finals L8 Flashcards
(35 cards)
- covers generator, tube, cables, control panel, table, tube stand, and bucky.
- important to ensure that all equipment are working efficiently and accurate records are kept.
xray equipment
- part of the electrical system of an x-ray machine which determines radiation output.
- inconsistency in radiation output will adversely influence the quality of the radiograph.
- The kV, mA, and exposure time must be consistent on all settings.
generator
Many factors can influence radiation output. The ones to be considered are:
- Fluctuations in the main electrical supply
- Inconsistency or inaccuracies in kVp, mA, and exposure time.
(cleaning of generator) Check that the equipment is earthed by ___ attached to the metal covers of the unit or by an earthly conductor carried in the main cable of a mobile unit
copper tape
radiation output reproducibility test frequency
at the start of a QC program
yearly
as necessary
radiation output reproducibility test equipment
- water phantom, consisting of a plastic flagon or bucket filled to a depth of 10 cm
- 2 sheets of lead rubber
- 1 loaded cassette
(radiation output reproducibility test procedure) Cover three quarters of the cassette with lead rubber, ____, leaving an end section uncovered.
crosswise
(radiation output reproducibility test procedure) Set a ___ SID.
100 cm
(radiation output reproducibility test procedure)
Set an exposure that will produce a low density on the film (___). ___ and a ____ is suggested. It should be just possible to read newspaper print through the density produced (you may find it necessary to experiment before starting the test)
density of 1
80 kVp
low mAs
(radiation output reproducibility test procedure)
Repeat this procedure for each of the remaining three sections. Making sure that you have identical conditions. When placing the lead rubber allow for a space of about ___ between each exposed area.
2 mm
radiation output reproducibility test
If the densities are not the same
repeat the test at different times of the day. Mains fluctuations may be the problem.
radiation output reproducibility test
If the densities remain inconsistent,
call the x-ray engineer
radiation output reproducibility test
If the inconsistencies are unacceptable,
then stop using the unit until the fault has been corrected.
constancy of radiation output at different ma settings frequency
start of a QC program
yearly
as necessary
constancy of radiation output at different ma settings equipment
- step wedge or 10 cm water phantom (consisting of a plastic flagon or bucket filled)
- 2 sheets of lead rubber
- 1 loaded cassette
constancy of radiation output at different ma settings
suggested exposures
Exposure 1: 80kV, 10mAs (50mA, 0.2s)
Exposure 2: 80kV, 10mAs (100mA, 0.1s)
Exposure 3: 80kV, 10mAs (200mA, 0.05s)
If the out of step density in each test was the second exposure (100 mA), this indicates that the fault lies with the 100 mA setting.
80 kV 50 mAs (50 mA, 1.0 s)
80 kV 50 mAs (100 mA, 0.5 s)
80 kV 50 mAs (200 mA, 0.25 s)
constancy of radiation output at different ma settings action
If the densities continue to be inconsistent
call an x-ray technician
constancy of radiation output at different ma settings action
If the inconsistencies are minimal
continue to use the unit.
constancy of radiation output at different ma settings action
If the inconsistencies are large
stop using the unit until the fault has been corrected.
test for kvp accuracy frequency
When the equipment is first installed.
Yearly
As necessary
test for kvp accuracy equipment
kVp meter
generator to be tested
test for kvp accuracy
Place the sensor in the middle of the x-ray beam at an SID of
100 cm
test for kvp accuracy
Take readings at 10 kVp intervals from ___ to the maximum kVp available.
50 kVp