may still be attached to the stationary phase
over 100% –> not all solvent has been evaporated
In this experiment, you were able to “see” the separation of the compounds on the column due to
the fact that they are colored. However, most organic compounds are not colored making it more
difficult to run the column. Since many organic compounds do absorb UV light, it would seem like
we could just hold a UV light up to the glass column to observe the separation. However, this is not
possible. Why? (Consider whether or not you get a sun tan with sunlight shining through a window.)
what does the glass column do?
The glass of the column absorbs the UV rays making it unable to reach the product itself
A chemist makes a 1 M solution of compound A and another 1 M solution of compound B and
spots both on a TLC plate and runs the TLC. Upon visualizing the plate under UV light, the chemist
notices that the spot corresponding to “A” is much darker than the spot corresponding to “B”. Since
identical amounts were applied to the plate, how can the chemist explain the differences in spot
intensity? 2 POINTS
1) B could have absorbed into the TLC paper more
2) more uv absorbance = darker spot
(1) You are instructed to insulate the fractionating column for fractional distillation but not the
condenser. Why don’t you wrap the condenser with a cotton insulation?
(2) You are told to wrap the fractional distillation column with cotton. Why is this done?
(1) You dont wrap the condenser bc it needs to be cold so the vapor can be converted back into liquid
(2) to contain the heat of the column for vaporization
You are told that distillations should never be conducted to dryness. Why do you think this is important?
what will expand?
If done to dryness, no water will be left in the flask and heat will start expanding the glass which can lead to it shattering
Describe, using the ideal gas law, why heating a liquid in a closed system can lead to an explosion.
How do we make sure this does not happen in this experiment?
PV=nRT
Increased temp + pressure = constant volume
increased pressure causes an explosion
To avoid this, an opening into the external environment is needed to release the buildup of pressure the system is heated up.
If the thermometer in a distillation apparatus were positioned too high, what effect would this have
on the experimental results? What if it is positioned too low?
in both scenarios readings would NOT be accurate
too high = they would be slightly lower
too low = they would be slightly higher
You are instructed to tightly cap the vials containing your distillate. Why is this important and what
would happen if you did not cap the vials?
volatile compounds would evaporate + composition would change
if not capped, then sample evaporates (loss of sample)
For fractional distillation, you add copper mesh to the column. What is the purpose of the copper
mesh?
what is increased and why?
copper mesh increases the surface area, allowing for multiple repeated vaporizations + condensations cycles
You are instructed to use the same GC instrument for all your analyses and to record a GC trace of
the original mixture of unknowns on each day of analysis. Why is this important? 3 points
bc each GC instrument has different response factor, retention times and peak areas differ trial to trial, even for the same sample
You are instructed to rinse the syringe several times first with the mixture to be analyzed before
taking a sample for analysis. Why is this done?
You rinse the syringe to get rid of any leftover mixture that may have been remaining
Why is it important to press the start button immediately after injecting your sample of the GC?
2 points
to minimize the disturbance of gas flow
to make sure the entire sample reaches the column at the same time. (you dont want overlapping peaks!)
In GC, what is the most typical factor that determines the order of elution for different components
in a mixture
its a physical characteristic
boiling points, compounds with lower bp elude first
You are instructed to estimate the boiling points of the unknown compounds using data from the plots of the vapor temperature vs. volume of distillate.
Suppose the compositions indicate that the estimation is not accurate, **how do you expect the accurate boiling point to differ from the estimation? **
For each component (higher boiling and lower boiling), do you expect the accurate boiling point of the pure substance to be higher or lower than the estimation, and why?
increased/decreased bp from estimation
**lower boiling ** - accurate bp is expected to be lower than the estimated bp bc of contamination by higher boiling component
higher boiling - accurate bp is expected to be higher than the estimated bp bc of contamination by the lower boiling component