most important patient protection characteristic og a radiographic imaging system
filtration
state statutes require that general purpose of radiographic units have a minimum total filtration of ___
2.5 mm Al
filtration is measured through ___ of the xray beam
half value layer HVL
filtration frequency
annually or anytime a change has occurred in the xray tube or tube housing
Most systems today are equipped with _____ collimators
positive beam-limiting (PBL)
- x-ray field must coincide with the
light field of the variable-aperture
light-localizing collimator. - these fields are misaligned, the intended anatomy will be missed and unintended anatomy irradiated.
collimation
(collimation) Distance and centering indicators must be accurate to within ___ of the source-to-image receptor distance (SID).
2% and 1%
(collimation) distance indicator can be checked simply with a ___
location of the focal spot is usually marked on the ___
tape measure
x-ray tube housing.
- determines the spatial resolution of a radiographic imaging system.
focal spot size
3 tools are used for measurement of a focal-spot size:
pinhole camera
star pattern
slit camera
difficult to use and requires excessive exposure time.
pinhole camera
easy to use but has significant limitations for FS sizes less than 0.3 mm
star pattern
standard for measurement of effective FS size
slit camera
focal spot size frequency
annually or whenever an xray tube is replaced
alternative to focal spot size measurement is the use of ____ to determine limiting spatial frequency
line pair test tool
kvp calibration frequency
annually or whenever high-voltage generator components have changed significantly
measured kVp should be within ___ of the indicated kVp.
10%
variation in kVp of approximately __ is necessary to affect image optical density and radiographic contrast.
4%
- operator selectable on most radiographic consoles.
- particularly responsible for patient radiation dose and image optical density.
exposure time
exposure time accuracy frequency
annually or more frequently if a component of the console has undergone major repairs
Accuracy of __ is acceptable for exposure times of 10 ms or less.
20%
Automatic exposure control (AEC) must also be evaluated. It is done by exposing an IR through various thicknesses of ____
aluminum or acrylic
I(exposure time) nsertion of a ___ allows one to adequately assess the functioning of the backup timer.
lead filter
If the phototimer fails, the backup timer should terminate the exposure at ___, whichever occurs first
6s or 600 mAs