finals L7 Flashcards

(40 cards)

1
Q

most important patient protection characteristic og a radiographic imaging system

A

filtration

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

state statutes require that general purpose of radiographic units have a minimum total filtration of ___

A

2.5 mm Al

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

filtration is measured through ___ of the xray beam

A

half value layer HVL

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

filtration frequency

A

annually or anytime a change has occurred in the xray tube or tube housing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Most systems today are equipped with _____ collimators

A

positive beam-limiting (PBL)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q
  • x-ray field must coincide with the
    light field of the variable-aperture
    light-localizing collimator.
  • these fields are misaligned, the intended anatomy will be missed and unintended anatomy irradiated.
A

collimation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

(collimation) Distance and centering indicators must be accurate to within ___ of the source-to-image receptor distance (SID).

A

2% and 1%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

(collimation) distance indicator can be checked simply with a ___
location of the focal spot is usually marked on the ___

A

tape measure
x-ray tube housing.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q
  • determines the spatial resolution of a radiographic imaging system.
A

focal spot size

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

3 tools are used for measurement of a focal-spot size:

A

pinhole camera
star pattern
slit camera

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

difficult to use and requires excessive exposure time.

A

pinhole camera

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

easy to use but has significant limitations for FS sizes less than 0.3 mm

A

star pattern

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

standard for measurement of effective FS size

A

slit camera

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

focal spot size frequency

A

annually or whenever an xray tube is replaced

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

alternative to focal spot size measurement is the use of ____ to determine limiting spatial frequency

A

line pair test tool

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

kvp calibration frequency

A

annually or whenever high-voltage generator components have changed significantly

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

measured kVp should be within ___ of the indicated kVp.

18
Q

variation in kVp of approximately __ is necessary to affect image optical density and radiographic contrast.

19
Q
  • operator selectable on most radiographic consoles.
  • particularly responsible for patient radiation dose and image optical density.
A

exposure time

20
Q

exposure time accuracy frequency

A

annually or more frequently if a component of the console has undergone major repairs

21
Q

Accuracy of __ is acceptable for exposure times of 10 ms or less.

22
Q

Automatic exposure control (AEC) must also be evaluated. It is done by exposing an IR through various thicknesses of ____

A

aluminum or acrylic

23
Q

I(exposure time) nsertion of a ___ allows one to adequately assess the functioning of the backup timer.

24
Q

If the phototimer fails, the backup timer should terminate the exposure at ___, whichever occurs first

A

6s or 600 mAs

25
- ability of a radiographic unit to produce a constant radiation output for various combinations of mA and exposure time - determined by a precision radiation dosimeter that measures radiation intensity at various combinations of mA and exposure time.
exposure linearity
26
(exposure linearity) radiation output for adjacent mA stations should be within
10%
27
exposure linearity frequency
annually or after significant change or repair of operating console
28
method of assessing exposure linearity is valid or not valid if the exposure timer is inaccurate
not valid
29
(exposure linearity) the mGy /mAs value should be within ___ between adjacent mA stations.
10%
30
Two methods are used to evaluate exposure reproducibility:
- One makes a series of at least three exposures at the same technique factors, having changed technique controls between each exposure. - If the result is not reproducible, this is usually the result of error in the kVp control.
31
filtration tolerance
> 2.5 mm Al
31
exposure reproducibility - Mathematical formulas can be used to determine reproducibility in both instances. - These formulas basically require that output radiation intensity should not vary by ___
more than plus or minus 5%.
32
Two methods are used to evaluate exposure reproducibility:
One can select a combination of technique factors and hold them constant for a series of 10 exposures.
32
collimation frequency
semiannually
33
collimation tolerance
+2% SID
33
focal spot size tolerance
+50%
33
calibration of kvp tolerance
+10%
34
exposure time accuracy tolerance
+5% > 10 ms + 20% < 10ms
34
exposure linearity tolerance
+10%
35
exposure reproducibility tolerance
+5%